A Study on clinical Considerations caused by inevitably Extended SSD for Electron beam therapy (확장된 SSD에 기인한 Electron beam의 Output 및 특성 변화에 관한 연구)

  • Lee, Jeong-U;Kim, Jeong-Man
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.8 no.1
    • /
    • pp.29-35
    • /
    • 1996
  • We are often faced with the clinical situations that is inevitably extended SSD for electron beam therapy due to anatomical restriction or applicator structure. But there are some difficulties in accurately predicting output and properties. In electron beam treatment , unlike photon beam the decrease in output for extended SSD does not follow inverse-square law accurately because of a loss of side scatter equilibrium, which is particularly significant for small cone size and low energies. The purpose of our study is to analyze the output in changing with the energy, cone size, air gap beyond the standard SSD and to compare inverse-square law factor derived from calculated effective SSD, mominal SSD with measured output factor. In addition, we have analyzed the change of PDD for several cones with different SSDs which range from 100cm to 120cm with 5cm step and with different energies(6MeV, 9MeV, 12MeV, 16MeV, 20MeV). In accordance with our study, an extended SSD produces a significant change in beam output, negligible change in depth dose which range from 100cm to 120cm SSDs. In order to deliver the more accurate dose to the neoplastic tissue, first of all we recommend inverse-square law using the table of effective SSDs with cone sizes and energies respectively or simply to create a table of extended SSD air gap correction factor. The second we need to have an insight into some change of dose distribution including PPD, penumbra caused by extended SSD for electron beam therapy.

  • PDF

Determination of output factors by 1D method for 6MeV electron (1D 방법에 의한 6MeV 전자선의 output 인자 결정)

  • 유명진
    • Progress in Medical Physics
    • /
    • v.13 no.1
    • /
    • pp.27-31
    • /
    • 2002
  • Output factors for 6MeV electron were directly measured under the condition of an individual beam cutout and these factors were compared with the output factors by 1D method which is an easy means to predict the output factors of electron beam. Output factors by 1D method are defined as output factors of rectangular fields where one side is always equal to the side of the square reference field and the output of an arbitrary rectangular field X, Y is given by the product of the 1D output factors. The output of very large square fields is overestimated using 1D method for the 6MeV electron, but it results in agreement with measured data under the condition of an individual cutout within 1% error adopting a correction factor $CF=C\times$[(X-10)(Y-10)/$\mid$(X-10)(Y-10)$\mid^{1/2}]$.

  • PDF

Dose Characteristics for IORT Applicator of ML-15MDX Electron Beam (ML-15MDX 술중조사용 Applicator에 의한 전자선선량 특성)

  • Choi, Tae-Jin;Lee, Ho-Joon;Kim, Yeung-Ae;Kim, Jin-Hee;Kim, Ok-Bae
    • Radiation Oncology Journal
    • /
    • v.11 no.2
    • /
    • pp.455-461
    • /
    • 1993
  • Experimental measurements of dose characteristics with pentagonal applicator at nominal energy of 4, 6, 9, 12 and 15 MeV electron beam were performed for intraoperative radiotherapy (IORT) in ML-15MDX linear accelerator. This paper presents the percent depth dose, surface dose, beam flatness and output factors of using the IORT applicator in different electron beam energy. The output factor showed as a 24 percent higher in IORT applicator than that of reference $10{\times}10cm^2$ applicator. The surface dose of using the IORT applicator showed 7.7 and 2.7 percent higher than that of reference field in 4 and 15 MeV electron beam, respectively. In our experiments, the variation of percent depth dose was very small but the output factor and flatnees at 0.5 cm depth have showed a large value in IORT applicator.

  • PDF

Determination of Electron Beam Output Factors of Individual Applicator for ML-15MDX Linear Accelerator (선형가속기 ML-15MDX의 각 Applicator에 대한 전자선 출력선량 계수 결정)

  • Park, Tae-Jin;Kim, Ok-Bae
    • Progress in Medical Physics
    • /
    • v.5 no.1
    • /
    • pp.87-99
    • /
    • 1994
  • Purpose : The determination of electron beam output factor was investigated from individual applicator for various energy of ML-15MDX linear accelerator. The output factor of electron beam was extended from square to rectangular field in individual applicator size through with a least-square fit to a polynomial expression. Materials : In this experiments. the measurement of output was obtained from 2${\times}$cm$^2$ to 20${\times}$20cm$^2$ of field size in different applicator size for 4 to 15 MaV electron beam energy. The output factor was defined as the ratio of maximum dose output on the central axis of the field of individual applicator size to that of a given field size. Applicator factors were derived from comparing with the output dose of reference field size 10${\times}$10cm$^2$. The thickness of block was specially designed as 10mm in thickness of Lipowitz metal for field shaping in all electron energy. Two types of output curves are included as output factors versus side of square fields and that of variable side length for X and Y in one-dimensional to compare the expected values to that of experiments. Results : Expected output factors of rectangular which was derived from that of square fields in individual applicator size from 2${\times}$2cm$^2$ to 20${\times}$20cm$^2$ in different electron energy was very closed to that of experimental measurements within 2% uncertainty. However 1D method showed a 3% discrepancy in small rectangular field for low energy electron beam. Conclusion : Emperical non-linear polynomial regressions of square root and 1D method were performed to determin the output factor in various field size and electron energy. The expected output of electron beam of square root method for square field and 1D method for rectangular field were very closed to that of measurement in all selected electron beam energy.

  • PDF

Measurement of Energy Parameters for Electron Gun Heater Currents and Output Dose Rate for Electron Beams from a Prototype Linac (연구용 선형가속기의 전자총 가열 전류에 따른 전자선의 에너지 인자 측정과 출력 측정 연구)

  • Lim, Heuijin;Lee, Manwoo;Kim, Me Young;Yi, Jungyu;Lee, Mujin;Kang, Sang Ku;Rhee, Dong Joo;Jeong, Dong Hyeok
    • Progress in Medical Physics
    • /
    • v.27 no.1
    • /
    • pp.25-30
    • /
    • 2016
  • The dosimetric characteristics were experimentally evaluated for electron beams from the prototype linac developed for radiotherapy units. This paper focuses on the electron beam output and energy variations as a function of electron gun heater current. The electron energy was derived from its mean and most probable energies measured by film dosimetry. The electron beam output at the maximum electron energy was measured with the plane parallel ionization chamber in water using TRS-398 dosimetry protocol. The mean energy and the most probable energy of the electron beam were 6.54~3.31 MeV and 5.94~2.80 MeV at electron gun current of 2.02~2.50 A respectively. The output dose rate for an electron beam of mean energy 6.54 MeV was 5.41 Gy/min ${\pm}1.5%$ at the reference depth in water.

Assessment of the usefulness of the Machine Performance Check system that is an evaluation tools for the determination of daily beam output (일간 빔 출력 확인을 위한 평가도구인 Machine Performance Check의 유용성 평가)

  • Lee, Sang Hyeon;Ahn, Woo Sang;Lee, Woo Seok;Choi, Jin Hyeok;Kim, Seon Yeon
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.29 no.2
    • /
    • pp.65-73
    • /
    • 2017
  • Purpose: Machine Performance Check (MPC) is a self-checking software based on the Electronic Portal Imaging Device (EPID) to measure daily beam outputs without external installation. The purpose of this study is to verify the usefulness of MPC by comparing and correlating daily beam output of QA Beamchecker PLUS. Materials and Methods: Linear accelerator (Truebeam 2.5) was used to measure 10 energies which are composed of photon beams(6, 10, 15 MV and 6, 10 MV-FFF) and electron beams(6, 9, 12, 16 and 20 MeV). A total of 80 cycles of data was obtained by measuring beam output measurement before treatment over five months period. The Pearson correlation coefficient was used to evaluate the consistency of the beam output between the MPC and the QA Beamchecker PLUS. In this study, if the Pearson correlation coefficient is; (1) 0.8 or higher, the correlation is very strong (2) between 0.6 and 0.79, the correlation is strong (3) between 0.4 and 0.59, the correlation is moderate (4) between 0.2 and 0.39, the correlation is weak (5) lower than 0.2, the correlation is very weak. Results: Output variations observed between MPC and QA Beamchecker PLUS were within 2 % for photons and electrons. The beam outputs variations of MPC were $0.29{\pm}0.26%$ and $0.30{\pm}0.26%$ for photon and electron beams, respectively. QA Beamchecker PLUS beam outputs were $0.31{\pm}0.24%$ and $0.33{\pm}0.24%$ for photon and electron beams, respectively. The Pearson correlation coefficient between MPC and QA Beamchecker PLUS indicated that photon beams were very strong at 15 MV, and strong at 6 MV, 10 MV, 6 MV-FFF and 10 MV-FFF. For electron beams, the Pearson correlation coefficient were strong at 16 MeV and 20 MeV, moderate at 9 MeV and 12 MeV, and very weak at 6 MeV. Conclusion: MPC showed significantly strong correlation with QA Beamchecker PLUS when testing with photon beams and high-energy electron beams in the evaluation of daily beam output, but the correlation when testing with low-energy electron beams (6 MeV) appeared to be low. However, MPC and QA Beamchecker PLUS are considered to be suitable for checking daily beam output, as they performed within 2 % of beam output consistency during the observation. MPC which can perform faster than the conventional daily beam output measurement tool, is considered to be an effective method for users.

  • PDF

Measurement of Bremsstrahlung Radiation with Electron Beam Energy

  • Srivastava, R.P.;Chaurasia, P.P.;Prasiko, G.;Jha, A.K.
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2002.09a
    • /
    • pp.235-236
    • /
    • 2002
  • A Klystron powered dual photon energy electron linear accelerator 2300 C/D from Varian Associates has been installed in our center. From the radiological safety view as well as treatment planning, the output (contamination) of Bremsstrahlung Radiation with electron beam energy determined accurately. It has been found 0.5% to 4.7% with increasing the electron beam energy which is the clinically not much significant in the treatment of the malignant diseases with the treatment of electron beam.

  • PDF

Using Nanosecond Electron Beam to Produce Silver Nanopowder

  • Balezin, M.E.;Timoshenkova, O.R.;Sokovnin, S.Yu.;Lee, Hi-Min;Rhee, C.K.
    • Journal of Powder Materials
    • /
    • v.15 no.6
    • /
    • pp.466-470
    • /
    • 2008
  • Experiments with a URT-0.5 accelerator (0.5 MeV, 50 ns, 1 kW) generating a nanosecond electron beam for irradiation of silver nitrate in various liquid solutions (water and toluene) were performed with the aim of producing silver nanopowders. A radiochemical reaction allows making weakly agglomerated pure Ag powders with particles of 10-15 nm and 30-50 nm in size by irradiation in toluene and water respectively. The injection of the nanosecond electron beam energy to the solution is optimal. As the absorbed dose increases, the output of the radiochemical reaction does not grow, but more agglomerated powders are synthesized.

Study on Characteristics of Dose Distribution in Tissue of High Energy Electron Beam for Radiation Therapy (방사선 치료용 고에너지 전자선의 조직 내 선량분포 특성에 관한 연구)

  • Na, Soo-Kyung
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.14 no.1
    • /
    • pp.175-186
    • /
    • 2002
  • The purpose of this study is directly measure and evaluate about absorbed dose change according to nominal energy and electron cone or medical accelerator on isodose curve, percentage depth dose, contaminated X-ray, inhomogeneous tissue, oblique surface and irradiation on intracavitary that electron beam with high energy distributed in tissue, and it settled standard data of hish energy electron beam treatment, and offer to exactly data for new dote distribution modeling study based on experimental resuls and theory. Electron beam with hish energy of $6{\sim}20$ MeV is used that generated from medical linear accelerator (Clinac 2100C/D, Varian) for the experiment, andwater phantom and Farmer chamber md Markus chamber und for absorbe d dose measurement of electron beam, and standard absorbed dose is calculated by standard measurements of International Atomic Energy Agency(IAEA) TRS 277. Dose analyzer (700i dose distribution analyzer, Wellhofer), film (X-OmatV, Kodak), external cone, intracavitary cone, cork, animal compact bone and air were used for don distribution measurement. As the results of absorbed dose ratio increased while irradiation field was increased, it appeared maximum at some irradiation field size and decreased though irradiation field size was more increased, and it decreased greatly while energy of electron beam was increased, and scattered dose on wall of electron cone was the cause. In percentage depth dose curve of electron beam, Effective depth dose(R80) for nominal energy of 6, 9, 12, 16 and 20 MeV are 1.85, 2.93, 4.07, 5.37 and 6.53 cm respectively, which seems to be one third of electron beam energy (MeV). Contaminated X-ray was generated from interaction between electron beam with high energy and material, and it was about $0.3{\sim}2.3\%$ of maximum dose and increased with increasing energy. Change of depth dose ratio of electron beam was compared with theory by Monte Carlo simulation, and calculation and measured value by Pencil beam model reciprocally, and percentage depth dose and measured value by Pencil beam were agreed almost, however, there were a little lack on build up area and error increased in pendulum and multi treatment since there was no contaminated X-ray part. Percentage depth dose calculated by Monte Carlo simulation appeared to be less from all part except maximum dose area from the curve. The change of percentage depth dose by inhomogeneous tissue, maximum range after penetration the 1 cm bone was moved 1 cm toward to surface then polystyrene phantom. In case of 1 cm and 2 cm cork, it was moved 0.5 cm and 1 cm toward to depth, respectively. In case of air, practical range was extended toward depth without energy loss. Irradiation on intracavitary is using straight and beveled type cones of 2.5, 3.0, 3.5 $cm{\phi}$, and maximum and effective $80\%$ dose depth increases while electron beam energy and size of electron cone increase. In case of contaminated X-ray, as the energy increase, straight type cones were more highly appeared then beveled type. The output factor of intracavitary small field electron cone was $15{\sim}86\%$ of standard external electron cone($15{\times}15cm^2$) and straight type was slightly higher then beveled type.

  • PDF