• 제목/요약/키워드: microbeam concentrator

검색결과 1건 처리시간 0.017초

Modification of conventional X-ray diffractometer for the measurement of phase distribution in a narrow region

  • Park, Yang-Soon;Han, Sun-Ho;Kim, Jong-Goo;Jee, Kwang-Yong;Kim, Won-Ho
    • 분석과학
    • /
    • 제19권5호
    • /
    • pp.407-414
    • /
    • 2006
  • An X-ray diffractometer for spatially resolved X-ray diffraction measurements was developed to identify phase in the narrow (micron-scaled) region of high burn-up fuels and some nuclear materials. The micro-XRD was composed of an X-ray microbeam alignment system and a sample micro translation system instead of a normal slit and a fixed sample stage in a commercial XRD. The X-ray microbeam alignment system was fabricated with a microbeam concentrator having two Ni deposited mirrors, a vertical positioner, and a tilt table for the generation of a concentrated microbeam. The sample micro translation system was made with a sample holder and a horizontal translator, allowing movement of a specimen at $5{\mu}m$ steps. The angular intensity profile of the microbeam generated through a concentrator was symmetric and not distorted. The size of the microbeam was $4,000{\times}20{\mu}m$ and the spatial resolution of the beam was $47{\mu}m$ at the sample position. When the diffraction peaks were measured for a $UO_2$ pellet specimen by this system, the reproducibility ($2{\Theta}={\pm}0.01^{\circ}$) of the peaks was as good as a conventional X-ray diffractometer. For the cross section of oxidized titanium metal, not only $TiO_2$ in an outer layer but also TiO near an oxide-metal interface was observed.