• Title/Summary/Keyword: X-ray unit

Search Result 562, Processing Time 0.028 seconds

A Study on Measurement of Output Dose in X-ray Unit (X-선장치에서 출력선량 측정에 관한 연구)

  • Kim, Jong Eon;Lee, Sang Hun
    • Journal of the Korean Society of Radiology
    • /
    • v.14 no.3
    • /
    • pp.289-294
    • /
    • 2020
  • In order to control the quality of X-ray images and patient exposure, it is necessary to document the output dose(air absorption dose(mGy)) output from the X-ray unit from the measurement. The purpose of this study is to find an equation that can calculate the output dose from the measurement of the output dose and output factor(Of) of the X-ray Unit. The output dose and output factors of the X-beam irradiated from the X-ray unit were measured using an XR multi-detector. The output dose calculation formula was obtained by fitting the measured output dose divided by the tube current-exposure time product(mAs) and the set tube voltage with Allometric1. The final output dose calculation formula was obtained by multiplying this formula with the output factor. It is considered that the obtained final output dose calculation formula will be useful for all tube voltages, tube currents, exposure times, field sizes, and distances.

Comparison of X-ray Shielding Performance according to the Weight of unit volume of Heavy Weight Concrete Utilizing Electric Arc Furnace Oxidizing Slag. (전기로 산화슬래그 골재를 활용한 중량 콘크리트의 단위 용적 중량 변화에 따른 X-선 차폐 성능 비교)

  • Lim, Hee Seob;Lee, Han Seung;Choi, jae Seok
    • Proceedings of the Korean Institute of Building Construction Conference
    • /
    • 2013.05a
    • /
    • pp.35-36
    • /
    • 2013
  • Electric arc furnace oxidizing slag from massively produced steel slag has been used in road bases and subbases, hot mix asphalt, and landfill. Electric arc furnace oxidizing slag contains iron (15%~30%) and has a high density of 3.0~3.7 ton/m3. Depending on the type and amount of concrete aggregates, the radiation-shielding characteristics can vary. Therefore, aggregates of electric arc furnace oxidizing slag can be considered for the production of radiation-shielding concrete. The experimental design of this study is experiments on Compressive strength experiments, X-ray irradiation experiments, and experiments related to the unit volume weight were carried out on hardened concrete. This experiment compared the performance evaluation of radiation shielding of concrete using electric arc furnace oxidizing slag.

  • PDF

Measurement of Skin Dose Distribution for the Mobile X-ray Unit Collimator Shielding Device (이동형 X선 장치 차폐도구 제작을 통한 표면선량 분포 측정)

  • Hong, Sun-Suk;Kim, Deuk-Yong
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.12 no.1
    • /
    • pp.5-8
    • /
    • 2010
  • Opened a court in February 10, 2006, a rule of safety management of the diagnosis radiation system was promulgated for safety of the radiation worker, patients and patients' family members. The purpose of this rule is to minimize the risk of being exposed to radiation during the process of handling X-ray. For this reason, we manufactured shielding device of mobile X-ray unit collimator for diminution of skin dose. Shielding device is made to a thickness of Pb 0.375mm. For portable chest radiography, we measured skin dose 50cm from center ray to 200cm at intervals of 20cm by Unfors Xi detector. As a result, a rule of safety management of the diagnosis radiation system has been strengthened. But there are exceptions, such as ER, OR, ICU to this rule. So shielding device could contribute to protect unnecessary radiation exposure and improve nation's health.

  • PDF

Usefulness of Non-Invasive Measurement Tool on Performance Evaluation of Inverter Type X-ray Unit (인버터식 X선장치의 성능평가 시 비접속형 측정기의 유용성)

  • Kang, Se-Sik;Kim, Chang-Soo;Ko, Sung-Jin
    • Journal of radiological science and technology
    • /
    • v.31 no.2
    • /
    • pp.123-127
    • /
    • 2008
  • Purpose: As the demand of a simple and precise method increases to evaluate the performance of the inverter type x-ray unit, we evaluated the usefulness of the recently-introduced X-ray Multi-Function Test Device (moldel : Xi (unfors)-prestige). Method: We compared the performance of X-ray Multi-Function Test Device (XMFTD) which is non-inveasive type device with the performance of Dynalyzer III that has been most widely used inveasive type measure device. Result: X-ray output dose was increased a little in the XMFTD, but both devices were below the performance evaluation standard, 0.002 in the output reproducibility. Linearity of XMFTD were below 0.1 which means that Dynalyzer III showed more excellency in linearity. As the the accuracy of exposure factor, 1.8 and 2 tube voltage, 2.01 and 2.3 tube current were measured. The exposure time was also measured by 0.01 sec ${\pm}10%$. Both devices were within the acceptance of performance evaluatioin standard. Conclusion: We proved the usefulness of X-ray Multi-Function Test Device (model: Xi (unfors)-prestige) to evaluated the performance on reproductibility and linearity of X-ray output and accuracy of exposure factor of inverter type unit.

  • PDF

Development of PC based High Voltage Generator for Dental CT (PC기반 치과 CT용 고전압 펄스 발생장치 개발)

  • Kim, Hack-Seong;Oh, Jun-Yong;Song, Sang-Hoon;Won, Choong-Yeon
    • Proceedings of the KIPE Conference
    • /
    • 2008.06a
    • /
    • pp.580-582
    • /
    • 2008
  • The object of this paper is develope the PC based controlled high voltage power supply and studies 1.2kW(120kV, 10mA) pulse power X-ray generator possible to adapt fluoroscopy of Dental CT X-ray generator and industrial X-ray pulse power equipment. The developed pulse power X-ray generator consisted of mono-block tank include X-ray tube and high voltage X-ray power supply circuit and high voltage control unit with RS232C/422 communication port. The PC control program of pulse power X-ray generator uses LabVIEW, and the size of high voltage transformer and high voltage generator is minimized by high voltage high frequency inverter has 100kHz switching frequency. Also this paper shows result of X-ray tube voltage and tube current correspond to variable load.

  • PDF

A Study on PAE of Tube Voltage of Diagnostic X-ray Units -At the area of Sungnam, Inchon, Seoul- (진단용(診斷用) X선장치(線裝置)의 관전압(管電壓) 백분율(百分率) 평균오차(平均誤差)에 관한 조사(調査) 연구(硏究))

  • Lim, Han-Young;Oh, Hyun-Joo;Kim, Young-Il
    • Journal of radiological science and technology
    • /
    • v.16 no.2
    • /
    • pp.45-50
    • /
    • 1993
  • In this paper, the PAE of tube voltage of 93 diagnostic X-ray units those are using at the 19 general hospitals and 9 local clinics in the area of Sungnam, Inchun, and Seoul have been surveyed. The results are summarized as following. 1. PAE of tube Voltage 70, 100KV at the 56 (60%) X-ray units were ${\pm}7%$ below, at the 8 units (9%), were ${\pm}7%$ above, and those 29 X-ray units (31%) were ${\pm}7%$ below or above with the 70, 100KV tube Voltage in a unit. 2. PAE ${\pm}7%$ below were 68 X-ray units, at the low tube voltage (70KV), and 76 X-ray units at the high tube Voltage (100KV). 3. For six year-use abofe, used X-ray units were PAE ${\pm}7%$ above, and so, the longer used years the more X-ray units became declined in those performance. 4. PAE of tube Voltage were better efficiency at the large capacity 30 X-ray units than small capacity 10. 5. PAE ${\pm}7%$ abovein were only 8 units among 84 diagnostic X-ray units in the general hospitals.

  • PDF

Chromium(III) Complex Obtained from Dipicolinic Acid: Synthesis, Characterization, X-Ray Crystal Structure and Electrochemical Studies

  • Ghasemi, Khaled;Rezvani, Ali Reza;Razak, Ibrahim Abdul;Moghimi, Abolghasem;Ghasemi, Fatemeh;Rosli, Mohd Mustaqim
    • Bulletin of the Korean Chemical Society
    • /
    • v.34 no.10
    • /
    • pp.3093-3097
    • /
    • 2013
  • The synthesis, X-ray crystallography, spectroscopic (IR, UV-vis), and electrochemical properties of the title compound, $[H_3O][Cr(dipic)_2][H_3O^+.Cl^-]$ (1), ($H_2dipic$ = 2,6-pyridinedicarboxylic acid), are reported. This complex crystallizes in the monoclinic space group Cc with a = 14.9006(10) ${\AA}$, b = 12.2114(8) ${\AA}$, c = 8.6337(6) ${\AA}$, ${\alpha}=90.00^{\circ}$, ${\beta}=92.7460(10)^{\circ}$, ${\gamma}=90.00^{\circ}$, and V = 1569.16(18) ${\AA}^3$ with Z = 4. The hydrogen bonding and noncovalent interactions play roles in the stabilization of the structure. In order to gain a better understanding of the most important geometrical parameters in the structure of the complex, atoms in molecules (AIM) method at B3LYP/6-31G level of theory has been employed.

Mechanical properties and radiation shielding performance in concrete with electric arc furnace oxidizing slag aggregate

  • Lim, Hee Seob;Lee, Han Seung;Kwon, Seung Jun
    • Journal of Ceramic Processing Research
    • /
    • v.20 no.4
    • /
    • pp.363-371
    • /
    • 2019
  • In this study, physical properties of normal concrete, magnetite concrete, EAF concrete, and EAF concrete with added iron powder were evaluated and a feasibility of radiation shielding is also evaluated through irradiation tests against X-rays and gamma-rays. While the unit weight of EAF concrete (3.21 t/㎥) appeared lower than that of magnetite concrete (3.50 t/㎥), the results in compressive strength of EAF concrete were greater than those in magnetite and normal concrete. While the radiation transmission rate of normal concrete reaches 26.0% in the X-ray irradiation test, only 6.0% and 9.0% of transmission rate were observed in magnetite concrete and linear relationship with unit volume weight and radiation shielding. In the gamma-ray irradiation test, the performance of EAF and magnetite concretes appeared to be similar. Through the results on the excellent physical properties and radiation shielding performance a potential applicability of EAF concrete to radiation shielding was verified.

A Experimental Study on the Correction Methods of X-ray Exposure Factors (촬영조건의 교정에 관한 실험)

  • Huh, Joon;Kim, Chang-Kyun;Lee, Sun-Sook;Kim, Hak-Seong
    • Journal of radiological science and technology
    • /
    • v.7 no.1
    • /
    • pp.67-70
    • /
    • 1984
  • Radiographic units have changeable factors in x-ray outputs, Therefore, an exposure factor by basic experimental study must be made out about each x-ray installation, but this is very intricate methods and not a practical business. Authors tried out a new method to make simple exposure factors. The first, we had a experimental object radiograph taken to find the output of each radiographic unit. The second, by obtained x-ray density we found the difference in x-ray output. and lastly, we made a new and simple correction method to use the obtained output and x-ray density.

  • PDF

THE CHARACTERISTICS OF OUTPUT FOR INVERTER TYPE X-RAY GENERATOR

  • Lim, Hong-Woo;Han, Euam-Yong;Baek, Hyung-Lae;Lee, Seong-Kil
    • Proceedings of the KIPE Conference
    • /
    • 1998.10a
    • /
    • pp.431-435
    • /
    • 1998
  • This paper deals with the output characteristics of resonant PWM inverter type X-ray generators connected to different DC power units i.e. a single phase full bridge rectifier, a three phase full bridge rectifier and a power storage unit(PSU). The quality of X-ray beam depend on the pulsation waveforms of DC voltage supplied to the X-ray tulbe. In a X-ray generator, the waveforme of DC output voltage can be affected from hramonic distortion of DC input power. When a tube voltage waveform is distorded, the property of X-ray beam such as reproducibility, direcibility and doesage can be reduced. Therefore, we compared DC output waveforms and doesages with three thpe of DC power units and show the experimental results in this paper

  • PDF