Improved Lateral Resolution of Interferometric Microscope Using Precision Scanner (정밀 스캐너를 이용한 간섭 현미경의 가로방향 분해능 향상)

  • 박성림;박도민;류재욱;권대갑
    • Journal of the Korean Society for Precision Engineering
    • /
    • v.15 no.6
    • /
    • pp.116-123
    • /
    • 1998
  • An interferometric microscope with an improved lateral resolution is presented. The nanometer resolution XY stage is integrated into standard temporal phase shifting interferometer. The nanometer resolution XY stage is used to position specimen in subpixel of CCD detector, therefore CCD detector's sampling is improved. Two scanning algorithms and those simulation results are also presented. The simulation results show that scanning algorithms improve CCD detector's sampling significantly, and interferometeric microscope's lateral resolution is improved also.