• 제목/요약/키워드: geometric mean attenuation correction

검색결과 4건 처리시간 0.017초

위 배출시간 측정의 분석방범에 대한 연구 (Analysis in Measurements of Gastric Emptying Time)

  • 이춘호;이만구
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제20권1호
    • /
    • pp.35-38
    • /
    • 1997
  • Scintigraphic measurement of gastric emptying time has been reported to be influenced by the variation in depth of radionuclide within the stomach. This study was designed to clarify whether a part of the variability in gastric emptying could be ascribed to a relationship between anterior image, the total anteroposterior Image and the tissue attenuation correction(geometric mean). A dual-head scintillation camera(ADAC, USA) was used to investigate effect of such changes. We were performed 16 normal subject gastric emptying studies with $^{99m}TC$ labelled scramble egg, milk and solid meal(610 Kcal, 300 g) The results are as follows; On anterior Image, $T_{1/2}$ emptying time was delayed by 5 min, 6.5%(range $3{\sim}18\;min,\;5{\sim}31.4%$) compared with the geometric mean. But there was no different gastric emptying time between the total anteroposterior image and geometric mean. Therefore, if will be useful to use the method of geometric mean or the total anteroposterior image to evaluate the gastric emptying time accurately.

  • PDF

신장 방사선 섭취량 결정을 위한 Conjugate View 방법에 대한 평가 (Evaluation of a Conjugate View Method for Determination of Kidney Uptake)

  • 봉정균;윤미진;이종두;김희중;손혜경;권윤영;박해정;김유선
    • 대한핵의학회지
    • /
    • 제39권3호
    • /
    • pp.191-199
    • /
    • 2005
  • 목적: 본 연구의 목적은 신장의 정확한 방사능 섭취량을 얻기 위해서 기하학적 평균 감쇠보정을 이용한 conjugate view 방법(CVM)을 평가하고 Gate 방법과 비교하는 것이다. 대상 및 방법 : 본 연구는 신장의 방사능 섭취량을 시뮬레이션하기 위해서 몬테칼로 코드, SIMIND와 Zubal 팬텀을 사용하였다. 또한 이중 감마카메라를 이용하여 직경 5cm인 팬텀들을 직경 20cm인 실제 팬텀에 삽입하여 실험을 하였다. CVM 방법을 평가하기 위해서 산란과 감쇠가 없는 이상적 데이터와 비교되었다. 또한, CVM 방법을 Gate 방법과 비교하였고, 산란보정의 적용 또는 비적용으로 나누어 CVM 방법을 실행하였다. Gate 방법은 임상에서 사용하는 것처럼 산란보정을 적용하지 않았으며, $0.12cm^{-1}$$0.15cm^{-1}$ 감쇠계수들을 적용하였다. 관심영역 내에 있는 평균계수, 신장영상 위에서 얻어진 프로파일, 선형회귀분석을 이용하여 데이터를 분석하였고, 이상적 데이터와의 상관계수를 계산하였다. 결과: 컴퓨터 시뮬레이션의 경우, 이상적 데이터, CVM 방법, Gate 방법으로부터 얻어진 평균계수들은 각각 (오른쪽: $998{\pm}209$, 왼쪽: $896{\pm}249$), (오른쪽: $911{\pm}207$, 왼쪽: $815{\pm}265$), (오른쪽: $1065{\pm}267$, 왼쪽 $1546{\pm}267$)이었다. CVM 방법은 이상적 데이터와 좋은 상관관계를 보여주었고, 이상적 데이터와 대한 CVM 방법, Gate방법의 상관계수는 각각 (오른쪽: 0.91, 왼쪽: 0.93)와 (오른쪽: 0.85, 왼쪽 0.90)이었다. 결론: 기하학적 평균 감쇠보정을 이용한 CVM 방법은 Gate 방법보다 정량적으로 더 정확한 값을 제공하였다. 결론적으로, 신장 깊이에 영향을 받지 않는 CVM 방법으로 더욱 정확하게 신장의 방사능 섭취량을 측정할 수 있을 것으로 생각된다.

위배출시간 측정시 감쇠보정에 의한 효과 (Attenuation Correction in Measurements of Gastric Emptying Time)

  • 고은미;김덕윤;김병호;장영운;김광원;장린;최영길;김용봉
    • 대한핵의학회지
    • /
    • 제24권2호
    • /
    • pp.260-266
    • /
    • 1990
  • Scintigraphic measurement of gastric emptying time has been reported to be influenced by the variation in depth of radionuclide within the stomach. In order to determine the effect of tissue attenuation in the measurement of gastric emptying time, 15 gastric emptying studies were performed with Tc-99m labeled egg sandwiches. Single anterior detector method overestimated the T1/2 by an average of 13% than geometric mean method and range of overestimation was wide (from -13% to 132%). Therefore, to evaluate the gastric empthying time accurately, methods of attenuation correction are needed.

  • PDF

99mTc-MAG3를 이용한 상대적 신장 기능 평가 시 신장 깊이에 따른 기하평균 적용의 유용성 평가 (Utility Evaluation on Application of Geometric Mean Depending on Depth of Kidney in Split Renal Function Test Using 99mTc-MAG3)

  • 이은별;이왕희;안성민
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제39권2호
    • /
    • pp.199-208
    • /
    • 2016
  • $^{99}mTc-MAG_3$ 동적 신장 검사는 $^{99}mTc-MAG_3$를 이용하여 방사성 의약품이 신장에 섭취되고 배설되는 과정을 멈춤 없이 동적으로 영상화 하는 방법이다. 검사 시작 후 1~2.5분 사이 좌우 신장에서의 비율을 측정하여 상대적 신장기능비를 측정할 수 있으며, 총 신장 기능을 기준으로 상대적 신장기능비를 비율로 나타낼 수 있다. 본 연구는 일반적인 신장 검사 방법인 후면 검출기만 이용해 계수된 데이터로 구해진 상대적 신장기능비와 전면, 후면 검출기 모두로 계수한 뒤 기하평균을 이용해 얻어진 데이터로 구해진 상대적 신장기능비를 비교하여, 좌우 신장 깊이 차이가 상대적 신장기능비에 미치는 영향을 알아보고, 기하평균의 신장 깊이차에 따른 감쇠 보정 효과의 유용성에 대해 연구하고자 한다. 2015년 7월부터 2016년 2월 사이에 $^{99}mTc-MAG_3$ 동적 신장 검사를 시행한 33명의 환자(남자 13명, 여자 20명, 평균나이 44.66세, 범위 5세~70세, 평균키 160.40 cm, 평균 몸무게 55.40 kg)를 대상으로 하였다. 신장 깊이를 측정한 결과. 왼쪽 신장은 평균 65.82 mm, 오른쪽 신장은 71.62 mm로 나타났다. 33명의 환자들 중 30명의 환자에게서 전면과 후면 검출기로 계수하여 기하평균을 적용한 상대적 신장기능비가 후면 검출기로만 얻어진 정보를 가지고 구해진 상대적 신장기능비보다 깊게 위치한 신장의 비가 상승하고, 얕게 위치하고 있는 신장의 비가 떨어지는 결과가 나왔다. 깊게 위치하고 있는 신장과 검출기 사이의 감쇠가 보정된 결과라고 사료되며, 신장 혹은 신장 주변에 병변이 있는 경우, 척추의 기형, 이소 신장의 경우, 이식신장을 가지고 있는 환자 중 본래의 신장을 제거하지 않은 환자의 경우 등 좌우 신장의 깊이 차이가 나는 경우 깊게 위치하고 있는 신장의 비를 보정해 주어 기존에 일반적으로 시행되어지고 있는 검사 방법(후면 검출기 계수)과 비교했을 때, 정확한 상대적 신장기능비를 계산할 수 있을 것이라 사료된다.