The Study on the Head and Neck Phantom for Quality Assurance of Intensity Modulated Radiotherapy (세기변조방사선치료의 정도관리를 위한 두경부 팬톰 제작에 관한 연구)

  • Shin Dongho;Park Sung-Yong;Kim Joo Young;Lee Se Byeong;Cho Jung Keun;Kim Dae Yong;Cho Kwan Ho
    • Progress in Medical Physics
    • /
    • v.16 no.1
    • /
    • pp.39-46
    • /
    • 2005
  • For the QA of IMRT treatment of head and neck cancer by using M3 (BrainLAB Inc. Germany), it is not easy to measure delivery dose exactly because the dose attenuation appears by the couch according to the position of table and gantry. In order to solve this problem, we fabricated head and neck phantom which would be implemented on the couch mount of Brain Lab Inc. We investigated dose attenuation by the couch and found the difference of dose distribution by the couch, in the applying this phantom to the clinic. After measurement, we found that point dose attenuation was 35% at maximum and dose difference was 5.4% for a point dose measurement of actual patient quality assurance plan.

  • PDF

The effective quality assurance for image guided device using the AMC G-Box (AMC G-Box를 이용한 영상유도장치의 효율적인 정도관리)

  • Kim, Chong Mi
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.2
    • /
    • pp.199-206
    • /
    • 2014
  • Purpose : According to the rapid increase recently in image-guided radiation therapy, It is necessary to control of the image guidance system completely. In particular for the main subject to the accuracy of image guided radiation therapy device to be done essentially the quality assurance. We made efficient phantom in AMC for the management of the accurate and efficient. Materials and Methods : By setting up of five very important as a quality assurance inventory of the Image guidance system, we made (AMC G-Box) phantom for quality assurance efficient and accurate. Quality assurance list were the Iso-center align, the real measurement, the center align of four direction, the accuracy of table movement and the reproducibility of Hounsfield Unit. The rectangular phantom; acrylic with a thickness of 1 cm to $10cm{\time}10cm{\time}10cm$ was inserted the three materials with different densities respectively for measure the CBCT HU. The phantom was to perform a check of consistency centered by creating a marker that indicates the position of the center fixed. By performing the quality assurance using the phantom of existing, comparing the resulting value to the different resulting value using the AMC G-Box, experiment was analyzed time and problems. Therapy equipment was used Varian device. It was measured twice at 1-week intervals. Results : When implemented quality assurance of an image guidance system using AMC G-Box and a phantom existing has been completed, the quality assurance result is similar in $0.2mm{\pm}0.1$. In the case of the conventional method, it was 45 minutes at 30 minutes. When using AMC G-Box, it takes 20 minutes 15 minutes, and declined to 50% of the time. Conclusion : The consistency and accurate of image guidance system tend to decline using device. Therefore, We need to perform thoroughly on the quality assurance related. It needs to be checked daily to consistency check especially. When using the AMC G-Box, It is possible to enhance the accuracy of the patient care and equipment efficiently performing accurate quality assurance.

A Study on Mechanical Errors in Cone Beam Computed Tomography(CBCT) System (콘빔 전산화단층촬영(CBCT) 시스템에서 기계적 오류에 관한 연구)

  • Lee, Yi-Seong;Yoo, Eun-Jeong;Kim, Seung-Keun;Choi, Kyoung-Sik;Lee, Jeong-Woo;Suh, Tae-Suk;Kim, Joeng-Koo
    • Journal of radiological science and technology
    • /
    • v.36 no.2
    • /
    • pp.123-129
    • /
    • 2013
  • This study investigated the rate of setup variance by the rotating unbalance of gantry in image-guided radiation therapy. The equipments used linear accelerator(Elekta Synergy TM, UK) and a three-dimensional volume imaging mode(3D Volume View) in cone beam computed tomography(CBCT) system. 2D images obtained by rotating $360^{\circ}$and $180^{\circ}$ were reconstructed to 3D image. Catpan503 phantom and homogeneous phantom were used to measure the setup errors. Ball-bearing phantom was used to check the rotation axis of the CBCT. The volume image from CBCT using Catphan503 phantom and homogeneous phantom were analyzed and compared to images from conventional CT in the six dimensional view(X, Y, Z, Roll, Pitch, and Yaw). The variance ratio of setup error were difference in X 0.6 mm, Y 0.5 mm Z 0.5 mm when the gantry rotated $360^{\circ}$ in orthogonal coordinate. whereas rotated $180^{\circ}$, the error measured 0.9 mm, 0.2 mm, 0.3 mm in X, Y, Z respectively. In the rotating coordinates, the more increased the rotating unbalance, the more raised average ratio of setup errors. The resolution of CBCT images showed 2 level of difference in the table recommended. CBCT had a good agreement compared to each recommended values which is the mechanical safety, geometry accuracy and image quality. The rotating unbalance of gentry vary hardly in orthogonal coordinate. However, in rotating coordinate of gantry exceeded the ${\pm}1^{\circ}$ of recommended value. Therefore, when we do sophisticated radiation therapy six dimensional correction is needed.