• Title/Summary/Keyword: MV 광자선

Search Result 163, Processing Time 0.022 seconds

A Study on the Variation of Transmission Factors, Output Factors and Percent Depth Doses by Wedge Filters for 4~10 MV X-Ray Beams (4~10 MV X-선의 쐐기 (wedge) 필터의 투과율과 출력계수, 선축상 선량분포의 변화에 관한 연구)

  • 강위생
    • Progress in Medical Physics
    • /
    • v.8 no.2
    • /
    • pp.3-17
    • /
    • 1997
  • Because a wedged beam consists of attenuated primary photons and scattered radiations from wedge, the spectrum of the wedged beam does not coincide with that of an open beam with same geometry. The aims of current report are to get exact information about whether effects of 15-60$^{\circ}$ wedge for 4 -10 MV photon beams should be considered for dose calculation or not, and to suggest a reference condition for measurement of wedge transmission factor. Percent depth dose of both open and wedged fields with angles of 15, 30, 45, 60$^{\circ}$ for beams of 4 MV(Clinac 4/100, Varian), two 6 MV(Clinac 6/100 and Clinac 2100C, Varian), 10 MV(Clinac 2100C, Varian) X-rays were measured to 30cm deep in water using ionization chambers. Hardening factors of photon beams were calculated with measured PDDs. Both field size factors and transmission factors of wedge filters were measured at d$_{max}$ in water. Beam hardening factors of wedged fields of 4 and 6 MV X-ray were larger than 1 for all wedge angles, field sizes and depths deeper than d$_{max}$ Beam hardening factors for wedge angles 15, 30, 45, 60$^{\circ}$ for 10$\times$10cm were respectively 1.010, 1.014, 1.023 and 1.034 for 4MV X-ray, 1.005, 1.008, 1.019, and 1.024 for 6MV X-ray of Clinac 6/100, 1.011, 1.021, 1.032, 1.036 for 6MV X-ray of Clinac 2100C, and 1.008, 1.012, 1.012 and 1.012 for 10MV X-ray. Beam hardening factors of 10MV X-ray were 1 within 1.2% difference for all wedge angles, depths and field sizes. It was made clear that for 6MV X-rays, the beam hardening factor depends on treatment machine. The relationship of the factor and depth was linear. Field size factor at d$_{max}$ was independent of wedge angle except for the field of 15$\times$15cm. and maximum difference of the field size factors for the field size was 1.4% for 4MV X-ray. When the wedge factor is determined, dependence of the factor on field size is negligible at d$_{max}$ but should be considered at deeper depth. Calculating dose distribution or MU, the beam hardening factor should be applied for 4~6MV X-ray beams, but might not be considered for 10MV beam. When wedge transmission factor was determined at d$_{max}$ or in air, field size factors for open field are also applicable to wedged fields, but otherwise, field size factor for each wedge or wedge factor depending on field size should be applied.

  • PDF

Comparison and Analysis of Photon Beam Data for Hospitals in Korea and Data for Quality Assurance of Treatment Planning System (국내 의료기관들의 광자 빔 데이터의 비교 분석 및 치료계획 시스템 정도관리자료)

  • Lee, Re-Na;Cho, Byung-Chul;Kang, Sei-Kwon
    • Progress in Medical Physics
    • /
    • v.17 no.3
    • /
    • pp.179-186
    • /
    • 2006
  • Purpose: Photon beam data of linear accelerators in Korea are collected, analyzed, and a simple method for checking and verifying the dose calculations in a TPS are suggested. Materials and Methods: Photon beam data such as output calibration condition, output factor, wedge factor, percent depth dose, beam profile, and beam quality were collected from 26 institutions in Korea. In order to verify the accuracy of dose calculation, ten sample planning tests were peformed. These Include square, elongated, and blocked fields, wedge fields, off-axis dose calculation, SSD variation. The planned data were compared to that of manual calculations. Results: The average and standard deviation of photon beam quality for 6, 10, and 15 MV were $0.576{\pm}0.005,\;0.632{\pm}0.004,\;and\;0.647{\pm}0.006$, respectively. The output factors of 6 MV photon beam measured at depth of dose maximum for $5{\times}5cm,\;15{\times}15cm,\;20{\times}20cm\;were\;0.944{\pm}0.006,\;1.031{\pm}0.006,\;and\;1.055{\pm}0.007$. For 10 MV photon beam, the values were $0.935{\pm}0.006,\;1.031{\pm}0.007,\;1.054{\pm}0.0005$. The collected data were not enough to calculate average, the output factors for 15MV photon beam with field size of $5{\times}5cm,\;15{\times}15cm,\;20{\times}20cm\;were\;0.941{\pm}0.008,\;1.032{\pm}0.004,\;1.049{\pm}0.014$. There was seven institutions $e{\times}ceeding$ tolerance when monitor unit values calculated from treatment planning system and manually were compared. The measured average MU values for the machines calibrated at SAD setup were 3 MU and 5 MU higher than the machines calibrated at SSD for 6 MV and 10 MV, respectively except the wedge case. When the wedges were inserted, the MU values to deliver 100 cGy to 5 cm depends on manufactures. When the same wedge angle was used, Siemens machine requires more MUs then Varian machine. Conclusion: In this study, photon beam data are collected and analyzed to provide a baseline value for chocking beam data and the accuracy of dose calculation for a treatment planning system.

  • PDF

Assessment of the usefulness of the Machine Performance Check system that is an evaluation tools for the determination of daily beam output (일간 빔 출력 확인을 위한 평가도구인 Machine Performance Check의 유용성 평가)

  • Lee, Sang Hyeon;Ahn, Woo Sang;Lee, Woo Seok;Choi, Jin Hyeok;Kim, Seon Yeon
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.29 no.2
    • /
    • pp.65-73
    • /
    • 2017
  • Purpose: Machine Performance Check (MPC) is a self-checking software based on the Electronic Portal Imaging Device (EPID) to measure daily beam outputs without external installation. The purpose of this study is to verify the usefulness of MPC by comparing and correlating daily beam output of QA Beamchecker PLUS. Materials and Methods: Linear accelerator (Truebeam 2.5) was used to measure 10 energies which are composed of photon beams(6, 10, 15 MV and 6, 10 MV-FFF) and electron beams(6, 9, 12, 16 and 20 MeV). A total of 80 cycles of data was obtained by measuring beam output measurement before treatment over five months period. The Pearson correlation coefficient was used to evaluate the consistency of the beam output between the MPC and the QA Beamchecker PLUS. In this study, if the Pearson correlation coefficient is; (1) 0.8 or higher, the correlation is very strong (2) between 0.6 and 0.79, the correlation is strong (3) between 0.4 and 0.59, the correlation is moderate (4) between 0.2 and 0.39, the correlation is weak (5) lower than 0.2, the correlation is very weak. Results: Output variations observed between MPC and QA Beamchecker PLUS were within 2 % for photons and electrons. The beam outputs variations of MPC were $0.29{\pm}0.26%$ and $0.30{\pm}0.26%$ for photon and electron beams, respectively. QA Beamchecker PLUS beam outputs were $0.31{\pm}0.24%$ and $0.33{\pm}0.24%$ for photon and electron beams, respectively. The Pearson correlation coefficient between MPC and QA Beamchecker PLUS indicated that photon beams were very strong at 15 MV, and strong at 6 MV, 10 MV, 6 MV-FFF and 10 MV-FFF. For electron beams, the Pearson correlation coefficient were strong at 16 MeV and 20 MeV, moderate at 9 MeV and 12 MeV, and very weak at 6 MeV. Conclusion: MPC showed significantly strong correlation with QA Beamchecker PLUS when testing with photon beams and high-energy electron beams in the evaluation of daily beam output, but the correlation when testing with low-energy electron beams (6 MeV) appeared to be low. However, MPC and QA Beamchecker PLUS are considered to be suitable for checking daily beam output, as they performed within 2 % of beam output consistency during the observation. MPC which can perform faster than the conventional daily beam output measurement tool, is considered to be an effective method for users.

  • PDF

Effect of Transverse Magnetic Field on Build-up Region of 6 MV Photon Beam (6 MV 광자선의 선량 상승 영역에 대한 자기장 영향)

  • Shin, Seong Soo;Choi, Wonsik;Ahn, Woo Sang;Kwak, Jungwon
    • Journal of the Korean Magnetics Society
    • /
    • v.27 no.1
    • /
    • pp.18-22
    • /
    • 2017
  • The purpose of this study was to present an improved method of dose modulation over the increase of build-up for existing 6 MV photon beam. Two neodymium permanent magnets with a strength of 0.5 T (Tesla) were applied with a magnetic field perpendicular to the photon beam. The effects of dose on build-up region with or without the magnetic field were measured according to the magnet-to-magnet distance (MMS) and the magnet-to-surface distance (MSD). For MMS = 6 cm and MSD = 2.5 cm, $D_{0mm}$, $D_{2mm}$, $D_{5mm}$, and $D_{10mm}$ showed improved doses of 6.8 %, 14.6 %, 6.9 %, and 2.1 %, respectively, as compared with 6 MV open beam. In this study, the device with low strength magnetic field can be applied directly to the outside of the human body when the target volume located close to the skin is delivered with radiation. It is expected that the method of build-up modulation using a low strength magnetic field will be feasible in the clinical applications.

Calculation of Photon Spectra from the Tungsten Target for 10 MeV Electron Beam (10 MeV의 전자선이 텅스텐 표적에 충돌하여 생성되는 광자선 스펙트럼의 계산)

  • 이정옥;정동혁;문성록;강정구;김승곤
    • Progress in Medical Physics
    • /
    • v.10 no.1
    • /
    • pp.55-62
    • /
    • 1999
  • In an effort to study the characteristics of x-rays utilized in radiation therapy, we calculated the energy distribution and the mean energy of x-rays generated from a tungsten target bombarded by 6, 10, and 15 MeV electron beams, using a Monte Carlo technique. The average photon energies calculated as a function of the beam radius lied in 1.4 ∼ 1.6, 2.1 ∼ 2.5 and 2.8 ∼ 3.3 MeV ranges for 4, 10, and 15 MV electron beams, respectively, which turned out to have no strong dependence on the radius. Using the energy distributions of 6,10, and 15 MV x-rays obtained for the target distance of 100 cm, percentage depth doses were determined using Monte Carlo calculations. For the case 10 MV, a comparison was made between our calculation and measurement performed by others. The calculated percentage depth dose appeared somewhat smaller than the measured one except in the surface region. We conclude that this is due to the fact that the beam hardening effect resulting from the flattening filter was not properly allowed for in our Monte Carlo calculations.

  • PDF

Analysis of dose from surface to near the buildup region in the therapeutic X-ray beam (표피로 부터 buildup 영역까지 흡수되는 암치료용 방사선의 선량분석)

  • Vahc, Young-Woo
    • Progress in Medical Physics
    • /
    • v.6 no.2
    • /
    • pp.41-50
    • /
    • 1995
  • The absorbed dose and contaminant electron distribution of therapeutic X-ray beam (15MV photon) was studied with a half blocked beams of 30$\times$30$\textrm{cm}^2$ and field size ranging from 5$\times$5 to 30$\times$30$\textrm{cm}^2$. For a 15MV photon beam energy, the value of the depth of dose maximum, d$_{max}$, gradually decrease with increasing field size from 5$\times$5 to 30$\times$30$\textrm{cm}^2$ due to mainly by contaminant electrons which are produced in the flattening filter and scattered by collimator jaws, tray holder〔Lucite〕, blocking block and air. The results suggest that separate dosimetry data should be kept for blocked and unblocked field. The inherence of the contaminant electrons to the open field depth of maximum dose can lead to mistaken results if attenuation measurements are made at that depth. A nurmerous contaminant electrons mainly were distributed as shape of corn in the central photon beam and their path length in the water were shorter than 30mm because of the electrons energy having around 6MeV. These results clearly appears that the substraction of scattered electrons (electrons and positrons) from the total depth dose curve not only lowers the absolute dose in the bulidup region and surface dose, it also causes a shift of d$_{max}$ to a deeper depth. In the terapeutic high energy photon beam, the absorbed dose near the buildup region is the combined result of incident contaminant electrons and phantom generated electronsrons.

  • PDF

Wedge Factors in Various Field Sizes (조사야의 크기에 따른 쐐기 인자)

  • 조철우
    • Progress in Medical Physics
    • /
    • v.6 no.2
    • /
    • pp.29-34
    • /
    • 1995
  • The wedge factor is defined as a ratio of the absorbed dose in a phantom at a depth of reference point on the central axis with the wedge in the place to the absorbed dose at the same point with the wedge removed. We attempted to show the wedge factors dependence on the field sizes. The wedge factors were measured at various field sizes on 6MV and 15MV x-ray of Varian Clinac 1800 and 5MV x-ray of Philips SL75/5. The single wedge factor measured for a reference field size(10cmx10cm) may not be valid for all field sizes. For the thick wedge, especially an autowedge on Philips SL75/5 for maximum field size width 30cm. the error can be significant(6.6%). Therefore, in the presence of a wedge filter in the beam, a field size dependent wedge factor may be necessary in the treatment dose calculations.

  • PDF

두경부암의 6MV 광자선 치료 시 표면선량 증가를 위한 Spoiler의 유용성 평가

  • 이강혁;김원택;이화중;김대영
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.14 no.1
    • /
    • pp.41-47
    • /
    • 2002
  • 1.목적 두경부암(head and neck Ca)과 쇄골상부(Supraclavicular)에 6MV 광자선으로 치료 시 치료부위(Target volume)가 피부에서 대략 $1^{\sim}2mm$정도 깊이에 위치할 경우, 6MV 광자선의 선량분포는 표면선량이 낮아서 치료에 적합하지 않기 때문에 Bolus와 같이 사용하지만 Skin Sparing(피부보호)효과의 손실로 피부의 손상이 발생할 수 있다. 이러한 이유로 피부의 보호와 치료 시 표면선량의 증가를 위해 Spoiler(산란판)를 제작하여 측정 후 그 특성을 이해하고 선량의 분포를 통하여 Bolus와 비교한 후에 Spoiler의 유용성에 대해 평가하고자 하였다. 2.방법 Siemens사 선형가속기(PRIMUS)의 6MV 광자선을 사용하여 Spoiler의 사용여부 및 Spoiler의 사용 시에는 조사면의 크기를 $5{\times}5,\;7{\times}7,\;10{\times}10,\;15{\times}15,\;20{\times}20cm^2$로 하고 Spoiler와 표면과의 거리는 6, 10, 15cm로 바꾸어 가면서 물팬톰(PTW. 독일)을 이용해 깊이와 측방에 따른 선량분포를 Markus 전리함(PTW. 독일)으로 측정하였으며 전리함의 방수를 위해 씌어진 방수 캡 때문에 표면선량을 별도의 고형 팬톰으로 측정하였다. 표면의 측정선량은 전리함의 측면 벽 등에 의한 선량 측정치의 증가 현상으로 과 반응을 보였으며 이를 교정하였다. 그리고 측정된 데이터를 치료계획 시스템(Pinnacle 6.0m)으로 비교, 분석하였다. 3.결과 Spoiler의 사용 시 3cm깊이 측정선량 백분율과 Spoiler를 사용하지 않은 해당 치료 조사면의 3cm깊이 선량의 백분율에 일치하도록 하여 가상의 치료 깊이인 2mm에서 측정값을 비교하여 본 결과 조사면 $5{\times}5,\;10{\times}10,\;20{\times}20cm^2$에서 OPEN시 62, 64, $70\%$, Bolus는 97, 97, $99\%$로 Spoiler의 사용 시 표면과의 거리가 6cm에서 82, 98, $103\%$, 10cm에는 72, 89, $101\%$, 15m에 65, 79, $96\%$로 나타났으며 표면에서의 측정값을 비교하여 본 결과 OPEN시 11, 17, $27\%$, Bolus는 84, 84, $86\%$, Spoiler의 사용 시 6cm에서 40, 71, $93\%$, 10cm에는 25, 50, $81\%$, 15cm에 18, 36, $67\%$를 나타내었다. 또한 3m깊이에서의 측방 선량분포에서 Spoiler의 거리변화(6, 10cm)는 심부선량의 변화에 영향을 주지 않는 것으로 확인할 수 있었다. 그리고 위의 실험측정치를 치료계획 시스템에 입력하여 선량분포를 확인한 결과 Spoiler를 사용하는 경우 OPEN에 비해 선량분포 영역을 표면으로 끌어 올릴 수 있으며 Bolus 보다 피부 보호효과는 어느 정도 유지가 되는 것을 보여주었다. 4.결론 이와 같이 Spoiler는 Bolus와 비교하여 6MV 광자선의 build up 영역을 표면으로 증가시키는 동시에 Skin Sparing(피부보호)효과를 유지할 수 있으며 두경부암의 치료에서 Spoiler의 사용이 가능한 조건으로는 조사면이 $5{\times}5cm^2$에서 Spoiler와 표면과의 거리가 6cm일 때, $7{\times}7cm^2$에서 6cm, 10cm였고 $10{\times}10cm^2$는 10cm, 15cm로, $15{\times}15cm^2$는 15cm의 간격으로 평가되었다. 또한 $20{\times}20cm^2$의 조사면, Spoiler가 6cm 간격 인 경우 Bolus를 사용한 것 보다 더욱 높은 표면선량을 나타내었다. 그러나 Spoiler와 표면간의 거리를 다르게 함으로써 깊이에 따라 선량분포를 다양하게 나타낼 수 있기 때문에 표면선량의 증가와 피부의 보호를 위해 환자의 피부 두께, 실제 치료 부위의 깊이 등을 고려한다면 Spoiler를 사용하는 것이 bolus를 사용하는 것보다 더 유용하게 적용할 수 있을 것으로 사료된다.

  • PDF

The Effects of Metal Plate loaded on TLD chip in 6 MV Photon and 6 MeV Electron Beams (6 MV 광자선과 6 MeV 전자선 하에서 TLD 기판 위에 얹힌 금속 박막의 효과)

  • Kim, Sookil;Byungnim Min
    • Progress in Medical Physics
    • /
    • v.10 no.1
    • /
    • pp.41-46
    • /
    • 1999
  • There is necessity for making a smaller and more sensitive detector in small field sizes. This report assesses the suitability of metal-loaded thermoluminescent dosimeters for this purpose. Measurements were performed in the 6 MV photon and 6 MeV electron beams of a medical linear accelerator with LiF thermoluminescence dosimeters (TLD-100) embedded in solid water phantom. TLD-100 chips(surface area 3.2 $\times$ 3.2 $\textrm{mm}^2$) loaded with a metal plate(Tin or gold respectively) were used to enhance dose readings to TLD-100. Surface dose was measured for field size 10 $\times$ 10 $\textrm{cm}^2$ and 100 em SSD. Measurements have been made of the enhanced signal intensity and good linearity for absorbed dose with each metal. Using a 1 mm each metal on TLD-l00 in the beam increased the surface dose to 14% and 56% respectively for 6MV photon. In the case of 6 MeV electron, gold plate enhanced the TL response to 13%, but there is no difference for tin plate. The specific dose response of TLD-100 with thin metal plate increases with electron concentration of metal film, this is most likely due to increased electron scattered from the additional material with electron density higher than TLD-100. This emphasizes the role of TL dosimeters with metal as amplified dosimeters for therapeutic high energy x-ray beams. Due to the enhanced dose reading of TLD-100 with metal plate, it could be possible to develop smaller TL dosimeter with high sensitivity.

  • PDF

Protection effect of metal balls against high energy photon beams (고에너지 광자선에 대한 금속구의 차폐효과)

  • 강위생;강석종
    • Progress in Medical Physics
    • /
    • v.9 no.3
    • /
    • pp.137-141
    • /
    • 1998
  • The purposes of this report are to evaluate whether lead ball and steel ball could be used as protective material of radiation and to acquire physical data of them for protecting 4-10 MV X-ray beams. Lead balls of diameter 2.0~2.5mm or steel balls of diameter 1.5~2.0 mm were filled in an acrylic box of uniform width. An MV radiograph of metal balls in a box were taken to ascertain uniformity of ball distribution in the box. Average density of metal ball and linear attenuation coefficient of metal balls for 4~10 MV X -rays were measured. At the time of measurement of linear attenuation coefficient, Farmer ionization chamber was used and to minimize the scatter effect, distance between the ball and the ionization chamber was 70 cm and field size was 5.5cm${\times}$5.5cm. For comparison, same parameters of lead and steel plates were measured. The distribution of metal balls was uniform in the box. The density of a mixture of lead-air was 6.93g/cm$^3$, 0.611 times density of lead, and the density of a mixture of steel-air was 4.75g/cm$^3$, 0.604 times density of steel. Half-value layers of a mixture of lead-air were 1.89 cm for 4 MV X-ray, 2.07 cm for 6 MV X-ray and 2.16 cm for 10 MV X-ray, and approximately 1.64 times of HVL of lead plate. Half-value layers of a mixture of steel-air were 3.24 cm for 4 MV X-ray, 3.70 cm for 6 MV X-ray and 4.15 cm for 10 MV X-ray, and approximately 1.65 times of HVL of lead plate. Metal balls can be used because they could be distributed evenly. Average densities of mixtures of lead-air and steel-air were 6.93g/cm$^3$, 4.75g/cm$^3$ respectively and approximately 1.65 times of densities of lead and steel. Product of density and HVL for a mixture of metal-air are same as the metal.

  • PDF