• Title/Summary/Keyword: linac

Search Result 305, Processing Time 0.027 seconds

KLYSTRON-MODULATOR SYSTEM PERFORMANCES FOR PLS 2-GeV LINAC (포항 20억 전자볼트 선형가속기 클라이스트론-모듈레이터 시스템의 성능)

  • Park, S.W.;Park, S.S.;Lee, K.T.;Oh, J.S.;Cho, M.H.;NamKung, W.
    • Proceedings of the KIEE Conference
    • /
    • 1995.07c
    • /
    • pp.1365-1367
    • /
    • 1995
  • The PLS 2-GeV linac employs 11 units of high-power pulsed klystrons(80MW) as the main RF sources. The matching modulators of 200 MW(400kV, 500 A) can provide a flat-top pulse width of 4.4 ${\mu}s$ with a maximum pulse repetition rate of 120 Hz at the full power level. For a good stability of electron beams, the pulse-to-pulse flat-top voltage variation of a modulator requires less than 0.5%. In order to achieve this goal, we stabilized high voltage charging power supplies within 1% by a phase controlled SCR voltage regulator. In addition, we employed ac/dc feedback together with a resistive De-Q'ing system to achieve far less than 0.5% variation of the PFN charging voltage. This paper presents the main feature of the klystron-modulator system and the characteristics of the pulsed high-power RF system performance during the beam injection operation for the Pohang Light Source commissioning.

  • PDF

Preliminary Study of the Siemens Primus Linac MLC modelling using BEAM Monte Carlo code (BEAM 몬테칼로 코드를 이용한 Siemens Primus 선형가속기 다엽콜리메이터의 모델링 예비연구)

  • Cheong, Kwang-Ho;Suh, Tae-Suk;Cho, Byung-Chul;Park, Sung-Ho
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2004.11a
    • /
    • pp.29-32
    • /
    • 2004
  • In this study, we had modelled Siemens type MLC using BEAM Monte Carlo code and tested the feasibility of the modelling. To model the Primus linac MLC, we had measured the actual dimensions of MLC and each leaves, then approximated the leaf shape. VARMLC component module was used for the modelling and leakage, tongue-and-groove effect were also considered. Simulation result showed the good agreement with the film measurement.

  • PDF

PHOTO-NEUTRON SOURCE USING 2 GEV ELECTRON LINAC FOR RADIATION SHIELDING RESEARCH

  • Lee, Hee-Seock;Bak, Joo-Shik;Chung, Chin-Wha;Ban, Syuichi;Shin, Kazuo;Sato, Tatsuhiko
    • Journal of Radiation Protection and Research
    • /
    • v.26 no.3
    • /
    • pp.333-335
    • /
    • 2001
  • The 2 GeV electron linac, the injector of the Pohang Light Source, was used as a photo-neutron source for radiation shielding research. The operational beam parameters are the nominal electron intensity of $0.5\;{\sim}5\;nC/sec$, the repetition rate of 10 Hz, and the beam pulse length of 1.0 nsec. One electron beam line was modified in order to install the target systems for producing pulsed photo-neutrons. The neutron spectrum and intensity were investigated by the time-of-flight technique. The reliable maximum energy of the measured neutrons was about 500 MeV. The number of neutrons above 20 MeV produced by one 1 GeV electron in a thick Pb target was about $6.45{\times}10^{-4}/sr$ at 90 degrees to the beam axis. The status of the photo-neutron source and the application research are presented.

  • PDF

Electron Energy Distribution for a Research Electron LINAC

  • Lim, Heuijin;Lee, Manwoo;Yi, Jungyu;Kang, Sang Koo;Kim, Me Young;Jeong, Dong Hyeok
    • Progress in Medical Physics
    • /
    • v.28 no.2
    • /
    • pp.49-53
    • /
    • 2017
  • The energy distribution was calculated for an electron beam from an electron linear accelerator developed for medical applications using computational methods. The depth dose data for monoenergetic electrons from 0.1 MeV to 8.0 MeV were calculated by the DOSXYZ/nrc code. The calculated data were used to generate the energy distribution from the measured depth dose data by numerical iterations. The measured data in a previous work and an in-house computer program were used for the generation of energy distribution. As results, the mean energy and most probable energy of the energy distribution were 5.7 MeV and 6.2 MeV, respectively. These two values agreed with those determined by the IAEA dosimetry protocol using the measured depth dose.

Optimization of CORVUS Planning System with PRIMART Linac for Intensity Modulated Radiation Therapy

  • Lee, Se-Byeong;Jino Bak;Cho, Kwang-Hwan;Chu, Sung-sil;Lee, Suk;Suh, Chang-ok
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2002.09a
    • /
    • pp.83-85
    • /
    • 2002
  • Yonsei Cancer Center introduced an IMRT System at the beginning of February, 2002. The system consists of CORVUS(NOMOS) inverse planning machine, LANTIS(SIEMENS), PRIMEVIEW and PRIMART Linac(SIEMENS). The optimization of CORVUS planning system with PRIMART is an important work to get an efficient treatment plan. So, we studied two Finite Size Pencil Beams, 1.0 x 1.0 cm$^2$ and 0.5 x 1.0 cm$^2$, and four leaf transmission sets, 5%, 10%, 20%, 33%. We compared the dose distribution of target volume and delivery efficiency of the plan results.

  • PDF

The Characteristic Curves of Commercial Medical X-ray Films (상용 의학용 X-ray 필름의 특성곡선)

  • Heo, Hoon;Jeong, Yeon-Tae;Lee, Jae-Sung
    • Journal of the Korean Graphic Arts Communication Society
    • /
    • v.19 no.2
    • /
    • pp.12-21
    • /
    • 2001
  • For the purpose of determining characteristics of widely used commercial medical x-ray films, which are used for obtaining the Linac-grams for radiational treatment of cancers, we placed several commercial x-ray films at a fixed distance form the linear accelerator. After varying the exposed amount of radiation step by step, we could obtain a continually increasing density image for each film whose densities were determined by microdensitometer readings. The characteristic curves of the films were obtained by plotting the densities vs. the exposed radiation amounts, and their ${\gamma}$ values were determined. These values can be used to suggest a minimum necessary amount of exposed radiation to get a useful Linac-gram. The measured ${\gamma}$ values of the characteristic curves of the Kodak-DVP/RA-1 film were 1.73 when used 6MV x-ray, 1.70 when used 15MV of intensity. For the Konica-AX film, ${\gamma}$ values were 1.29 and 1.18 respectively. The result suggests that the effective conditions for high resolution of a L-gram be 6 MV of x-ray intensity and about 3 rad of exposed dose on a Kodak-DVP/RA-1 film.

  • PDF

Dose Characteristics of Stereotatic Radiosurgery CONE used for ML-6M Linear Accelerator (ML-6M LINAC에 장착한 Radiosurgery Cone의 Beam 특성에 대한 고찰)

  • Moon, Un-Chull;Kim, Hyun-Jin;Kim, Young-Suk
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.7 no.1
    • /
    • pp.15-31
    • /
    • 1995
  • The radiosurgery treatment is one time, non surgical approach to the treatment of patients with intracranial disorders whose conditions would be difficult or dangerous to treat with conventional sugical procedures. The LINAC based radiosurgery is based on the combination of multiple isocentric arc irradiation with small fields centered in the stereotactic target. The absorption of the beam in a tissue equivalent medium, such as water, as well as the uniformity, or profile, of the beam must be precisely documented. The beam characteristics and dosimetric measurememts of the 6MV X-ray beam from a ML-6M linear accelerator are examined. The percent depth dose (PDD) and beam profile (including flatness, symmetry and penumbra) is calibrated with the radiosurgery cone in water phantom. The cone is made of lead which size is from $10{\times}10mm{\phi}$ to $30{\times}30mm{\phi}$ All of these dosimetric measurements sufficiently characterized the beam to permit safe clinical use.

  • PDF

Improvement of Beam-Quality Evaluation Method for Medical Linear Accelerator Using Magnetic Field

  • Kim, Jeongho;Han, Manseok;Yoo, Sejong;Kim, Kijin;Cho, Jae-Hwan
    • Journal of Magnetics
    • /
    • v.20 no.2
    • /
    • pp.120-128
    • /
    • 2015
  • Beam-quality of medical linac evaluations vary by diverse factors. Because conventional beam-quality evaluation methods yield fragmentary results, a new beam-evaluation method is suggested, and its feasibility is evaluated. The PDDs (percentage depth doses) of 6 MV (Mega-voltage) and 10 MV photon, R (Range) of a 6 MeV (Mega Electron-voltage) and 9 MeV electron were measured and compared with the conventional evaluation methods, and the improved methods $PDD^{10}{_5}$, $PDD^{20}{_{10}}$, $PDD^{30}{_{20}}$, $PDD^{20}{_5}$, $PDD^{30}{_{10}}$, and $R^{70}{_{50}}$, $R^{50}{_{30}}$, $R^{70}{_{30}}$ as the magnetic field of the bending magnet was changed to +2% to -2%, and the results were compared. The comparison showed that the improved methods exhibit a higher discrimination than the conventional methods in each energy regime. $PDD^{10}{_5}$, $PDD^{30}{_{20}}$, $PDD^{30}{_{10}}$ and $R^{70}{_{50}}$, $R^{50}{_{30}}$ should be applied. These methods exhibit a higher discrimination in each energy regime than conventional beam-quality evaluation methods; therefore, they should be used for beam-quality evaluation according to the magnetic field variation.

The Variation of PSF Induced Enlarged Wedged Fields (확장된 쐐기조사야에 의한 조직산란계수의 변화)

  • Lee, Jeong-Woo;Cho, Hwa-Seop;Park, Seong-Ryul
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.10 no.1
    • /
    • pp.97-101
    • /
    • 1998
  • In recent days, although many kinds of beam modifiers are developing and using for clinical purposes in accordance with progressing medical engineering, physical wedges are preferred to use as a beam modifier by a lot of institutions until now because of cost, complexities of dosimetry and mechanical uncertainties. According to progressing technology, available field size of wedge is more enlarger than that of old model LINAC. Because field size dependence of wedged fields increases in new model LINAC, we was trying to know that how much different PSFs are in enlarged wedged fields compared with open fields. In small or middle size of fields($4{\times}4{\sim}15{\times}15cm$), there are only a few percents of PSF variation between open and wedged fields. But there are $2{\sim}8\%\;variations\;in\;relatively\;large\;fields(20{\times}20{\sim}30{\times}40cm)$.

  • PDF