• Title/Summary/Keyword: Termoluminescence dosimeter

Search Result 1, Processing Time 0.014 seconds

Evaluation of Depth Dose and Surface Dose According to Treatment Room Wall Distance (방사선 치료실 벽면 거리에 따른 심부선량과 표층선량 평가)

  • Je, Jae-Yong
    • Journal of the Korean Society of Radiology
    • /
    • v.5 no.3
    • /
    • pp.121-125
    • /
    • 2011
  • This study was intended to evaluate the surface dose and depth dose of according to the distance of the treatment room wall. High energy photon beams from linear accelerators produce large scattered radiation by various components of the treatment head, collimator and walls or objects in the treatment room including the patient. The scattered radiation measured by thermoluminescence dosimeter(TLD). Linear accelerators rotation center of the four walls(X) distance was measured to be 236, 272, 303, and 337 cm. The result of 100 cGy and 200 cGy of 6 MV photon irradiation, surface dose was 0.49, 0.83 mSv at 236 cm of the shortest distance to the wall, In 272 cm 0.41, 0.53 mSv, 303 cm in the 0.28, 0.57 mSv, and 337 cm distance from the wall in the 0.33, 0.76 mSv surface dose respectively. There was remarkable difference in the surface dose among the treatment room wall distance. The results of useful data in relation to stochastic effect for radiation therapy patients.