• Title/Summary/Keyword: CLYC

Search Result 3, Processing Time 0.018 seconds

Characterization of a CLYC Detector and Validation of the Monte Carlo Simulation by Measurement Experiments

  • Kim, Hyun Suk;Smith, Martin B.;Koslowsky, Martin R.;Kwak, Sung-Woo;Ye, Sung-Joon;Kim, Geehyun
    • Journal of Radiation Protection and Research
    • /
    • v.42 no.1
    • /
    • pp.48-55
    • /
    • 2017
  • Background: Simultaneous detection of neutrons and gamma rays have become much more practicable, by taking advantage of good gamma-ray discrimination properties using pulse shape discrimination (PSD) technique. Recently, we introduced a commercial CLYC system in Korea, and performed an initial characterization and simulation studies for the CLYC detector system to provide references for the future implementation of the dual-mode scintillator system in various studies and applications. Materials and Methods: We evaluated a CLYC detector with 95% $^6Li$ enrichment using various gamma-ray sources and a $^{252}Cf$ neutron source, with validation of our Monte Carlo simulation results via measurement experiments. Absolute full-energy peak efficiency values were calculated for gamma-ray sources and neutron source using MCNP6 and compared with measurement experiments of the calibration sources. In addition, behavioral characteristics of neutrons were validated by comparing simulations and experiments on neutron moderation with various polyethylene (PE) moderator thicknesses. Results and Discussion: Both results showed good agreements in overall characteristics of the gamma and neutron detection efficiencies, with consistent ~20% discrepancy. Furthermore, moderation of neutrons emitted from $^{252}Cf$ showed similarities between the simulation and the experiment, in terms of their relative ratios depending on the thickness of the PE moderator. Conclusion: A CLYC detector system was characterized for its energy resolution and detection efficiency, and Monte Carlo simulations on the detector system was validated experimentally. Validation of the simulation results in overall trend of the CLYC detector behavior will provide the fundamental basis and validity of follow-up Monte Carlo simulation studies for the development of our dual-particle imager using a rotational modulation collimator.

Design and Fabrication of CLYC-Based Rotational Modulation Collimator (RMC) System for Gamma-Ray/Neutron Dual-Particle Imager

  • Kim, Hyun Suk;Lee, Jooyub;Choi, Sanghun;Bang, Young-bong;Ye, Sung-Joon;Kim, Geehyun
    • Journal of Radiation Protection and Research
    • /
    • v.46 no.3
    • /
    • pp.112-119
    • /
    • 2021
  • Background: This work aims to develop a new imaging system based on a pulse shape discrimination-capable Cs2LiYCl6:Ce (CLYC) scintillation detector combined with the rotational modulation collimator (RMC) technique for dual-particle imaging. Materials and Methods: In this study, a CLYC-based RMC system was designed based on Monte Carlo simulations, and a prototype was fabricated. Therein, a rotation control system was developed to rotate the RMC unit precisely, and a graphical user interface-based software was also developed to operate the data acquisition with RMC rotation. The RMC system was developed to allow combining various types of collimator masks and detectors interchangeably, making the imaging system more versatile for various applications and conditions. Results and Discussion: Operational performance of the fabricated system was studied by checking the accuracy and precision of the collimator rotation and obtaining modulation patterns from a gamma-ray source repeatedly. Conclusion: The prototype RMC system showed reliability in its mechanical properties and reproducibility in the acquisition of modulation patterns, and it will be further investigated for its dual-particle imaging capability with various complex radioactive source conditions.

Gamma/neutron classification with SiPM CLYC detectors using frequency-domain analysis for embedded real-time applications

  • Ivan Rene Morales;Maria Liz Crespo;Mladen Bogovac;Andres Cicuttin;Kalliopi Kanaki;Sergio Carrato
    • Nuclear Engineering and Technology
    • /
    • v.56 no.2
    • /
    • pp.745-752
    • /
    • 2024
  • A method for gamma/neutron event classification based on frequency-domain analysis for mixed radiation environments is proposed. In contrast to the traditional charge comparison method for pulse-shape discrimination, which requires baseline removal and pulse alignment, our method does not need any preprocessing of the digitized data, apart from removing saturated traces in sporadic pile-up scenarios. It also features the identification of neutron events in the detector's full energy range with a single device, from thermal neutrons to fast neutrons, including low-energy pulses, and still provides a superior figure-of-merit for classification. The proposed frequency-domain analysis consists of computing the fast Fourier transform of a triggered trace and integrating it through a simplified version of the transform magnitude components that distinguish the neutron features from those of the gamma photons. Owing to this simplification, the proposed method may be easily ported to a real-time embedded deployment based on Field-Programmable Gate Arrays or Digital Signal Processors. We target an off-the-shelf detector based on a small CLYC (Cs2LiYCl6:Ce) crystal coupled to a silicon photomultiplier with an integrated bias and preamplifier, aiming at lightweight embedded mixed radiation monitors and dosimeter applications.