• Title/Summary/Keyword: Marginal bone resoprtion

Search Result 1, Processing Time 0.014 seconds

Three-dimensional finite element analysis for influence of marginal bone resorption on stress distribution in internal conical joint type implant fixture (변연골 흡수가 내측연결 임플란트 매식체의 응력분포에 미치는 영향)

  • Yun, Mi-Jung;Yoon, Min-Chul;Eom, Tae-Gwan;Huh, Jung-Bo;Jeong, Chang-Mo
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.50 no.2
    • /
    • pp.99-105
    • /
    • 2012
  • Purpose: The change of the marginal bone around dental implants have significance not only for the functional maintenance but also for the esthetic success of the implant. The purpose of this study was to investigate the load transfer of internal conical joint type implant according to marginal bone resorption by using the three-dimensional finite element analysis model. Materials and methods: The internal conical joint type system was selected as an experimental model. Finite element models of bone/implant/prosthesis complex were constructed. A load of 300 N was applied vertically beside 3 mm of implant axis. Results: The pattern of stress distribution according to marginal bone resorption was similar. The maximum equivalent stress of implant was increase according to marginal bone resorption and the largest maximum equivalent stress was shown at model of 1 mm marginal bone resorption. Although marginal bone loss more than 1mm was occurred increasing of stress, the width of the stress increase was decreasing. Conclusion: According to these results, the exposure of thin neck portion of internal conical joint type implant is most important factor in stress increasing.