• Title/Summary/Keyword: Electron beam therapy

Search Result 125, Processing Time 0.023 seconds

The Dosimetric evaluation of the standard electron cone for the extended cone for the extended SSD and The Dosimetric characteristics of the custom-made electron cone (표준 전자선 cone의 확장된 SSD에서의 선량평가 및 자체제작한 전자선 cone의 특성)

  • Chung Se Young;Chung Hui Young;Kim Young Bum;Kwon Young Ho
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.11 no.1
    • /
    • pp.73-78
    • /
    • 1999
  • In general, the patients of the head and neck cancer are treated with 4MV photon beam up to prescribed dose, but spinal cord should be excluded in the treatment field. When its absorbed dose is limited at the tolerance dose. In case of the patients who has the positive posterior neck nodes need a boost electron beam treatment to the prescribed dose. In that case, the anatomical structure of the neck and the physical structure of the standard electron cone interrupt to allow proper access to the disease site. Therefore, we extended treatment SSD for the remove of the those hindrances. In this study, we evaluated the dosimetric variation of the standard electron cone for the extended SSD, from 100cm to 120cm, 5 cm increment, and compare to the custom-made electron cone. As a result, the $\%$ depth dose, the point of maximum dose and the range of maximum were changed within the $2\%$. The penumbra width was increased from 1.0cm to 2.0cm. However, the dosimetric characteristics of the custom-made electron cone was very similar to that of the 100cm SSD standard electron cone and due to its characteristic of physical structure, patients didn't need re-positioning after photon beam treatment, therefore accurate treatment was possible, we conclude that the custom-made electron cone was very useful for the clinical practice.

  • PDF

Fabrication and Characterization of a Fiber-Optic Radiation Sensor for High Energy Electron Beam Therapy (치료용 고에너지 전자선 계측을 위한 광섬유 방사선 센서의 제작 및 특성 분석)

  • Jang, Kyoung-Won;Cho, Dong-Hyun;Yoo, Wook-Jae;Lee, Bong-Soo;Yi, Jeong-Han;Tack, Gye-Rae;Cho, Hyo-Sung;Kim, Sin
    • Journal of Biomedical Engineering Research
    • /
    • v.27 no.6
    • /
    • pp.332-336
    • /
    • 2006
  • In this study, we have fabricated a fiber-optic radiation sensor using an organic scintillator for high energy electron beam therapy. The intensities of scintillating light from a fiber-optic radiation sensor are measured with different field size, electron beam energy and monitor unit of a clinical linear accelerator. To obtain percent depth dose(PDD), the amount of scintillating light is measured at different depth of polymethylmethacrylate(PMMA) phantom. Also the intensity of Cerenkov light is measured and characterized as a function of incident angle of electron beam and a subtraction method is investigated using a background optical fiber to remove a Cerenkov light.

A Study on the Dosimetry of the Total Skin Electron Beam Therapy in Cutaneous T-Cell Lymphoma (피부 T 세포림프종의 전 피부 전자선 치료를 위한 dosimetry 연구)

  • 신교철;윤형근
    • Progress in Medical Physics
    • /
    • v.7 no.2
    • /
    • pp.57-65
    • /
    • 1996
  • Total Skin Electron Beam Therapy (TSEBT) is one of the most effective treatment methods for superficially disseminated skin cancer or cutaneous T-cell lymphoma. We have treated a patient with cutaneous T-cell lymphoma. We have used Stanford technique using six dual field. The nominal energy of electron beam was 4MeV. SSD was 390cm and the gantry angles of dual fields were 76$^{\circ}$ and 104$^{\circ}$. The dose profiles of single field and dual fields were measured with films and a Farmer type ion chamber. The field uniformity was 10% over the patient's surface. During treatment, the patient was placed in six different positions for homogenous dose distribution over the body surface. The areas not directly exposed to the path of the electron beam (soles of feet, perineum and vertex of scalp) were boosted with 7MeV electron beam. During the treatment, lens, fingernails and toenails were shielded.

  • PDF

Study on Absorbed Dose Determination of Electron Beam Quality for Cross-calibration with Plane-parallel Ionization Chamber (평행평판형이온함의 교차교정 시 전자선 선질에 따른 흡수선량 결정에 대한 연구)

  • Rah, Jeong-Eun;Shin, Dong-Oh;Park, So-Hyun;Jeong, Ho-Jin;Hwang, Ui-Jung;Ahn, Sung-Hwan;Lim, Young-Kyung;Kim, Dong-Wook;Yoon, Myong-Geun;Shin, Dong-Ho;Lee, Se-Byeong;Suh, Tae-Suk;Park, Sung-Yong
    • Progress in Medical Physics
    • /
    • v.20 no.2
    • /
    • pp.97-105
    • /
    • 2009
  • Absorbed dose to water based protocols recommended that plane-parallel chambers be calibrated against calibrated cylindrical chambers in a high energy electron beam with $R_{50}$>7 $g/cm^2$ (E${\gtrsim}$16 MeV). However, such high-energy electron beams are not available at all radiotherapy centers. In this study, we are compared the absorbed dose to water determined according to cross-calibration method in a high energy electron beam of 16 MeV and in electron beam energies of 12 MeV below the cross-calibration quality remark. Absorbed dose were performed for PTW 30013, Wellhofer FC65G Farmer type cylindrical chamber and for PTW 34001, Wellhofer PPC40 Roos type plane-parallel chamber. The cylindrical and the plane-parallel chamber to be calibrated are compared by alternately positioning each at reference depth, $Z_{ret}=0.6R_{50}-0.1$ in water phantom. The $D_W$ of plane-parallel chamber are derived using across-calibration method at high-energy electron beams of 16, 20 MeV. Then a good agreement is obtained the $D_W$ of plane-parallel chamber in 12 MeV. The agreement between 20 MeV and 12 MeV are within 0.2% for IAEA TRS-398.

  • PDF

Evaluation of Detection Performance of TlBr Materials for the Development of Electron Beam Quality Assurance Dosimeters (전자선 Quality Assurance 선량계 개발을 위한 TlBr 물질의 검출성능 평가)

  • Yang, Seung-Woo;Park, Sung-Kwang
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.5
    • /
    • pp.513-518
    • /
    • 2022
  • Electron beam quality assurance (QA) should be done regularly for accurate radiation therapy. However, QA tools used in clinical practice are designed mainly for X-rays. So, a dosimeter for electron beam QA is required. Therefore, in this study, the electron beam detection performance was measured by using a thorium bromide material as an electron beam sensor. In addition, it was evaluated whether it could be applied with an electron beam QA dosimeter. Reproducibility, linearity, and dose rate dependence were evaluated at 6 MeV and 9 MeV energies. As a result of reproducibility, it showed a maximum output change of 0.92% at 6 MeV and 1.15% at 9 MeV. The linearity result evaluation and determination coefficient were presented as 0.9998. As a result of dose rate dependence evaluation, relative standard deviation 0.51% at 6 MeV and relative standard deviation 1.07% at 9 MeV were presented. The manufactured TlBr sensor shows the ability to detect radiation that meets the criteria for evaluation of reproducibility, linearity, and dose rate dependence. These results mean that the TlBr dosimeter is applicable as an electron beam QA dosimeter.

Design and Application of Acrylic Electron Wedge to Improve Dose Inhomogeneities at the Junction of Electron Fields (전자선 조사야 결합부분의 선량분포 개선을 위한 Acrylic Electron Wedge의 제작 및 사용)

  • Kim Young Bum;Kwon Young Ho;Whang Woong Ku;Kim You Hyun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.10 no.1
    • /
    • pp.60-68
    • /
    • 1998
  • Treatment of a large diseased area with electron often requires the use of two or more adjoining fields. In such cases, not only electron beam divergence and lateral scattering but also fields overlapping and separation may lead to significant dose inhomogeneities(${\pm}20\%$) at the field junction area. In this study, we made Acrylic Electron Wedges to improve dose homogeneities(${\pm}5\%$) in these junction areas and considered application it to clinical practices. All measurements were made using 6, 9, 12, 16, 20MeV Electron beams from a linear accelerator for a $10{\times}10cm$ field at 100cm SSD. Adding a 1 mm sheet of acryl gradually from 1 mm to 15 mm, We acquired central axis depth dose beam profile and isodose curves in water phantom. As a result, for all energies, the practical range was reduced by approximately the same distance as the thickness of the acryl insert, e.g. a 1 mm thick acryl insert reduce the practical range by approximately 1 mm. For every mm thickness of acryl inserted, the beam energy was reduced by approximately 0.2MeV. These effects were almost independent of beam energy and field size. The use of Acrylic Electron Wedges produced a small increase $(less\;than\;3\%)\;in\;the\;surface\;dose\;and\;a\;small\;Increase(less\;than\;1\%)$ in X-ray contamination. For acryl inserts, thickness of 3 mm or greater, the penumbra width increased nearly linear for all energies and isodose curves near the beam edge were nearly parallel with the incident beam direction, and penumbra width was $35\;mm{\sim}40\;mm$. We decide heel thickness and angle of the wedge at this point. These data provide the information necessary to design Acrylic Electron Wedge which can be use to improve dose uniformity at electron field junctions and it will be effectively applicated in clinical practices.

  • PDF

Monte Carlo Calculation of the Dose Profiles for a 6 MeV Electron Beam with Longitudinal Magnetic Fields

  • Jeong, Dong-Hyeok;Oh, Young-Kee;Kim, Jhin-Kee;Kim, Jeung-Kee;Shin, Kyo-Chul;Kim, Ki-Hwan;Lee, Jeong-Ok;Kang, Jeong-Ku;Moon, Sun-Rock
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2002.09a
    • /
    • pp.109-112
    • /
    • 2002
  • Using Monte Carlo calculations the effects of longitudinal magnetic fields on the beam profiles produced by clinical electron beam were studied. The Monte Carlo calculations were performed using the EGS4 code system modified to account for external magnetic fields. The beam profiles for a 6 MeV electron beam with longitudinal magnetic fields of 0.5-3.0 T were calculated. As a result of these calculations we found that the penumbra widths can be reduced with increased magnetic fields. This means that the electron therapy benefits from the external magnetic fields.

  • PDF

Study of Acute Myelocytic Leukemia Patient Treatment That Used Total Skin Electron Beam (Total Skin Electron Beam을 이용한 급성 골수성 백혈병 환자 치료에 대한 연구)

  • Lee, Sang-Ryul;Kang, Min-Kyu;Kim, Sung-Kyu
    • Progress in Medical Physics
    • /
    • v.20 no.3
    • /
    • pp.152-158
    • /
    • 2009
  • Total Skin Electron Beam Therapy (TSEBT) of linear accelerator has become use so as to be useful, 2~9 MeV of energy territories came to be used with mycosis fungoides and cutaneous lymphomas in the superficial lesion treatment which covers the major portion of the body. I treat a patient to Stanford technique in this study, and it is $60^{\circ}$ around the patients whom Stanford technique irradiated electronic beam to a linear accelerator in horizontal directions and there is a way a standard of TSEBT treat it to six located field (anterior, posterior, and four obliques) becoming. An each field does horizontally it and consist to beam of the two component which fitted the center to a suitable angle. a patient treats it to three dual field a day in order to make short treatment time. when a first day, we treat one dual field at anterior position and two dual field at posterior position. when the second day, treat one dual field at posterior position and two dual field at anterior position. Therefore, six dual field is finished in perfect periodic two days. we made cylindrical acrylic phantom, and I inserted a dosimeter film between phantom. in order to measure a dose distribution calculation before treat a patient, and a patient checked it in six field directions that got from a treatment. It is after that thermoluminescent dosimetry (TLD) as it uses Rando phantom and then measurement dose distribution in six field directions after attaching at chest, the right and left flank, a back after irradiation.

  • PDF

Evaluation of dose variation at the vertex during Total Skin Electron Beam (전신 피부 전자선 조사(TSEB)시 두정부(Vertex)에서의 선량 변화 평가)

  • Jeon Byeong-Chul;An Seung-Kwon;Lee Sang-Gyu;Kim Joo-Ho;Cho Kwang-Hwan;Cho Jung-Hee;Park Jae-Il
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.12 no.1
    • /
    • pp.112-116
    • /
    • 2000
  • Purpose : The vertex scalp is always tangentially irradiated during total skin electron beam(TSEB) This study was discuss to the dose distribution at the vertex scalp and to evaluate the use of an electron reflector. positioned above the head as a means of improving the dose uniformity. Methods and Materials Vetex dosimetry was performed using ion-chamber and TLD. Measurements were 6 MeV electron beam obtained by placing an acrylic beam speller in the beam line. Studies were performed to investigate the effect of electron scattering on vertex dose when a lead reflector $40{\times}40cm$ in area, was positioned above the phantom. Results : The surface dose at the vertex, in the without of the reflector was found to be less than $37.8\%$ of the skin dose. Use of the lead reflector increased this value to $62.2\%$ for the 6 MeV beam. Conclusion : The vertex may be significantly under-dosed using standard techniques for total skin electron beam. Use of an electron reflector improves the dose uniformity at the vertex and may reduce or eliminate the need for supplemental irradiation.

  • PDF

Evaluation of the Usefulness of Tungsten Nanoparticles as an Alternative to Lead Shielding Materials in Electron Beam Therapy (전자선 치료시 납 차폐체 대체물질로서의 텅스텐 나노입자의 유용성 평가)

  • Kim, Ji-Hyang;Kim, Na-Kyoung;Lee, Gyu-Yeong;Jung, Da-Bin;Heo, Yeong-Cheol
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.7
    • /
    • pp.949-956
    • /
    • 2021
  • The purpose of this paper is to evaluate whether tungsten nanoparticles have a shielding effect on scattered light generated at high doses as an alternative material to lead used to shield scattered light in electron beam therapy. A plate was manufactured to set the position of the dosimeter and the size of the radiation field to be constant. The glass dosimeter was placed at 12 points, which were 1, 2, and 4 cm apart from the center of the field of 10 × 10 cm2 in the cross direction. A total of 12 types of tungsten nanoparticle shields were developed with a thickness of 0.75 mm to 4.00 mm and a size of 10 × 10 cm2 using 0.4, 0.75, and 1 mm materials. Using a linear accelerator, measurements were made four times at 6 MeV and four times at 12 MeV, and the dose intensity was investigated at 100 MU. The 4 mm shielding plate showed the highest shielding effect at 1 cm from the irradiation field. The 1 mm shielding plate at 2 cm from the irradiation field had the lowest shielding effect. As the thickness of the tungsten shielding plate increased, the electron beam's shielding effect increased sharply. It was confirmed that tungsten nanoparticles can reduce the amount of scattered light generated by electron beam therapy. Therefore, this study will provide basic data when follow-up studies are conducted on the shielding ability of tungsten nanoparticles.