• 제목/요약/키워드: Dichloroacetamide (DCAD)

검색결과 2건 처리시간 0.016초

Fluorimetric Determination of Dichloroacetamide by RPLC with Postcolumn Detection

  • Choi, Yong-Wook;Reckhow, David A.
    • Bulletin of the Korean Chemical Society
    • /
    • 제25권6호
    • /
    • pp.900-906
    • /
    • 2004
  • An RPLC-postcolumn detection method has been developed for the fluorimetric determination of dichloroacetamide (DCAD) in water. After ammonia and DCAD were separated on a $C_{18}$ nonpolar stationary phase with 2.5% methanol-0.02 M phosphate buffer at pH 3, the column eluant was reacted with post column reagents, o-phthaldialdehyde (OPA) and sulfite ion at pH 11.5, to produce a highly fluorescent isoindole fluorophore, which was measured with a fluorescence detector ( ${\lambda}_{ex}$ = 363 nm, ${\lambda}_{em}$ = 425 nm). With the optimized conditions for RPLC and the postcolumn derivatization, the calibration curve was found to be linear in the concentration ranges of 0.5 and 20 ${\mu}$M for DCAD, and the detection limit for DCAD was 0.18 ${\mu}$M (23${\mu}$g/L). This corresponded to 18 pmol per 100 ${\mu}$L injection volume for a signal-to-noise ratio of 3, and the repeatability and reproducibility of this method were 1.0% and 2.5% for five replicate analyzes of 2 ${\mu}$M DCAD, respectively. The degradation yields DCAD to ammonia were 94 and 99%, and the percent recoveries of DCAD from 4 and 6 ${\mu}$M DCAD-spiked tap water were shown mean more than 97%.

Simultaneous Fluorimetric Determination of On-line Preconcentrated HANs, DCAD and TCAD by Using RPLC with a Postcolumn Derivatization System

  • Jung, Sung-Woon;Choi, Yong-Wook
    • Bulletin of the Korean Chemical Society
    • /
    • 제34권6호
    • /
    • pp.1783-1790
    • /
    • 2013
  • A simultaneous analytical method has been developed for the fluorimetric determination of haloacetonitriles (HANs) [dichloroacetonitrile (DCAN), trichloroacetonitrile (TCAN), dibromoacetonitrile (DBAN), haloacetamides [dichloroacetamide (DCAD), and trichloroacetamde (TCAD)] in drinking water by using the combined on-line perconcentration/reversed phase liquid chromatography (RPLC)-postcolumn detection system. This on-line perconcentration system was achieved by employing a precolumn packed with a commercial solid phase extraction (SPE) sorbent for the enrichment and purification of the target analytes. The haloacetonitriles and haloacetamides were separated on CN analytical column in a 7.5% methanol-0.02 M phosphate buffered mobile phase at pH 3. The column effluents were reacted with postcolumn reagents of ophthaldialdehyde (OPA) and sulfite ion at pH 11.5, to produce a highly fluorescent isoindole fluorophore, which were measured with a fluorescence detector. Under the optimized conditions for RPLC and the postcolumn derivatization system all of the coefficient of determination of the standard calibration curves for the target analytes were over 0.99 and had a linear range from 5 to 100 ${\mu}g/L$. The detection limits showed 1.6 ${\mu}g/L$ for DCAD, 0.1 ${\mu}g/L$ for TCAD, 0.6 ${\mu}g/L$ for DCAN, 1.6 ${\mu}g/L$ for TCAN and 1 ${\mu}g/L$ for DBAN, and the recoveries were ranged from 64 to 99% except for DCAD with precisions less than 4.9% in distilled water, and from 72(${\pm}4%$) to 116%(${\pm}2%$) in tap water.