• 제목/요약/키워드: heat flux gauge

검색결과 17건 처리시간 0.015초

Measurement of Heat Flux in Rocket Combustors Using Plug-Type Heat Flux Gauges

  • Kim, Min Seok;Yu, I Sang;Kim, Wan Chan;Shin, Dong Hae;Ko, Young Sung
    • International Journal of Aeronautical and Space Sciences
    • /
    • 제18권4호
    • /
    • pp.788-796
    • /
    • 2017
  • This paper proposes a new measurement method to improve the shortcomings of an existing integral method for measuring heat flux in plug-type heat flux gauges in the high-temperature and high-pressure environments of liquid-rocket combustors. Using the existing integral measurement method, the calculation of the surface area for the heat flux in the gauge exhibits error in relation to the actual surface area. To solve this problem, transient profiles obtained from ANSYS Fluent were used to calculate unsteady heat flux as it adjusted to the measured temperature. First, a heat flux gauge was designed and manufactured specifically for use in the high-temperature and high-pressure conditions that are similar to those of liquid rocket combustors. A calibration test was performed to prove the reliability of the manufactured gauge. Then, a combustion experiment was conducted, in which the gauge was used to measure unsteady heat flux in a liquid rocket combustor that used kerosene and liquid oxygen as propellants. Reasonable heat flux values were obtained using the gauge. Therefore, the proposed measurement method is considered to offer significant improvement over the existing integral method.

강재 급속냉각용 고온 열유속게이지 개발 (Development of High-Temperature Heat Flux Gauge for Steel Quenching)

  • 이정호
    • 열처리공학회지
    • /
    • 제23권6호
    • /
    • pp.323-330
    • /
    • 2010
  • The present study was motivated by increasing demands on quantitative measurements of the heat flux through the water cooling and quenching process of hot steel. The local heat flux measurements are employed by a novel experimental technique that has a function of high-temperature heat flux gauge in which test block assemblies are directly used to measure the heat flux variation during water cooling and quenching of hot steel. The heat flux can be directly achieved by Fourier's law and is also compared with numerical estimation which is solved by inverse heat conduction problem (IHCP). The high-temperature heat flux gauge developed in this study can be applicable to measure cooling rate and history during the actual cooling applications of steelmaking process. In addition, the measurement uncertainty of heat flux is calculated by a quantitative uncertainty analysis which is based on the ANSI/ASME PTC 19.1-2005 standard.

접착형 박막 측온게이지식 열류센서의 개발에 관한 연구

  • 한응교;박두원;최규철;노병옥;이명호
    • 한국정밀공학회:학술대회논문집
    • /
    • 한국정밀공학회 1991년도 추계학술대회 논문집
    • /
    • pp.80-86
    • /
    • 1991
  • In this study an the development of the heat flux sensor with thermocouple, the heat-treated adhesive type film nickel gauge(Ni-Gauge) for measuring temperature was used. and this Ni-Gauge is not only comparable for platinum gauge(Pt-Gauge) in linearity, but also economically cheap. And from this viewpoint. numerical analysis is essential to investigate charateristics of sensor, since this analysis is capable of simulating precise boudary condition and practical conditions and so on. By the way. there are many types of heat flux sensor. of which adhesive type flux sensor is common, in the study this type of heat flux sensor was chogen. and analysis of the sensor is considered as a kind of open cavity figure. performed by SIMPLER algorithm. As a result. through temperature distributial of the sensor by numerical analysis in steady and unsteady state, the dracteristics of the acdhesive type heat flux sensor(lst heat flux sensor) acooodinge to heat flux, that is, outvoltage, sensitivity. and responsibilty could be evaluated. in addition, time of improved heat flux sensor(2nd heat flux sensor) could be predicted fran the reflectian of proper operating temperature(l50 $^{\circ}C$)of theNi-Gauge.

  • PDF

접착형 박막 온도측정 게이지식 열전대센서의 개발에 관한 연구 (Development of Heat Flux Sensor Using Adhesive Type Film Gauge for Measuring Temperature)

  • 한응교;최규철;노병욱;박두원
    • 한국정밀공학회지
    • /
    • 제9권2호
    • /
    • pp.52-60
    • /
    • 1992
  • In this study on the development of the heat flux sensor, unlike the common heat-flux sensor with thermocouple, the heat-treated adhesive-tupe film nickel-gauge was used in measuring temperature. The proposed its Ni-gauge is bound to be compatible with platinum gauge(Pt-Gauge) in its linearity. It is also considered to be cheap in economical sense. In the evaluation of it's performance, the numerical analysis is essential to investigate charateristics of proper sensor and the adequate analsis is depended upon boundary conditions and actual conditions. There are many types of heat flux sensor in the market, and adhexive type flux sensor is most common. In the present investigation, this type of heat flux sensor had been chosen. The figure of the sensor under consideration is an open cavity type, which is calculated numerically by SIMPLER algorithm. The temperature distributions of the sensor predicted by numerical calculation for steady and unsteady states are able to give the chacteristics of the adhesive type heat flux sensor(1st heat flux sensor) according to the heat flux. It means that the outvoltage, the sensitivity, and the performances of responsibility could be evaluated as a result. Through this analysis improved heat flux sensor(2nd heat flux sensor) could be predicted with the reflection of proper operating temperature($150^{\circ}C$) of the Ni-gauge.

  • PDF

강재 열처리용 다점 열유속 측정 기술 개발 (Development of Multi-point Heat Flux Measurement for Steel Quenching)

  • 이정호;오동욱;도규형;김태훈
    • 열처리공학회지
    • /
    • 제25권4호
    • /
    • pp.181-189
    • /
    • 2012
  • The demand on quantitative measurement of the heat flux is motivated in making higher-quality steel product through a water quenching process of plate mill. To improve a spatial degree of heat flux measurement, the multi-point heat flux measurement was carried out by a unique experimental technique that has a combination of the existing single-point heat flux gauge. The corresponding heat flux can be easily determined by Fourier's law in a conventional way. The multi-point heat flux gauge developed in this study can be applicable to measure the surface heat flux, the surface heat transfer coefficient during a water quenching applications of steelmaking process. The results exhibit different heat transfer regimes; such as single-phase forced convection, nucleate boiling, and film boiling, that are occurred in close proximity on the multi-point heat flux gauge quenched by water impinging jet.

제조사 보정상수에 기인한 열유속계의 측정 오차 (Measurement Bias of Heat Flux Gauge based on Calibration Constant supported by Manufacturer)

  • 김성찬
    • 한국화재소방학회논문지
    • /
    • 제32권4호
    • /
    • pp.1-6
    • /
    • 2018
  • 본 연구는 Schmidt-Boelter형 열유속계의 측정 불확실도를 정량화하기 위한 기초연구로서 제조사에서 제공되는 보정상수의 측정오차를 평가하였다. 보정실험은 미국립표준기술원(NIST) 화재연구부의 2000 W급 할로겐-텅스텐 램프를 열원으로 적용한 열유속계 보정장치를 이용하였으며 동일 열유속 조건들에 대해 보정대상 열유속계와 표준열유속계를 비교하여 보정상수를 얻었다. 3개 제조사의 열유속계에 대한 보정실험 결과를 제조사의 보정상수와 비교하였다. 열유속계의 출력신호의 섭동값은 $1{\cdot}$를 넘지 않았으며 보정실험결과와 제조사의 보정상수의 상대오차는 1.5% ~ 14.3%까지 차이를 보였다. 본 연구는 정확한 열유속을 측정하기 위해 지속적이고 주기적인 보정과정이 필요함을 보여준다.

수치해석을 통한 유도가열 코일의 설계 및 설계인자의 민감도 해석 (Design and Sensitivity Analysis of Design Factors for Induction Heating System)

  • 오동욱;김태훈;도규형;박장민;이정호
    • 열처리공학회지
    • /
    • 제26권5호
    • /
    • pp.233-240
    • /
    • 2013
  • Rapid and homogeneous heating in heat treatment has been a challenging engineering issue throughout a heating temperature over $1,000^{\circ}C$. Induction heating has been widely used in field of heat treatment compared with conventional heating system. Advantages in homogeneous heating, simple fabrication, and repeatable use can be efficiently made with the induction heater. In this paper, numerical analysis of an induction coil system for heat flux gauge heating is performed. The effect of configuration on the heating performance was considered in various cases of the coil radius, distance between the winding, relative height difference between the heat flux gauge and the coil, and the applied current frequency. Temperature distribution within the heat flux gauge at frequency-steady state was calculated with a finite element method. Sensitivity analysis was also performed and the relative importance of 2 key parameters; coil radius, distance between the winding, were taken as main contributors for induction heating.

냉각수 온도에 따른 수분류 충돌제트의 열전달 특성 연구 (Effect of Cooling Water Temperature on Heat Transfer Characteristics of Water Impinging Jet)

  • 이정호;유청환;도규형
    • 열처리공학회지
    • /
    • 제23권5호
    • /
    • pp.249-256
    • /
    • 2010
  • Water jet impingement cooling has been widely used in a various engineering applications; especially in cooling of hot steel plate of steelmaking processes and heat treatment in hot metals as an effective method of removing high heat flux. The effects of cooling water temperature on water jet impingement cooling are primarily investigated for hot steel plate cooling applications in this study. The local heat flux measurements are introduced by a novel experimental technique that has a function of high-temperature heat flux gauge in which test block assemblies are used to measure the heat flux distribution during water jet impingement cooling. The experiments are performed at fixed flow rate and fixed nozzle-to-target spacing. The results show that effects of cooling water temperature on the characteristics of jet impingement heat transfer are presented for five different water temperatures ranged from 5 to $45^{\circ}C$. The local heat flux curves and heat transfer coefficients are also provided with respect to different boiling regimes.

고온 재료 테스트를 위한 고속 산소 연료 토치 흐름에서의 열유속 측정 (Heat Flux Measurements in High Velocity Oxygen-Fuel Torch Flow for Testing High Thermal Materials)

  • ;최성만;홍성민
    • 한국추진공학회지
    • /
    • 제25권2호
    • /
    • pp.34-41
    • /
    • 2021
  • 상업용 HVOF 토치 (원래 코팅 용도로 설계됨)가 재료 테스트를 위한 고온 유동원으로 수정되었다. 이 연구에서는 수냉식 상용 Gardon 게이지를 사용하여 노즐 출구에서 떨어진 네 위치에서 열유속을 측정하였다. 냉각수 온도 데이터는 동일한 위치에서 열량 측정 열유속 (calorimetric heat flux)을 계산하는 데 사용되었다. 두 방법의 열유속을 비교 한 결과 열량 측정 열유속이 Gardon 게이지 열유속보다 몇 배 더 높은 것으로 나타났다. 두 가지 방법 사이에 나타나는 불일치를 이해하기 위해 열량 측정방법에 대한 가설을 적용하였다. 이것은 가설에 대한 상당한 검증으로 간주될 수 있지만 적절한 수치모델을 사용하여 추가 개선이 필요하다.

판열유속계를 이용한 복사열유속 측정 실험 (Measurement of Radiative Heat Flux Using Plate Thermometer)

  • 박원희;윤경범
    • 대한기계학회논문집B
    • /
    • 제37권1호
    • /
    • pp.95-98
    • /
    • 2013
  • 고온환경의 복사열유속을 측정하기 위하여 판열유속계를 제작하였다. 열유속은 스테인리스스틸로 제작된 판열속계 뒷면에서 측정된 온도를 이용하여 계산되어진다. 스미트-볼터계로 측정된 열유속을 기준값으로 이용하였다. 단열재로의 열손실, 판열유속계 두께방향으로의 전도 및 평판의 위치에 따라 온도가 상이하여 발생하는 열전달 등을 고려하기 위하여 전도상수를 도입하였다. 이 상수는 반발입자군집 최적화법으로 구하였다.