• Title/Summary/Keyword: XKnife computer

Search Result 1, Processing Time 0.015 seconds

Verification of stereotactic target point and CT image transfer (정위적 target point 및 CT 영상전환 입증)

  • 유명진
    • Progress in Medical Physics
    • /
    • v.10 no.1
    • /
    • pp.47-54
    • /
    • 1999
  • Purpose: To verify the BRW coordinates of target located within the limit of XKnife hardware, and to verify the successful transfer of image data, rod detection, anatomical structure when CT images are transferred into a XKnife computer. Materials and Methods: Target coordinates of 13 patients were calculated by SCS1 computer through the rod image on the console screen and film. BRW coordinates of target and landmark calculated by SCS1 computer were compared to those acquired by XKnife localizer. Results : Vertical components of BRW coordinates of target for 13 patients are larger than -50 mm, and then the vertical components of BRW coordinates of target are localized within the limit of XKnife hardware. Average differences between XKnife and SCS1 for BRW coordinates of target and landmark were within 1 mm for AP and LAT components, 0.5 mm for VERT component. Conclusion : It was verified that the SCS1 computer is adequate tool to calculate BRW coordinates of target quickly. And by the comparison between SCS1 computer and XKnife localizer, it was verified that the image transfer into the XKnife computer was performed successfully.

  • PDF