• 제목/요약/키워드: Palmaz-Schatz Stent

검색결과 3건 처리시간 0.026초

하중 방향에 따른 관상동맥용 스텐트의 유연성 해석 (Flexibile Analysis of Coronary Stents due to Loading Directions)

  • 조승관;조은정;김한성
    • 한국정밀공학회:학술대회논문집
    • /
    • 한국정밀공학회 2004년도 추계학술대회 논문집
    • /
    • pp.331-334
    • /
    • 2004
  • In clinical use, coronary stents keep coronary arteries open after expansion with a balloon catheter and prevent the expanded artery from collapsing. Coronary stents are positioned in artery by catheter with a balloon along a guide wire to the lesion site. Flexibility is one of important ability for delivery. In this paper, Palmaz-Schatz stent and Tenax complete stent were selected because these are the most representative of tubular stents. Finite element analyses for the stent system were performed using ABAQUS/Standard code. The present study estimated the flexibility of coronary stents due to loading directions. Moreover the present paper suggests a numerical method to test the flexibility of stents. In conclusion this paper shows how the finite element analysis can be effectively organized in the stent development.

  • PDF

유한요소법을 이용한 스텐트의 기계적 성능 비교 및 평가 (Using Finite Element Analysis for Mechanical Properties of Coronary Stents)

  • 조승관;김한성;이상헌;탁계래;탁승제
    • 한국정밀공학회:학술대회논문집
    • /
    • 한국정밀공학회 2003년도 춘계학술대회 논문집
    • /
    • pp.1306-1309
    • /
    • 2003
  • In the present paper, finite element analyses were carried out to evaluate the most important feature of a stent, that is. high radial strength and flexibility. Palmaz-Schatz 154 stent and two new models(stent A, stent B) were selected because they are the most representatives of tubular stents. Finite element analyses for the stent system were performed using ABAQUS/Explicit code. As a result, Palmaz-Schatz 154 stent shows sufficient radial strength but it needs some improvement in general properties such as high flexibility, low elastic recoil, low longitudinal contraction and low metal coverage area. Other two models show that sufficient flexibility, foreshortening and longitudinal recoil.

  • PDF

유한요소 해석을 이용한 스텐트 최적형상 설계 (A Study on Optimal Shape of Stent by Finite Element Analysis)

  • 이태현;양철호
    • 한국산학기술학회논문지
    • /
    • 제18권11호
    • /
    • pp.1-6
    • /
    • 2017
  • 스텐트는 인체의 좁아지거나 막힌 부위에 삽입되어 혈류의 흐름을 정상화 시키는데 사용되는 금속망 형태의 임플란트로 관상 동맥 질환을 치료하는데 가장 일반적인 방법으로 널리 사용된다. 본 논문에서는 유한요소 해석과 다꾸지 방법을 이용하여 설정한 스텐트 설계 인자의 변화에 따른 기계적 반응을 고찰하였다. 스텐트 모델은 팔마즈 스텐트를 사용하였고 스텐트의 재료 모델로서는 탄소성 모델, 풍선은 초탄성 모델을 사용하였다. 연구의주요 관심은 스텐트의 탄성 회복량 조절을 통한 혈관의 재 협착 문제를 감소시킬 수 있는 방안에 대한 설계 인자의 영향을 고찰하는데 있다. 스텐트 두께, 슬롯의 가로 길이, 슬롯의 세로 길이의 각도를 설계인자로 선택하여 직교배열표를 구성하였다. 유한요소 해석을 이용하여 혈관 내 스텐트의 반경 방향 탄성회복율과 길이 방향 탄성회복율을 계산하였고 다구찌 기법을 이용한 통계적 분석을 하여 재 협착의 가능성을 감소시킬 수 있는 스텐트의 최적 형상 설계 방향을 제시하였다. 최적형상은 기본 모델에 비하여 탄성회복율은 반경방향으로 약 1%, 길이방향으로 약 0.1% 감소함을 보였다.