• Title/Summary/Keyword: stopped-flow instrument

Search Result 1, Processing Time 0.016 seconds

Real Time Scale Measurement of Inorganic Phosphate Release by Fluorophore Labeled Phosphate Binding Protein (형광단이 붙어 있는 인산결합 단백질에 의한 인산 배출의 실시간 측정)

  • Jeong Yong-Joo
    • Journal of Life Science
    • /
    • v.15 no.6 s.73
    • /
    • pp.935-940
    • /
    • 2005
  • Fluorescence change of coumarin labeled phosphate binding protein (PBP-MDCC) was monitored to measure the amount of released inorganic phosphate ($P_{i}$) during nucleoside triphosphate (NTP) hydrolysis reaction. After purification of PBP-MDCC, fluorescence emission spectra showed that fluorescence responded linearly to $P_{i}$ up to about 0.7 molar ratio of $P_{i}$ to protein. The correlation of fluorescence signal and $P_{i}$ standard was measured to obtain [$P_{i}$] - fluorescence intensity standard curve on the stopped-flow instrument. When T7 bacteriophage helicase, double-stranded DNA unwinding enzyme using dTTP hydrolysis as an energy source, reacted with dTTP, the change of fluorescence was able to be converted to the amount of released $P_{i}$ by the $P_{i}$ standard curve. $P_{i}$ release results showed that single-stranded Ml3 DNA stimulated dTTP hydrolysis reaction several folds by T7 helicase. Instead of end point assay in NTP hydrolysis reaction, real time $P_{i}$-release assay by PBP-MDCC was proven to be very easy and convenient to measure released $P_{i}$.