• Title/Summary/Keyword: power Doppler ultrasonography

Search Result 11, Processing Time 0.015 seconds

A Study on Changes in Body Surface Temperature and Thermal Effect According to Ultrasound Mode (초음파 진단모드에 따른 체표면 온도변화와 열효과에 관한 고찰)

  • Yang, Sung-Hee;Lee, Jin-Soo
    • Journal of radiological science and technology
    • /
    • v.40 no.2
    • /
    • pp.213-218
    • /
    • 2017
  • Recently, as the number of high-risk pregnancies increases, the use of new techniques such as Doppler, which have higher acoustic power than in the past, has been increasingly used in prenatal diagnosis and guidelines have been set up by various organizations to prevent excessive exposure. Therefore, in this study, we tried to investigate the temperature change of the body surface for each test mode according to the long time ultrasound examination and to examine the exposure time which is not influenced by the thermal effect. B mode, C mode, and PD mode according to time, and the temperature difference between exposed and unexposed sites were compared. As a result, the B mode showed a significant difference in the temperature change from 10 minutes, 50 minutes after exposed, 20 minutes from the C mode, and 30 minutes from the PD mode (p<0.01). In all three modes, the temperature difference was different (p<0.000), and PD mode was the most sensitive to temperature change. Also, it was found that the temperature rise time was shortened with the increase of the ultrasonic exposure time. Therefore, it is recommended that ultrasonography to observe the embryo or fetus should be used only for diagnostic purposes, avoiding excessive test time.