• Title/Summary/Keyword: CTDIvol

Search Result 39, Processing Time 0.029 seconds

A study of image evaluation and exposure dose with the application of Tube Voltage and ASIR of Low dose CT Using Chest Phantom (흉부 Phantom을 이용한 Low Dose CT의 관전압과 ASIR(Adaptive Statistical Iterative Reconstruction)적용에 따른 영상평가 및 피폭선량에 관한 연구)

  • Hwang, Hyeseong;Kim, Nuri;Jeong, Yoonji;Goo, Eunhoe;Kim, Kijeong
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.16 no.2
    • /
    • pp.9-14
    • /
    • 2014
  • Purpose: The purpose of this study has attempted to evaluate and compare the image evaluation and exposure dose by respectively applying Filtered Back Projection(FBP), the existing test method, and Adaptive Statistical Iterative Reconstruction(ASIR) with different values of tube voltage during the Low Dose Computed Tomography(LDCT). Materials and Methods: With the image reconstruction method as basis, Chest Phantom was utilized with the FBP and ASIR set at 10%, 20% respectively, and the change of Tube Voltage (100kVp, 120kVp). For image evaluation, Back ground noise, Signal to Noise ratio(SNR) and Contrast to Noise ratio(CNR) were measured, and, for dose evaluation, CTDIvol and DLP were measured respectively. The statistical analysis was tested with SPSS(ver. 22.0), followed by ANOVA Test conducted after normality test and homogeneity test. (p<0.05). Results: In terms of image evaluation, there was no outstanding difference in Ascending Aorta(AA) SNR and Infraspinatus Muscle(IM) SNR with the different values of ASIR application(p<0.05), but a significant difference with the different amount of tube voltage(p>0.05). Also, there wasn't noticeable change in CNR with ASIR and different amount of Tube Voltage (p<0.05). However, in terms of dose evaluation, CTDIvol and DLP showed contrasting results(p<0.05). In terms of CTDIvol, the measured values with the same tube voltage of 120kVp were 2.6mGy with No-ASIR and 2.17mGy with 20%-ASIR respectively, decreased by 0.43mGy, and the values with 100kVp were 1.61mGy with No-ASIR and 1.34mGy with 20%-ASIR, decreased by 0.27mGy. In terms of DLP, the measured values with 120kVp were $103.21mGy{\cdot}cm$ with No-ASIR and $85.94mGy{\cdot}cm$ with 20%-ASIR, decreased by $17.27mGy{\cdot}cm$(about 16.7%), and the values with 100kVp were $63.84mGy{\cdot}cm$ with No-ASIR and $53.25mGy{\cdot}cm$ with 20%-ASIR, a decrease by $10.62mGy{\cdot}cm$(about 16.7%). Conclusion: At lower tube voltage, the rate of dose significantly decreased, but the negative effects on image evaluation was shown due to the increase of noise. For the future, through the result of the experiment, it is considered that the method above would be recommended for follow-up patients or those who get health checkup as long as there is no interference on the process of diagnosis due to the characteristics of Low Dose examination.

  • PDF

A Study on the Reduction of Patient's Exposure Dose according to the Arrival Time of Contrast Media in Abdominal CT Scan using Bolus Tracking Technique (Bolus tracking 기법을 이용한 복부 CT 검사 시 조영제 도달시간에 따른 환자 피폭선량 감소에 관한 연구)

  • Lee, Seung yong;Han, Dong kyoon
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.2
    • /
    • pp.93-100
    • /
    • 2021
  • The purpose of this study is to find out the factors that affect the patient's exposure dose during the abdominal CT scan using the bolus tracking technique, and reducing the radiation exposure to the patient during the abdominal CT scan using the bolus tracking technique by adjusting the delay time according to the corresponding factor. The experiment was divided into two parts, and the first experiment was conducted with 300 patients There were 188 males and 112 females, and their average age was 58±12.18(19~85). In the second experiment, 150 subjects (100 males and 50 females) who were undergoing their follow-up examination among the first experiment subjects, and the difference in dose before and after was compared by applying the delay time according to the influencing factor. As a result of the first experiment, there was a relationship between the arrival time of the contrast media and the heart rate, and it was found that the arrival time decreased as the heart rate increased for both men and women. As a result of the second experiment, the average dose of CTDIvol and DLP before/after applying the delay time according to the heart rate decreased 4.98 mGy and 5.33 mGy·cm in the male group, and 3.54 mGy and 3.88 mGy·cm in the female group. By applying proper delay time according to the patient's heart rate during abdominal CT scan with the bolus tracking technique, the radiation exposure dose of the patient can be reduced.

The Usability Assessment of Self-developed Phantom for Evaluating Automatic Exposure Control System Using Three-Dimensions Printing (자동노출제어장치 평가를 위한 3D 프린팅 기반의 자체 제작 팬텀의 유용성 평가)

  • Lee, Ki-Baek;Nam, Ki-Chang;Kim, Ho-Chul
    • Journal of Biomedical Engineering Research
    • /
    • v.41 no.4
    • /
    • pp.147-153
    • /
    • 2020
  • This study was to evaluate the usability of self-developed phantom for evaluating automatic exposure control (AEC) using three-dimensions (3D) printer. 3D printer of fused deposition modeling (FDM) type was utilized to make the self-developed AEC phantom and image acquisitions were conducted by two different type of scanners. The self-developed AEC phantom consisted of four different size of portions. As a result, two types of phantom (pyramid and pentagon shape) were created according to the combination of the layers. For evaluating the radiation dose with the two types of phantom, the values of tube current, computed tomography dose index volume (CTDIvol), and dose length product (DLP) were compared. As a result, it was confirmed that the values of tube current were properly reflected according to the thickness, and the CTDIvol and DLP were not significantly changed regardless of AEC functions of different scanners. In conclusion, the self-developed phantom by using 3D printer could assess whether the AEC function works well. So, we confirmed the possibility that a self-made phantom could replace the commercially expensive AEC performance evaluation phantom.

Dose Assessment in Accordance with the Measured Position of Size Specific Dose Estimates (Size Specific Dose Estimates(SSDE)측정 위치에 따른 피폭선량 평가)

  • Kim, Jung-Su;Hong, Sung-Wan;Kim, Jung-Min
    • Journal of radiological science and technology
    • /
    • v.38 no.4
    • /
    • pp.383-387
    • /
    • 2015
  • This study investigated the size specific dose estimates of difference localizer on pediatric CT image. Seventy one cases of pediatric abdomen-pelvic CT (M:F=36:35) were included in this study. Anterior-posterior and lateral diameters were measured in axial CT images. Conversion factors from American Association of Physicists in Medicine (AAPM) report 204 were obtained for effective diameter to determine size specific dose estimate (SSDE) from the CT dose index volume (CTDIvol) recorded from the dose reports. For the localizer of mid-slice SSDE was 107.63% higher than CTDIvol and that of xiphoid-process slices SSDE was higher than 92.91%. The maximum error of iliac crest slices, xiphoid process slices and femur head slices between mid-slices were 7.48%, 17.81% and 14.04%. In conclusion, despite the SSDE of difference localizer has large number of errors, SSDE should be regarded as the primary evaluation tool of the patient radiation in pediatric CT for evaluation.

Usability Evaluation of Applied Low-dose CT When Examining Urinary Calculus Using Computed Tomography (컴퓨터 단층촬영을 이용한 요로결석 검사에서 저선량 CT의 적용에 대한 유용성 평가)

  • Kim, Hyeon-Jin;Ji, Tae-Jeong
    • The Journal of the Korea Contents Association
    • /
    • v.17 no.6
    • /
    • pp.81-85
    • /
    • 2017
  • The aim of this study was to evaluate the usability of applied Low dose Computed Tomography(LDCT) protocol in examining urinary calculus using computed tomography. The subjects of this study were urological patients who visited a medical institution located in Busan from June to December 2016 and the protocol used in this study was Adaptive Statistical Iterative Reconstruction: low-dose CT with 50% Adaptive Statistical Iterative Reconstruction (ASIR). As results of quantitative analysis, the mean pixel value and standard deviation within kidney region of image(ROI)of the axial image were $26.21{\pm}7.08$ in abdomen CT pre scan and $20.03{\pm}8.16$ in low-dose CT. Also the mean pixel value and standard deviation within kidney ROI of the coronal image were $22.07{\pm}7.35$ in abdomen CT pre scan and $21.67{\pm}6.11$ in low dose CT. The results of qualitative analysis showed that four raters' mean values of observed kidney artifacts were $19.14{\pm}0.36$ when using abdomen CT protocol and $19.17{\pm}0.43$ in low-dose CT, and the mean value of resolution and contrast was $19.35{\pm}0.70$ when using abdomen CT protocol and $19.29{\pm}0.58$ in low-dose CT. Also the results of a exposure dose analysis showed that the mean values of CTDIvol and DLP in abdomen CT pre scan were 18.02 mGy and $887.51mGy{\cdot}cm$ respectively and the mean values of CTDIvol and DLP when using low-dose CT protocol were 7.412 mGy and $361.22mGy{\cdot}cm$ respectively. The resulting dose reduction rate was 58.82% and 59.29%, respectively.

Evaluation of Image Quality in Low Tube-Voltage Chest CT Scan (흉부 CT 검사 시 저 관전압 영상의 화질평가에 관한 연구)

  • Kim, Hyun-Ju;Cho, Jae-Hwan;Park, Cheol-Soo
    • Journal of Radiation Protection and Research
    • /
    • v.35 no.4
    • /
    • pp.135-141
    • /
    • 2010
  • Purpose : The patients who visited this department for pulmonary disease and need CT scans for Follow-up to observe change of CT value, evaluation of image quality and decrease of radiation dose as change of kVp. Subjects and Methods : Subjects were the patients of 20 persons visited this department for pulmonary disease and Somatom Sensation 16(Semens, Enlarge, Germany) was used. Measurement of CT value as change of kVp was done by setting up ROI diameter of 1cm at the height of thyroid, aortic arch, right pulmonary artery in arterial phase image using 100 kVp, measuring 3 times, and recorded the average. CT value of phantom was measured by scanning phantoms which means contrast media diluted by normal saline by various ratio with tube voltage of 80 kVp, 100 kVp, 120 kVp, 140 kVp and recorded the average of 3 CT values of center of phantom image. In analysing radiation dose, CTDIVOL values of the latest arterial phase image of 120 kVp and as this research set that of 100 kVp were analyzed comparatively. 2 observers graded quality of chest images by 5 degrees (Unacceptable, Suboptimal, Adequate, Good, Excellent). Results : CT value of chest image increased at 100 kVp by 14.06%~27.26% in each ROI than 120 kVp. CT value of phantom increased as tube voltage lowered at various concentration of contrast media. CTDIVOL decreased at 100 kVp(5.00 mGy) by 36% than 120 kVp(7.80 mGy) in radiation dose analysis. here were 0 Unacceptable, 1 Suboptimal, 3 Adequate, 10 Good, 6 Excellent in totally 20 persons. Conclusion : Chest CT scanning with low kilo-voltage for patients who need CT scan repeatedly can bring images valuable for diagnose, and decrease radiation dose against patients.

Evaluation for Optimization of CT Dose Reduction Methods in PET/CT (PET/CT 검사 시 CT 피폭선량 감소 방법들의 최적화 평가)

  • Do, Yong Ho;Lee, Hong Jae;Kim, Jin Eui
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.19 no.2
    • /
    • pp.55-62
    • /
    • 2015
  • Purpose Various methods for reducing radiation exposure have been continuously being developed. The aim of this study is to evaluate effectiveness of dose reduction, image quality and PET SUV changes by applying combination of automatic exposure dose(AEC), automated dose-optimized selection of X-ray tube voltage(CAREkV) and sinogram affirmed iterative reconstruction(SAFIRE) which can be controled by user. Materials and Methods Torso, AAPM CT performance and IEC body phantom images were acquired using biograph mCT64, (Siemens, Germany) PET/CT scanner. Standard CT condition was 120 kV, 40 mAs. Radiation exposure and noise were evaluated by applying AEC, CAREkV(120 kV, 40 mAs) and SAFIRE(120 kV, 25 mAs) with torso phantom compare to standard CT condition. And torso, AAPM and IEC phantom images were acquired with combination of 3 methods in condition of 120 kV, 25 mAs to evaluate radiation exposure, noise, spatial resolution and SUV changes. Results When applying AEC, CTDIvol and DLP were decreased by 50.52% and 50.62% compare to images which is not applying AEC. mAs was increased by 61.5% to compensate image quality according to decreasing 20 kV when applying CAREkV. However, CTDIvol and DLP were decreased by 6.2% and 5.5%. When reference mAs was the lower and strength was the higher, reduction of radiation exposure rate was the bigger. Mean SD and DLP were decreased by 2.2% and 38% when applying SAFIRE even though mAs was decreased by 37.5%(from 40 mAs to 25 mAs). Combination of 3 methods test, SD decreased by 5.17% and there was no significant differences in spatial resolution. And mean SD and DLP were decreased by 6.7% and 36.9% compare to 120 kV, 40 mAs with AEC. For SUV test, there was no statistical differences(P>0.05). Conclusion Combination of 3 methods shows dose reduction effect without degrading image quality and SUV changes. To reduce radiation exposure in PET/CT study, continuous effort is needed by optimizing various dose reduction methods.

  • PDF

Medical Radiation Exposure in Children CT and Dose Reduction (소아 CT 촬영시 방사선 피폭과 저감화 방법)

  • Lee, Jeong-Keun;Jang, Seong-Joo;Jang, Young-Ill
    • The Journal of the Korea Contents Association
    • /
    • v.14 no.1
    • /
    • pp.356-363
    • /
    • 2014
  • Recently pediatric CT has been performed by reduced dose according to tube current modulation이라고, this fact has a possibility more reduce a dose because of strong affect depend on tube current modulation. Almost all MDCT snow show and allow storage of the volume CT dose index (CTDIvol), dose length product (DLP), and effective dose estimations on dose reports, which are essential to assess patient radiation exposure and risks. To decrease these radiation exposure risks, the principles of justification and optimization should be followed. justification means that the examination must be medically indicated and useful. Results is using tube current modulation이라고 tend to the lower kV, the lower effective dose. In case of use a low dose CT protocol, we found a relatively lower effective dose than using tube current modulation. Average effective dose of our studies(brain, chest, abdomen-pelvis) less than 47%, 13.8%, 25.7% of germany reference dose, and 55.7%, 10.2%, 43.6% of UK(United Kingdom) reference dose respectively. when performed examination for reduced dose, we must use tube current modulation and low dose CT protocol including body-weight based tube current adaption.

Measurement of Radiation Dose of HR CT and Low Dose CT by using Anthropomorphic Chest Phantom and Glass Dosimetry (인체등가형 흉부팬텀과 유리선량계를 이용한 고해상력 및 저선량 CT의 선량측정)

  • Kweon, Dae Cheol
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.7
    • /
    • pp.933-939
    • /
    • 2019
  • The purpose of this study is to provide basic clinical data by evaluating images, measuring absorbed dose and effective dose by using high resolution CT and low dose CT by using anthropomorphic chest phantom and glass dosimeter. Tissue dose was measured by inserting a glass dosimeter into the anthropomorphic chest phantom. A 64-slice CT system (SOMATOM Sensation 64, Siemens AG, Forchheim, Germany) and CARE Dose 4D were used, and the parameters of the high resolution CT were 120 kVp, Eff. Scan parameters of mAs 104, scan time 7.93 s, slice 1.0 mm (Acq. 64 × 0.6 mm), convolution kernel (B60f sharp) were used, and low dose CT was 120 kVp, Eff. mAs 15, scan time 7.41 s, slice 3.0 mm (Acq. 64 × 0.6 mm), scan of convolution kernel B50f medium sharp. CTDIvol was measured at 8.01 mGy for high resolution CT and 1.18 mGy for low dose CT. Low dose CT scans showed 85.49% less absorbed dose than high resolution CT scans.

Dose Reduction and Image Quality Assessment of the CareDose 4D Technique on Abdomen Liver Computed Tomography (복부 간 CT 검사에서 CareDose 4D 사용에 따른 선량 감소 및 화질 평가)

  • Seok, Jong-Min;Jeon, Woo-Jin;Park, Young-Joon;Lee, Jin
    • Journal of the Korean Society of Radiology
    • /
    • v.11 no.3
    • /
    • pp.109-115
    • /
    • 2017
  • The purpose of this study was to evaluate the clinical efficacy of 128 MDCT (multi-detector computed tomography) for reducing the CareDose 4D dose and comparing the image quality with the fixed tube current technique. For this purpose, we conducted the phantom and clinical studies to evaluate the exposure dose and image of the subject before and after applying the CareDose 4D system in abdominal examination using 128 MDCT. In the phantom study, ROI (Region of interest) was located at the center, 3, 6, 9, 12 o'clock, into two groups: group A without CareDose 4D and Group B applied were measured. In the clinical study, ROI was located at the liver 8 segments, divided into two groups too. The measured items were CT number, noise, and dose length product (DLP) dose. The result of CTDIvol (CT Dose Index volume) measurements in phantom and clinical studies were lower than those before CareDose 4D application, and dose and effective dose were also measured lower (p<.05). There was no difference in CT number before and after application (p>.05). In conclusion, using CareDose 4D, we can obtain optimal image information without deteriorating image quality while reducing patient dose.