In this study, we tried to quantify the actual cross-sectional area inside the bore when scanning by the MRI system with various bore sizes. To this end, a comparative analysis was conducted by both of blueprint of each MRI equipment and actual measurement in the field. As a result of analysis, ACSA(Actual Cross-Sectional Area) in Ingenia CX, Elition X, uMR 780, Omega, Vida, Lumina, Architect, Premier is recorded as 171230, 232150, 242100, 309332, 230760, 230760, 229380 and 235990 ㎟, respectively ACSA% was 60.6, 60.3, 73.0, 70.0, 60.0, 60.0, 59.6, and 61,3%. In addition, DTB (Distance from Table top to Bore top) recorded 400, 407, 445, 495, 405, 405, 405, 403, and 412 mm. Through this study, it was confirmed that there is a difference between the bore size according to each MRI system and the actual cross-sectional area during MRI scanning. Accordingly, if we consider the internal actual area just not bore size at the clinical site, useful diagnostic images can be obtained in the end with better convenience.