• Title/Summary/Keyword: Stepping inflation

Search Result 1, Processing Time 0.013 seconds

Design of the Blood Pressure Measurement System Using the Inflatable Oscillometric Method (가압식 오실로메트릭 방법을 사용한 혈압측정 시스템의 설계)

  • 노동곤;이윤선;지정호;박성빈;이계형;김해관
    • Journal of Biomedical Engineering Research
    • /
    • v.24 no.4
    • /
    • pp.281-286
    • /
    • 2003
  • Blood Pressure is one of the most fundamental Parameters which reflects physical conditions medically and the blood pressure measurement system using oscillometric method is a Non-Invasive Blood Pressure measurement device by measuring arterial Pressure through a cuff. In this paper. we designed a inflatable wrist blood pressure system which measures blood Pressure during the stepping inflation in the wrist cuff. The hardware system consists of a main power unit, a bladder in cuff unit, signal detection units, signal Processing units. a wireless data transmission unit, and a data display unit. We evaluated the reliability of this system by comparing and analyzing systolic. diastolic blood Pressure, and heart rate with other commercial blood Pressure measurement devices. Characteristic ratio values used to determine systolic and diastolic blood Pressure using MAA(Maximum Amplitude Algorithm) were 0.436 and 0.671 respectively.