• 제목/요약/키워드: Impedance cardiogram

검색결과 5건 처리시간 0.019초

전기적 임피던스 방법을 통한 hand-held 타입의 심박출량 모니터링 시스템 개발 및 검증 (Development and Verification of a Hand-held Typed Monitoring System for Cardiac Output Measurement using Electrical Impedance Technique)

  • 서광석;심명헌;김민용;윤찬솔;정주홍;박성빈;윤형로
    • 전기학회논문지
    • /
    • 제61권12호
    • /
    • pp.1932-1938
    • /
    • 2012
  • The impedance cardiogram has been proposed as a non-invasive, continuous, operator independent, and cost-effective method for cardiac output monitoring. However, it can not be completely considered as non-restrictive method because of attached spot and band type electrodes. Therefore, we developed a improved convenient hand-held typed measurement system for cardiac output by electrical impedance technique. 80 subjects from Yonsei University and the surrounding areas, participated. All subjects measured stroke volume and cardiac output through Physioflow and developed system. To verify the developed system, statistical methods such as correlation, Wilcoxon signed ranks test, and the Bland-Altman analysis were used. The proposed system showed significant correlation in both male and female stroke volume(r=0.715, r=0.704) and cardiac output(r=0.826; r=0.804). From these results, it can be concluded that stroke volume and cardiac output could be improved convenient measurement using the both hands without the help of a specialist.

상호상관 관계를 이용한 운동중의 임피던스 파형에서의 특성점 검출 (Detection of Distinctive Points in Impedance Cardiogram during Exercise by Cross-Correlation Method)

  • 오인식;송철규
    • 대한의용생체공학회:의공학회지
    • /
    • 제12권4호
    • /
    • pp.261-266
    • /
    • 1991
  • As the ensemble averaged dz/dt signal during exercise gets smoothed, it is difficult to find the distinctive marks for estimation of stroke volume. The cross correlation function was made use of estimating these marks for automatic calculation by computer from the ensemble averaged dz/dt signal. LVET( Left Ventricular Ejection Time) and stroke volume were estimated based on the calculated parameters from the characteristic points. LVET, stroke volume calculated by hand, by the ensemble average and the cross correlation were compared for accuracy validation.

  • PDF

운동중의 임피던스 신호에서 상호상관 관계를 이용한 특성점의 검출 (Delection of Distinctive Points in Impedance Cardiogram during Exercise by Cross-Correlation Method)

  • 오인식;송철규;김덕원;차일환
    • 대한의용생체공학회:학술대회논문집
    • /
    • 대한의용생체공학회 1991년도 추계학술대회
    • /
    • pp.93-95
    • /
    • 1991
  • As the ensemble averaged dZ/dt signal during exercise is smoothed, it is difficult to find the distinctive marks. The cross correlation function was made use of estmating these marks. LVET was calculated based on the calculated parameters of the characteristic points. For the accuracy validation, LVET calculated by hand, by the ensemble average and the cross correl at ion were compared.

  • PDF

심전도와 심탄도의 무구속적 동시 측정을 위한 의자 등받이 개발 (Development of Chair Backrest for Non-intrusive Simultaneous Measurement of ECG and BCG)

  • 임용규
    • 융합신호처리학회논문지
    • /
    • 제19권3호
    • /
    • pp.104-109
    • /
    • 2018
  • 심전도와 심탄도를 무구속적으로 동시에 측정하는 시스템을 개발하였다. 제안된 시스템은 의자의 보조 등받이에 설치되어 맥파전달시간(PAT), Pre-ejection period (PEP) 등의 심혈관계 역학 정보를 일상생활에서 무구속적으로 측정하는 것을 목적으로 개발되었다. 본 연구에서는, 의자의 등받이에 부착된 용량성 능동전극과 용량성 접지를 이용한 간접접촉 심전도 (IDC-ECG) 측정 방법에 의해 무구속적으로 심전도를 측정하였으며, 등받이에 부착된 EMFi 압력 센서를 사용하여 무구속적으로 심탄도(ballistocardiogram)를 측정하였다. 제안된 시스템으로 심전도와 심탄도가 무구속적으로 동시에 측정됨을 확인하였으며, 측정된 심탄도는 피검자 사이에서 일정한 패턴을 보이는 것을 알 수 있었다. 또한 심전도의 Q파와 심탄도의 첫 번째 첨두간의 시간차가 pre-ejection period와 높은 상관관계를 보이는 것도 확인할 수 있었다. 이 연구에서 개발된 시스템은 심혈관계의 다양한 지표(심전도, 심박수, PAT, PEP 등)를 무구속적으로 일상생활에서 측정할 수 있음을 보였다.

색채 환경 기반의 인체 반응 정보 측정 시스템 (A Measurement System for Color Environment-based Human Body Reaction)

  • 김지언;정창원;주수종
    • 인터넷정보학회논문지
    • /
    • 제17권2호
    • /
    • pp.59-65
    • /
    • 2016
  • 색채 환경에 따라 인지반응의 상관관계를 분석한 결과가 다양한 산업분야에 적용되고 있다. 특히, 의학 분야에서 색채 환경기반 생체신호를 수집하여 환자 상태의 파악과 치료 그리고 뇌활동의 규명에 관련된 연구가 활발하게 진행되고 있다. 그러나 기존 연구들은 색채조명 또는 LED를 이용한 색채 환경을 구축하여 다양한 실험을 진행하고 있으나 다른 광원에 노출 되어 심리적인 영향을 주는 문제점을 갖고 있다. 따라서, 본 논문에서 제안하는 시스템은 HMD(Head-Mounted-Display)를 이용하여 피험자에게 완벽한 색채 환경조건을 제공하고자 한다. 그리고 인체반응정도 측정을 위해 기억력-주의력 테스트하는 동안 BMS(Bio-Medical-System)으로 생체정보를 수집하여 색채별 인체반응정도 측정뿐만 아니라 심리적인 변화에 대한 상관관계를 확인할 수 있는 시스템을 제안한다. 제안한 시스템을 통하여 수집하는 생체정보는 심전도(ECG), 호흡(Respiration), 산소포화도(SpO2), 임피던스(Bio-Impedance), 혈압(Blood-Pressure) 이며 데이터베이스에 저장한다. 이와함께, 인체의 인지반응을 측정하기위해 기억력-주의력 테스트는 앱으로 구현하여 색채 조건별 인지반응 정보를 측정하고 결과를 확인 할 수 있다. 제안한 시스템을 통하여 수집된 인체반응정보를 활용하여 생리학적 정보의 정량화하는 연구를 진행할 수 있으며, 색채 환경과 생리학적 정보의 상관관계를 분석하여 의학적 진단 및 치료에 응용이 될 것으로 기대한다. 향후 연구로 구축한 시스템을 통해 수집된 데이터를 분석 및 시각화 기능을 추가하고, 치매 질환의 예측 및 치료를 위한 시스템으로 확장하여 재활프로그램으로서의 유효성을 평가하고자 한다.