• 제목/요약/키워드: Indenter Area Function

검색결과 2건 처리시간 0.016초

AFM을 이용한 나노 인덴터 팁의 면적함수 결정에 관한 연구 (A Study on Determination of the Area Function of Nano Indenter Tip with AFM)

  • 박성조;이현우;한승우
    • 한국정밀공학회지
    • /
    • 제21권6호
    • /
    • pp.145-152
    • /
    • 2004
  • Depth-sensing indentation is wifely used for evaluation of mechanical properties of thin films. It is generally accepted that the most significant source of uncertainty in nanoindentation measurement is the geometry of the indenter tip. Therefore the successful application of the technique requires accurate calibration of the indenter tip geometry. The direct measurement of geometry of a Berkovich indenter was determined using a atomic force microscope. The indentation geometrical calibration of contact area was performed by analyzing the indenter tip profile. The equations of area functions were proposed for nanoscale thin films..

압입자 첨단마모에 따른 나노압입곡선의 변화 및 이의 보정기법 (Variation of Nanoindentation Curve due to Wear of Indenter Apex and Its Correction Method)

  • 이윤희;김용일;박종서;김광호
    • 비파괴검사학회지
    • /
    • 제33권2호
    • /
    • pp.129-137
    • /
    • 2013
  • 나노압입시험기의 힘교정과 압입자에 대한 3차원 형상 관찰 및 분석이 본 연구에서 진행되었다. 표준분동으로 교정한 마이크로밸런스로 나노압입시험기에서 발생시킨 하중을 측정하여 측정치와 발생치의 비로 압입하중을 교정하였고, 나노압입시험의 시작점인 초기 접촉 하중도 확인할 수 있었다. 삼각뿔 압입자를 원자현미경으로 관찰하여 분석한 결과 비교적 사용이력이 없는 압입자 A와 마모된 압입자 B의 첨단곡률반경은 각각 $19.71{\pm}3.03$ nm와 $1043.94{\pm}50.91$ nm로 결정되었다. 완벽한 삼각뿔 압입자 형상과 중첩하여 압입자 A와 B의 첨단무딘깊이(bluntness depth)를 1.22 nm와 64.56 nm로 결정하였고, 용해실리카 기준시편에 수행한 나노압입시험 결과를 살펴본 결과 두 압입자의 압입하중-변위곡선들이 무딘깊이 차이만큼 수평축으로 서로 어긋나 있음을 확인할 수 있었다. 수평 이동을 통해 보정된 압입곡선의 분석을 통해 개별 압입자 면적함수에 대한 고려없이 1.11 % 이내에서 동일한 용해실리카의 나노경도를 결정할 수 있었다.