Evaluation of Treatment Planning for Head Tilting in WBRT 3D-CRT by TomoDirect mode: a Phantom Study (토모다이렉트를 이용한 3차원 전뇌 방사선치료에서 두상 각도에 따른 치료계획평가: 팬톰 실험)

  • Dae-Gun, Kim;Sang-Hyun, Kim
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.7
    • /
    • pp.857-862
    • /
    • 2022
  • The purpose of this study was to evaluate a three-dimensional conformal radiotherapy (3D-CRT) treatment plan with regard to head tilting in whole-brain radiotherapy (WBRT) using TomoDirect (TD) mode in Tomotherapy. WBRT 3D-CRT by TD was compared for a total of five head tilt angles (-20°, -10°, 0°. +10° and +20°). The dose homogeneity index (HI) and prescription dose index (CI) were calculated to confirm the target coverage. The maximum and average doses for critical organs such as the lens, eyeball and parotid glands were calculated for different angles of head tilting. The HI and CI were closet to the result value of 1 at the head tilted angle +10° and +20°. At a head tilted angle of +10°, the dose to the lens and eyeballs decreased by about 74% and about 30%, when compared with the reference angle (0°), respectively. The results of this study suggest that a head angle of +10 with chin-up would save adequate target coverage and reduce exposure dose to the lens.

Usefulness assessment of secondary shield for the lens exposure dose reduction during radiation treatment of peripheral orbit (안와 주변 방사선 치료 시 수정체 피폭선량 감소를 위한 2차 차폐의 유용성 평가)

  • Kwak, Yong Kuk;Hong, Sun Gi;Ha, Min Yong;Park, Jang Pil;Yoo, Sook Hyun;Cho, Woong
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.27 no.1
    • /
    • pp.87-95
    • /
    • 2015
  • Purpose : This study presents the usefulness assessment of secondary shield for the lens exposure dose reduction during radiation treatment of peripheral orbit. Materials and Methods : We accomplished IMRT treatment plan similar with a real one through the computed treatment planning system after CT simulation using human phantom. For the secondary shield, we used Pb plate (thickness 3mm, diameter 25mm) and 3 mm tungsten eye-shield block. And we compared lens dose using OSLD between on TPS and on simulation. Also, we irradiated 200 MU(6 MV, SPD(Source to Phantom Distance)=100 cm, $F{\cdot}S\;5{\times}5cm$) on a 5cm acrylic phantom using the secondary shielding material of same condition, 3mm Pb and tungsten eye-shield block. And we carried out the same experiment using 8cm Pb block to limit effect of leakage & transmitted radiation out of irradiation field. We attached OSLD with a 1cm away from the field at the side of phantom and applied a 3mm bolus equivalent to the thickness of eyelid. Results : Using human phantom, the Lens dose on IMRT treatment plan is 315.9cGy and the real measurement value is 216.7cGy. And after secondary shield using 3mm Pb plate and tungsten eye-shield block, each lens dose is 234.3, 224.1 cGy. The result of a experiment using acrylic phantom, each value is 5.24, 5.42 and 5.39 cGy in case of no block, 3mm Pb plate and tungsten eye-shield block. Applying O.S.B out of the field, each value is 1.79, 2.00 and 2.02 cGy in case of no block, 3mm Pb plate and tungsten eye-shield block. Conclusion : When secondary shielding material is used to protect critical organ while irradiating photon, high atomic number material (like metal) that is near by critical organ can be cause of dose increase according to treatment region and beam direction because head leakage and collimator & MLC transmitted radiation are exist even if it's out of the field. The attempt of secondary shield for the decrease of exposure dose was meaningful, but untested attempt can have a reverse effect. So, a preliminary inspection through Q.A must be necessary.

  • PDF

A Study Radiograph Techniques for Diagnosis of Axillary Fracture Ribs in X-ray (X-ray 갈비뼈 촬영에서 겨드랑이쪽의 골절된 갈비뼈 진단을 위한 촬영법 연구)

  • Ahn, Byung-Ju;Lee, Jun-Haeng
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.7
    • /
    • pp.979-986
    • /
    • 2019
  • The purpose of this study devised a method for the examination of the images of the first and second rib fractures of the axillary using Rib Rando Phantom. The position of the phantom and the angle of the X-ray tube were changed to vertical, head 5° and foot 5°. The Radiological Technologists subjectively evaluated the acquired images and the evaluation data were analyzed by SPSS 3.0 ver. The signal to noise ratio(SNR) was calculated using the ImageJ Program. As a result, the cronbach alpha value was significantly higher at 0.789. The SNR was highest at 6.038 when the X-ray tube was tilted 5° toward the head in the front of the ribs and highest at 7.860 when the X-ray tube was tilted 5° toward the foot. Radiographic Techniques proposed axillary ribs this study are as follows. The anterior rib scan is examined by elevating the fractured area and changing the X-ray tube angle by 5° towards the head. The posterior ribs scan is examined by attaching the fractured area and changing to the X-ray tube angle by 5° towards the foot. It is considered that such an inspection method can obtain a sharp image.

3-Dimensional Dosimetry of Small Field Photon Beam (광자선의 소조사면에서의 3차원적 선량 측정)

  • Jang, Ji-Sun;Kwon, Soo-Il
    • Progress in Medical Physics
    • /
    • v.23 no.1
    • /
    • pp.54-61
    • /
    • 2012
  • A polymer gel dosimeter was fabricated. A 3-dimensional dosimetry experiment was performed in the small field of the photon of the cyberknife. The dosimeter was installed in a head and neck phantom. It was manufactured from the acrylic and it was used in dosimetry. By using the head and neck CT protocol of the CyberKnife system, CT images of the head and neck phantom were obtained and delivered to the treatment planning system. The irradiation to the dosimeter in the treatment planning was performed, and then, the image was obtained by using 3.0T magnetic resonance imaging (MRI) after 24 hours. The dose distribution of the phantom was analyzed by using MATLAB. The results of this measurement were compared to the results of calculation in the treatment planning. In the isodose curve on the axial direction, the dose distribution coincided with the high dose area, 0.76mm difference on 80%, rather than the low dose area, 1.29 mm difference on 40%. In this research, the fact that the polymer gel dosimeter and MRI can be applied for analyzing a small field in a 3 dimensional dosimetry was confirmed. Moreover, the feasibility of using these for the therapeutic radiation quality control was also confirmed.

Co-registration of PET-CT Brain Images using a Gaussian Weighted Distance Map (가우시안 가중치 거리지도를 이용한 PET-CT 뇌 영상정합)

  • Lee, Ho;Hong, Helen;Shin, Yeong-Gil
    • Journal of KIISE:Software and Applications
    • /
    • v.32 no.7
    • /
    • pp.612-624
    • /
    • 2005
  • In this paper, we propose a surface-based registration using a gaussian weighted distance map for PET-CT brain image fusion. Our method is composed of three main steps: the extraction of feature points, the generation of gaussian weighted distance map, and the measure of similarities based on weight. First, we segment head using the inverse region growing and remove noise segmented with head using region growing-based labeling in PET and CT images, respectively. And then, we extract the feature points of the head using sharpening filter. Second, a gaussian weighted distance map is generated from the feature points in CT images. Thus it leads feature points to robustly converge on the optimal location in a large geometrical displacement. Third, weight-based cross-correlation searches for the optimal location using a gaussian weighted distance map of CT images corresponding to the feature points extracted from PET images. In our experiment, we generate software phantom dataset for evaluating accuracy and robustness of our method, and use clinical dataset for computation time and visual inspection. The accuracy test is performed by evaluating root-mean-square-error using arbitrary transformed software phantom dataset. The robustness test is evaluated whether weight-based cross-correlation achieves maximum at optimal location in software phantom dataset with a large geometrical displacement and noise. Experimental results showed that our method gives more accuracy and robust convergence than the conventional surface-based registration.

Usability Evaluation of Lateral Sliding Table in CT Examination (CT 검사에서 Lateral Sliding Table의 유용성 평가)

  • Choi, Jeong Hun;Kong, Chang gi;Song, Jong Nam;Han, Jae Bok
    • Journal of the Korean Society of Radiology
    • /
    • v.14 no.5
    • /
    • pp.677-684
    • /
    • 2020
  • Miscentering in the left and right X axis direction during CT examination affects dose and quality. When the CT Gantry Isocenter and the center of the examination objective are matched using the Lateral Sliding Table, the image quality is improved and the exposure dose is reduced. CTDI Head Phantom (Kimda, Korea) and dosimeter (Ray Safe, Sweden) were used to measure dose comparison CTDI (mGy) due to center deviation, and Water Phantom (HITACHI, Japan) was used to measure noise to see the difference in uniformity due to center deviation. Measurements of doses for dose comparison CTDI (mGy) with a deviation showed that doses were consistently reduced and exact dose was not projected until they were moved to 80 mm by 20 mm from the Isocenter. SD values were measured to see the difference in uniformity due to center deviation and the noise continued to increase until it was moved by 20 mm to 80 mm. The range of collimation has increased by the extent of deviating from the center and the range of exposure has increased. Using the Lateral Sliding Table, you can easily adjust the Isocenter, increase the quality of the image by adjusting the Isocenter in areaa such as the cardiac examination of the location away from the Isocenter, Extreme bone and Shoulder, and greatly reduce the collimation to the Isocenter, so it can be used to reduce unnecessary exposure dose.

The Image Distortion Analysis of Levin-tube tip by Patient position and Incidence Angle when taking Mobile Chest AP Projection (Mobile Chest AP 검사 시 환자자세와 입사각도에 따른 Levin-tube tip의 영상왜곡 분석)

  • Lee, Jinsoo;Park, Hyonghu
    • Journal of the Korean Society of Radiology
    • /
    • v.9 no.7
    • /
    • pp.467-471
    • /
    • 2015
  • This study's purpose is improve image quality to keep accurate tube angle in order to recognize distortion degree conditions by patient's position or tube angle and to provide exact clinical informations when taking chest AP projection for patient which have L-tube in stomach. The experimental equipment was ELMO-T6S by SHIMADZU corporation, then we put L-tube which attached 1 mm gap scales ruler on chest phantom surface. The experiment set by 90 kVp, 4 mAs, 120 cm distance. Each phantom position which changed supine, 30degree, 45degree, 60degree on the table exposured direct, ${\pm}5degree$, ${\pm}10degree$, ${\pm}15degree$ to head and feet directions. As a result, L-tube tip's position was changed by patient's position and tube angle. When patient's position is supine, tip's position change was lower than 30degree, 45degree, 60degree. We have to adjust patient's position or tube angle in order to occur image distortion by fault tube angle when confirming correct position L-tube tip through chest x-ray. Also, Radiological technologist try to make accurate evaluation index for satisfied L-tube insertion.

The Study of New Reconstruction Method for Brain SPECT on Dual Detector System (Dual detector system에서 Brain SPECT의 new reconstruction method의 연구)

  • Lee, Hyung-Jin;Kim, Su-Mi;Lee, Hong-Jae;Kim, Jin-Eui;Kim, Hyun-Joo
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.13 no.1
    • /
    • pp.57-62
    • /
    • 2009
  • Purpose: Brain SPECT study is more sensitive to motion than other studies. Especially, when applying 1-day subtraction method for Diamox SPECT, it needs shorter study time in order to prevent reexamination. We were required to have new study condition and analysing method on dual detector system because triple head camera in Seoul National University Hospital is to be disposed. So we have tried to increase image quality and make the dual and triple head to have equivalent study time by using a new analysing program. Materials and Methods: Using IEC phantom, we estimated contrast, SNR and FWHM. In Hoffman 3D brain phantom which is similar with real brain, we were on the supposition that 5% of injected doses were distributed in brain tissue. To compare with existing FBP method, we used fan-beam collimator. And we applied 15 sec, 25 sec/frame for each SEPCT studies using LEHR and LEUHR. We used OSEM2D and Onco-flash3D reconstruction method and compared reconstruction methods between applied Gaussian post-filtering 5mm and not applied as well. Attenuation correction was applied by manual method. And we did Brain SPECT to patient injected 15 mCi of $^{99m}Tc$-HMPAO according to results of Phantom study. Lastly, technologist, MD, PhD estimated the results. Results: The study shows that reconstruction method by Flash3D is better than exiting FBP and OSEM2D when studied using IEC phantom. Flowing by estimation, when using Flash3D, both of 15 sec and 25 sec are needed postfiltering 5 mm. And 8 times are proper for subset 8 iteration in Flash3D. OSEM2D needs post-filtering. And it is proper that subset 4, iteration 8 times for 15sec and subset 8, iteration 12 times for 25sec. The study regarding to injected doses for a patient and study time, combination of input parameter-15 sec/frame, LEHR collimator, analysing program-Flash3D, subset 8, iteration 8times and Gaussian post-filtering 5mm is the most appropriate. On the other hands, it was not appropriate to apply LEUHR collimator to 1-day subtraction method of Diamox study because of lower sensitivity. Conclusions: We could prove that there was also an advantage of short study time effectiveness in Dual camera same as Triple gamma camera and get great result of alternation from existing fan-beam collimator to parallel collimator. In addition, resolution and contrast of new method was better than FBP method. And it could improve sensitivity and accuracy of image because lesser subjectivity was input than Metz filter of FBP. We expect better image quality and shorter study time of Brain SPECT on Dual detector system.

  • PDF

Field-in-Field Technique to Improve Dose Distribution in the Junction of the Field with Head & Neck Cancer (Field-in-Field Technique을 이용한 두경부암의 접합부위 선량개선에 관한 고찰)

  • Kim, Seon-Myeong;Lee, Yeong-Cheol;Jeong, Deok-Yang;Kim, Young-Bum
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.21 no.1
    • /
    • pp.17-23
    • /
    • 2009
  • Purpose: In treating head and neck cancer, it is very important to irradiate uniform dose on the junction of the bilateral irradiation field of the upper head and neck and the anterior irradiation field of the lower neck. In order to improve dose distribution on the junction, this study attempted to correct non uniform dose resulting from under dose and over dose using the field-in-field technique in treating the anterior irradiation field of the lower neck and to apply the technique to the treatment of head and neck cancer through comparison with conventional treatment. Materials and Methods: In order to examine dose difference between the entry point and the exit point where beam diffusion happens in bilateral irradiation on the upper head and neck, we used an anthropomorphic phantom. Computer Tomography was applied to the anthropomorphic phantom, the dose of interest points was compared in radiation treatment planning, and it was corrected by calculating the dose ratio at the junction of the lower neck. Dose distribution on the junction of the irradiated field was determined by placing low-sensitivity film on the junction of the lower neck and measuring dose distribution on the conventional bilateral irradiation of the upper head and neck and on the anterior irradiation of the lower neck. In addition, using the field-in-field technique, which takes into account beam diffusion resulting from the bilateral irradiation of the upper head and neck, we measured difference in dose distribution on the junction in the anterior irradiation of the lower neck. In order to examine the dose at interest points on the junction, we compared and analyzed the change of dose at the interest points on the anthropomorphic phantom using a thermoluminescence dosimeter. Results: In case of dose sum with the bilateral irradiation of the upper head and neck when the field-in-field technique is applied to the junction of the lower neck in radiation treatment planning, The dose of under dose areas increased by 4.7~8.65%. The dose of over dose areas also decreased by 2.75~10.45%. Moreover, in the measurement using low-sensitivity film, the dose of under dose areas increased by 11.3%, and that of over dose areas decreased by 5.3%. In the measurement of interest point dose using a thermoluminescence dosimeter, the application of the field-in-field technique corrected under dose by minimum 7.5% and maximum 17.6%. Thus, with the technique, we could improve non.uniform dose distribution. Conclusion: By applying the field-in-field technique, which takes into account beam divergence in radiation treatment planning, we could reduce cold spots and hot spots through the correction of dose on the junction and, in particular, we could correct under dose at the entry point resulting from beam divergence. This study suggests that the clinical application of the field-in-field technique may reduce the risk of lymph node metastasis caused by under dose on the cervical lymph node.

  • PDF

Study of the CatcherTM Couch's Usefulness (토모치료기 CatcherTM Couch의 유용성에 대한 고찰)

  • Um, Ki Cheon;Lee, Chung Hwan;Jeon, Soo Dong;Song, Heung Kwon;Back, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.31 no.2
    • /
    • pp.65-74
    • /
    • 2019
  • Purpose: Recently, A Catcher was added to prevent sagging in Radixact® X9. In this study, We quantitatively compared general couch of Tomo-HDA® with catcher couch of Radixact® X9 using the human phantom and evaluated usefulness of catcher. Materials and methods: We used rando phantom for phantom study and set the each iso-center of head and neck region and Pelvis region for region parameter. Furthermore, We used hand made low melting point alloys for weight parameter. MVCT(Mega Voltage Computed Tomography) images were acquired for vertical error and rotation(pitch) error measurement increasing weight(A: 15kg, A+B: 30kg, A+B+C: 45kg). We selected 120 patients who has been treated using Tomotherpy machine for patient study. 60 patients has been treated in Tomo-HDA® and the other 60 patients treated in Radixact® X9. In the patient study methods, vertical error and rotation(pitch) error was measured for mean value calculation using MVCT images acquired on first day of radiation therapy. Result: Result of phantom study, Vertical error and rotation(pitch) error was increased proportionally increased as the weight increases in general couch of Tomo-HDA®. each maximum value was 7.52mm, 0.38° in head and neck region and 11.94mm, 0.92° in pelvis region. However, We could confirm that there was stable error range(0.02~0.1mm, 0~0.04°) in Catcher couch of Radixact®. Result of patient study, The head and neck region was measured 4.79mm 0.33° lower, and the pelvis region was measured 7.66mm, 0.22° lower in Catcher couch of Radixact® X9. Conclusion: In this study, Vertical error and rotation(pitch) error was proportionally increased as the weight increases in general couch of Tomo-HDA®. Especially, The pelvis region error was more increased than the head and neck region error. However, Vertical error and rotation(pitch) error was regularly generated regardless of weight or regions in CatcherTM couch of Radixact® X9 that this study's purpose. In conclusion, CatcherTM couch of Radixact® X9 can minimize mechanical error that couch sagging. Furthermore, The pelvis region is more efficiency than head and neck region. In radiation therapy using Tomotherapy machine, it is regarded that may contribute to minimizing unadjusted pitch error due to characters of Tomotherapy.