• Title/Summary/Keyword: therapeutic efficacy monitoring

Search Result 21, Processing Time 0.014 seconds

Acoustic outputs from clinical extracorporeal shock wave lithotripsy devices (임상에서 사용중인 체외충격파쇄석기의 음향 출력 분포)

  • Jong Min Kim;Oh Bin Kwon;Jin Sik Cho;Sung Joung Jeon;Ki Il Nam;Sung Yong Cho;Min Joo Choi
    • The Journal of the Acoustical Society of Korea
    • /
    • v.42 no.5
    • /
    • pp.469-490
    • /
    • 2023
  • Survey was carried out on the acoustic outputs from 12 shock wave fields produced by the 10 extracorporeal shock wave lithotriptors whose technical documents are available, among the 33 devices approved by the Ministry of Food & Drug Safety (MFDS).The results show that the acoustic outputs (P+, P-, efd, and E), critical to the therapeutic efficacy and the patient safety, are largely different between the devices. The maximum values of P+, P-, efd, and E vary up to 2.08, 3.72, 3.89, and 15.98 times, respectively. The acoustic output parameters are not thoroughly provided in the technical documents, and some of data (eg. efd) are suspected to be abnormal outside usual ranges. The large device to device differences in the shock wave outputs are likely to undermine equivalence between the ESWL devices approved for the same indication. To verify the reliability of the data in the technical documents of the approved devices and to confirm if the acoustic outputs from the devices in clinical use are the same as those in their technical documents, an authorized test laboratory should be available. A postapproval monitoring led by the regulatory agency is suggested to maintain the acoustic outputs from the ESWL devices that suffer from degrading in performance due to aging.