An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files

  • Gehlot, Paras Mull;Manjunath, Vinutha;Manjunath, Mysore Krishnaswamy
    • Restorative Dentistry and Endodontics
    • /
    • 제41권1호
    • /
    • pp.6-11
    • /
    • 2016
  • Objectives: The purpose of this in vitro study was to evaluate the accuracy of working length (WL) determination of four electronic apex locators (EALs), namely, Root ZX (RZX), Elements diagnostic unit and apex locator (ELE), SybronEndo Mini Apex locator (MINI) and Propex pixi (PIXI) using Stainless steel (SS) and nickel-titanium (NiTi) hand files. The null hypothesis was that there was no difference between canal length determination by SS and NiTi files of 4 EALs. Materials and Methods: Sixty extracted, single rooted human teeth were decoronated and the canal orifice flared. The actual length (AL) was assessed visually, and the teeth were embedded in an alginate model. The electronic length (EL) measurements were recorded with all four EALs using SS and NiTi files at '0.5' reading on display. The differences between the AL and EL were compared. Results: The results obtained with each EAL with SS and NiTi files were compared with AL. A paired sample t test showed that there was a statistical significant difference between EAL readings with SS and NiTi files for RZX and MINI (p < 0.05). The accuracy of RZX, ELE, MINI and PIXI within ${\pm}0.5 mm$ of AL with SS/NiTi files were 93.3%/70%, 90%/91.7%, 95%/68.3%, and 83.3%/83.3%, respectively. Conclusions: The results of this study indicate that Root ZX was statistically more accurate with NiTi files compared to SS files, while MINI was statistically more accurate with SS files compared to NiTi files. ELE and PIXI were not affected by the alloy type of the file used to determine WL.

4종 전자근관장측정기의 정확성과 일관성에 관한 in vitro 연구 (IN VITRO EVALUATION OF ACCURACY AND CONSISTENCY OF FOUR DIFFERENT ELECTRONIC APEX LOCATORS)

  • 조재현;금기연;이승종
    • Restorative Dentistry and Endodontics
    • /
    • 제31권5호
    • /
    • pp.390-397
    • /
    • 2006
  • 이번 연구는 서로 다른 4개의 전자근관장측정기의 정확성을 측정하고 각각 0.5지점과 Apex지점에서의 일관성을 비교하고자 하였다. 40개의 발치된 상하악 소구치를 대상으로 치수강 개방 후 alginate model에 고정시키고 근관장을 측정하였다. 사용된 전자근관장측정기는 Root ZX (Merits, Tokyo, Japan), SmarPex (META, Seoul, Korea). Elements Diagnostic Unit (SybronEndo, CA, USA), E-Magic Finder Deluxe (S-Denti, Seoul, Korea)이다. 먼저 모든 치아에서 4개의 전자근관장측정기를 사용하여 0.5지점과 Apex지점에서 근관장을 측정하여 한 치아당 8개의 측정값을 얻었다. 다음으로 치아를 각 전자근관장측정기당 10개씩 4개의 그룹으로 나누어, 각각 제조사의 지시대로 Root ZX, Elements Diagnostic Unit 및 E-Magic Finder Deluxe는 "0.5"지점에서, SmarPex는 "Apex"지점에서 file을 치아에 cement로 고정시켰다. 이후 치근단부 4 mm를 삭제하여 100배율의 Image Proplus로 관찰하여 file 끝에서 주근단공의 외연까지의 실제거리를 측정한 후, 4개의 전자근관장측정기의 0.5지점 및 Apex지점에서 file끝과 주근단공 사이의 거리를 계산하여 비교하였다. 그 결과 Root ZX와 E-Magic Finder는 실험군 100%, SmarPex는 90%, Elements Diagnostic Unit는 70%에서 주근단공과의 거리가 임상적 허용범위인 ${\pm}0.5 mm$이내에 있었다. 또한 각 전자근관장측정기 마다 0.5지점과 Apex지점에서의 근관장의 표준편차와 사분위 범위를 구하여 두 지점간의 일관성을 비교한 결과, Root ZX, E-Magic Finder는 0.5지점과 Apex지점에서 비슷한 일관성을 보였으며 SmarPex와 Elements Diagnostic unit는 Apex지점에서 0.5지점보다 더 높은 일관성을 보였다. 전자근관장측정기는 근관 내의 조건에 관계없이 근첨협착부에서 항상 일정한 거리를 재현해 낼 수 있는 일관성이 중요하므로, 이렇게 0.5지점 또는 Apex지점에서의 일관성이 증명된다면 실제 임상에서 사용할 때 전자근관장에서 일정한 거 리를 가감하여 사용할 수 있다.