Determination of output factors by 1D method for 6MeV electron
1D 방법에 의한 6MeV 전자선의 output 인자 결정
Abstract
Output factors for 6MeV electron were directly measured under the condition of an individual beam cutout and these factors were compared with the output factors by 1D method which is an easy means to predict the output factors of electron beam. Output factors by 1D method are defined as output factors of rectangular fields where one side is always equal to the side of the square reference field and the output of an arbitrary rectangular field X, Y is given by the product of the 1D output factors. The output of very large square fields is overestimated using 1D method for the 6MeV electron, but it results in agreement with measured data under the condition of an individual cutout within 1% error adopting a correction factor
6MeV 전자선의 output 인자를 결정하기 위해 간편하게 output 인자를 예측하는 수단인 1D(Dimension) 방법을 이용하였으며 그리고 1D 방법으로 결정된 output 인자를 개별적 beam cutout 하에서 직접 측정한 output 인자와 비교하여 1D 방법의 정확성을 분석하였다. 1D 방법은 정방형 field의 한 변과 항상 동일한 한 변을 갖는 사각형 field의 output 인자로서 규정되며, 임의의 사각형 field (X,Y)의 output은 1D output 인자의 곱으로 주어진다. 6MeV 전자선의 대단히 큰 정방형 field의 output은 1D 방법을 사용하면 실제보다 높은 값을 나타내지만 교정인자 CF=C.[(X-10)(Y-10)/