Preliminary Study of the Siemens Primus Linac MLC modelling using BEAM Monte Carlo code

BEAM 몬테칼로 코드를 이용한 Siemens Primus 선형가속기 다엽콜리메이터의 모델링 예비연구

  • Cheong, Kwang-Ho (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Suh, Tae-Suk (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Cho, Byung-Chul (Department of Radiation Oncology, The Hallym University Sacred Heart Hospital) ;
  • Park, Sung-Ho (Department of Radiation Oncology, The Hallym University Sacred Heart Hospital)
  • 정광호 (가톨릭의대 의공학교실) ;
  • 서태석 (가톨릭의대 의공학교실) ;
  • 조병철 (한림성심병원 방사선종양학과) ;
  • 박성호 (한림성심병원 방사선종양학과)
  • Published : 2004.11.26

Abstract

In this study, we had modelled Siemens type MLC using BEAM Monte Carlo code and tested the feasibility of the modelling. To model the Primus linac MLC, we had measured the actual dimensions of MLC and each leaves, then approximated the leaf shape. VARMLC component module was used for the modelling and leakage, tongue-and-groove effect were also considered. Simulation result showed the good agreement with the film measurement.

본 연구에서는 방사선치료장치의 몬테칼로 시뮬레이션에 널리 이용되는 BEAM 코드를 이용하여 Siemens사의 Primus 선형가속기 다엽콜리메이터(MLC)를 모델링한 후 그 적합성을 측정 데이터와의 비교를 통하여 검증하였다. 모델링을 위해 Primus 선형가속기의 다엽콜리메이터의 치수를 실측하였고 근사화를 통해 VARMLC 모듈을 사용하여 모델링하였다. 모델링 시에 leakage 및 tongue-and-groove 효과를 고려하였다. 시뮬레이션 결과는 필름 측정 결과와 잘 일치함을 확인하였으며, 본 연구에서의 방법이 Siemens 사의 다엽콜리메이터의 모델링에 유효함을 확인할 수 있었다.

Keywords