• 제목/요약/키워드: Korean Artificial Heart

검색결과 282건 처리시간 0.028초

완전 이식형 인공심장의 심박출량 자동 제어 알고리즘 개발에 관한 연구 (Development of an Automatic Cardiac Output Control Algorithm for the Total Artificial Heart)

  • 최원우;김희찬;민병구
    • 전자공학회논문지B
    • /
    • 제32B권3호
    • /
    • pp.38-47
    • /
    • 1995
  • A new automatic cardiac output control algorithm for the motor-driven electromechanical total artificial heart(TAH) was developed based on the motor current waveform analysis without using any extra transducer. The basic control requirements of artificial heart can be described in terms of three features : preload sensitivity, afterload insensitivity, and balanced ventricular outputs. In the previous studies, many transducers were utilized to obtain informations of hemodynamic states for the automatic cardiac output control, But such automatic control systems with sensors have had reliability problems. We proposed a new sensorless automatic cardiac output control algorithm providing adequate cardiac output to the time-varying physiological demand without causing right atrial collapse, which is one of the critical problem in an active-filling type device. In-vitro tests were performed on a mock circulation system to evaluate the performance of the developed algorithm and the results show that the new algorithm satisfied the basic control requirements on the cardiac output response and the possibility of application of the developed algorithm to in vivo experiments.

  • PDF

인공심장의 비선형 혈류 역학 변수 예측에 관한 연구 (A Study on the Practical Estimation of Nonlinear Hemodynamic Variables for the Moving-Actuator type Total Artificial Heart)

  • 엄경식;안재목
    • 대한의용생체공학회:의공학회지
    • /
    • 제19권2호
    • /
    • pp.153-162
    • /
    • 1998
  • 인공 심장의 연구에 있어서 혈류 역학 변수의 추정은 두말할 필요도 없이 매우 중요하다. 그러나 이 중요도에 견주어 볼지라도 실세계에 적용될 수 있는 연구 결과는 희박하다. 본 논문에서는 이동 작동기형 완전 인공 심장을 위한 혈류 역학 변수 추정 문제가 연구되었다. 추정 방법론으로서 다차원 선형 보간 기법을 이용하였다. 모의 순환 장치의 실험으로 제안된 방법을 검증하였으며 좋은 성능을 보였다.

  • PDF

악성 췌장 병변 진단에서 인공지능기술을 이용한 초음파내시경의 응용 (Application of Endoscopic Ultrasound-based Artificial Intelligence in Diagnosis of Pancreatic Malignancies)

  • 안재희;정회훈;박재근
    • Journal of Digestive Cancer Research
    • /
    • 제12권1호
    • /
    • pp.31-37
    • /
    • 2024
  • Pancreatic cancer is a highly fatal malignancy with a 5-year survival rate of < 10%. Endoscopic ultrasound (EUS) is a useful noninvasive tool for differential diagnosis of pancreatic malignancy and treatment decision-making. However, the performance of EUS is suboptimal, and its accuracy for differentiating pancreatic malignancy has increased interest in the application of artificial intelligence (AI). Recent studies have reported that EUS-based AI models can facilitate early and more accurate diagnosis than other preexisting methods. This article provides a review of the literature on EUS-based AI studies of pancreatic malignancies.

개구리 심전도(EKG) 및 혈액상의 계절에 따른 변화 (Changes of the Electrocardiogram and Blood Picture of Frogs in Four Seasons)

  • 이정무;배성호;신현찬;채의업
    • The Korean Journal of Physiology
    • /
    • 제8권2호
    • /
    • pp.33-44
    • /
    • 1974
  • The electrocardiogram of frogs were obtained in winter (January), spring (April), summer (July) and autumn (September and November). Electrocardiograms were recorded applying electrodes to the atria, ventricle and apex of the heart by unipolar or bipolar leads. V wave was recorded prior to P wave, for the presence of the sinus venosus which controls the automaticity of the frog heart, in four seasons. Regardless of the leads or the position of the electrodes P wave was diphasic and wide. According to the rise of temperature the rate of heart beat was increased, and V-P and P-R interval were shortened. Two regression line between R-R interval and both V-P interval and P-R interval were drawn. These were calculated as V-P interval=1 0.276R-R $interva1+0.067{\pm}0.15$ (sec.) and P-R interval=0.179R-R $interva1+0.155{\pm}0.1$ (sec). From these calculation the larger gradient of V-P interval than P-R interval was suggestive that the heart rate is more dependent on the changes of V-P interval than that of P-R interval. Changes of the heart rate were also measured in four seasons and artificial temperatures. Two regression lines between the heart rate (H.H.) and both seasonal temperature (T) and artificial temperature, were drawn. These two lines were calculated as H.R.=20+3.71 (T-10) and H.R.=32+1.425 T respectively. From two gradients of the above equations it is considered that the changes of the heart rate in artificial temperature were milder than that in seasonal temperature. The number of RBC and WBC of frogs were measured in four seasons and a tendency of the changes was observed according to the seasonal variation.

  • PDF

인공 심장판막 하류 유동의 in vitro 연구 (In vitro study of downstream flow passing through heart valve prostheses)

  • 김형범
    • 한국가시화정보학회지
    • /
    • 제4권1호
    • /
    • pp.47-55
    • /
    • 2006
  • The left ventricular filling flow is now considered as an indicator which can be used for early diagnosing of cardiovascular diseases. Because the understanding of left ventricular flow physics is critical for this purpose, the downstream flow characteristics of the artificial heart valve are investigated using particle image velocimetry (PIV) method. In this study, we investigated the wake characteristics of flows passing through three different artificial valves (St.Jude medical bileaflet mechanical valve. Bjork-Shiley monostrut mechanical valve and St.Jude medical Biocor bio valve). The downstream flow field has remarkably altered according to the different valves. SJM MHV has the flow field similar to the pulsating circular jet and BS MHV has oblique pulsating jet. SJM BHV shows the similar flow field of clinical data of normal heart.

  • PDF

지지대 형상에 따른 단엽식 고분자 판막의 비선형 해석 (Nonlinear Analysis of the Monoleaflet Polymer Valve According to Shape of Supporting Members)

  • 한근조;안성찬;심재준;김성윤
    • 한국정밀공학회:학술대회논문집
    • /
    • 한국정밀공학회 2001년도 춘계학술대회 논문집
    • /
    • pp.748-751
    • /
    • 2001
  • Monoleaflet polymer artificial heart valve was known to show remarkable improvement in antithrombosis and pressure drop compared with other type of artificial valve. In this investigation monoleaflet the vertical and horizontal deflection pattern of polymer heart valve with three types of supporting members straight member, and two curved members was analysed using the large deformation nonlinear finite element method.

  • PDF

모터 구동형 인공심장의 설계 및 해석(II) 해석 (Design and Analysis of Motor-Driven Artificial Heart ( II );Analysis)

  • 천길정;한동철;민병구
    • 대한기계학회논문집
    • /
    • 제15권3호
    • /
    • pp.855-868
    • /
    • 1991
  • 본 연구에서는 위에서 언급한바와 같은 제반사항들에 대해 이론적인 해석을 행하고, 그 결과를 이용하여 설계된 모델이 생체내에 이식될 수 있는 가능성의 여부를 판단하며 생체내에 이식 되었을 때의 성능과 생체에 미치는 영향등을 예측하고자 한다.

MUSIC을 이용한 기계식 심장 판막의 음향 신호 특성 연구 (A Study of the Acoustical Properties of the Mechanical Heart Valve Using MUSIC)

  • 이서우;최민주;민병구
    • 한국음향학회:학술대회논문집
    • /
    • 한국음향학회 1999년도 학술발표대회 논문집 제18권 2호
    • /
    • pp.131-134
    • /
    • 1999
  • This paper considers the acoustical characteristics of the mechanical valve employed in the Korean type Artificial Heart. $Bj\"{o}rk-Shiley$ tilting disc valve was chosen for the study and acoustic measurements were performed for the artificial heart operated in a mock circulation system as well as implanted to an animal as a Bi Ventricular Assist Device (BVAD). In the mock system, three different conditions of the valve were examined which were normal, damaged (torn off), pseudothrombus attached. Microphone measurements for the BVAD were carried out at a regular time interval for 5 days after the implantation operation. Of the recorded acoustic emissions from the artificial heart, click sounds mainly originated from the valves were further analyzed using Multiple Signal Classification (MUSIC) for estimating their spectral properties. It was shown that the spectral peaks below 4 kHz and the optimal order number for MUSIC, equivalent to the number of the spectral component, might be the key parameters which were highly correlated to the physiological states of the valve like the mechanical damage of the valve or the formation of thrombus on the valves.

  • PDF

설계방법론을 이용한 기계식 인공심장판막의 설계 (Design of a Mechanical Artificial Heart Valve Prosthesis Appliing Design Methodology)

  • 천길정;류형태
    • 대한의용생체공학회:의공학회지
    • /
    • 제19권3호
    • /
    • pp.291-296
    • /
    • 1998
  • 설계방법론을 적용하여 새로운 기계식 인공심장판막을 설계 하였다. 심장판막의 기능을 정의하고 기능구조 결과에 따라 체계적인 설계변이 탐색을 수행하였으며, 이에 의하여 다양한 가능성을 검토하였다. 심장판막에 요구되는 제반 요구조건들을 기준으로 하여 최적 모델을 결정하였다. 설계된 심장판막의 시제품을 제작하여 모의순환계를 이용한 거동특성측정 실험을 수행하였다. 설계된 신규모델은 유량과 파형특성 및 거동특성면에서 대체로 양호한 결과를 보였다.

  • PDF