A Micro Cell Counter Integrated with Oxygen Micropump

  • Son, Sang-Uk;Choi, Yo-Han;Lee, Seung-S.
    • 제어로봇시스템학회:학술대회논문집
    • /
    • 2003.10a
    • /
    • pp.2441-2444
    • /
    • 2003
  • This paper describes fabrication of a micro cell counter integrated with an oxygen micropump and Sephadex G-25 beads counting experiment. The device utilized a phototransistor, microwindow, and light source of microscope for beads detection. Microheater and microchannel were used for pumping and guiding of beads to the microwindow. Counting capability of the device was tested with a peristaltic pump and the measured signals (${\sim}10\;mV$) with oscilloscope showed peak shape when beads passed the microwindow. Pumping of beads by the oxygen micropump was carried out by heating paraffin, which enveloped manganese dioxide (catalyst), to trigger the decomposition of hydrogen peroxide into water and oxygen. It lasted for 5 min with $7\;{\mu}l$ of wt. 30 % hydrogen peroxide. Beads counting by oxygen micropump showed peaks ($2{\sim}20\;mV$) with $30\;{\mu}l$ of beads sample and the number of peaks by magnitude was acquired.

  • PDF

Micro Cell Counter Integrated with An Oxygen Micropump (산소 미세 펌프가 내장된 미세 세포 계수기)

  • Son, Sang-Uk;Choi, Yo-Han;Lee, Seung-Seob
    • Transactions of the Korean Society of Mechanical Engineers A
    • /
    • v.28 no.8 s.227
    • /
    • pp.1159-1165
    • /
    • 2004
  • This paper describes fabrication of a micro cell counter integrated with an oxygen micropump and counting experiment with Sephadex G-25 beads ($70{\sim}100\;{\mu}m$). The pumping part consisted of a microheater, catalyst (manganese dioxide) enveloped with paraffin, hydrogen peroxide, and microchannel, and the counting part consisted of collimated light, a microwindow, and a phototransistor including an external circuit. The micropump generated oxygen gas by decomposing hydrogen peroxide with manganese dioxide, which was initiated by melting the paraffin with the microheater, and pumped beads in the microchannel. When the beads passed the microwindow, they shaded the collimated light and changed the illumination on the phototransistor, which caused the current variation in the circuit. The signals, according to the bead size, reached up to 22 mV with noise level of 2 mV during 50 seconds and the numbers of peaks were analyzed by magnitude.