• Title/Summary/Keyword: Build-up modifier (BM)

Search Result 1, Processing Time 0.014 seconds

Characteristics of Photon Beam through a Handmade Build-Up Modifier as a Substitute of a Bolus (Bolus를 대체하기 위해 자체 제작된 선량상승영역 변환기를 투과한 광자선의 특성)

  • Kim, Sung Joon;Lee, Seoung Jun;Moon, Su Ho;Seol, Ki Ho;Lee, Jeong Eun
    • Progress in Medical Physics
    • /
    • v.25 no.4
    • /
    • pp.225-232
    • /
    • 2014
  • We evaluated the effect of scatter on a build-up region based on the measured percent depth dose (PDD) of high-energy photon beams that penetrated a handmade build-up modifier (BM) as a substitute of bolus. BM scatter factors ($S_{BM}$) were calculated based on the PDDs of photon beams that penetrated through the BM. The calculated $S_{BM}$ values were normalized to 1 at the square field side (SFS) of 30 mm without a BM. For the largest SFS (200 mm), the SBM values for a 6-MV beam were 1.331, 1.519, 1.598, 1.641, and 1.657 for the corresponding BM thickness values. For a 10-MV beam, the $S_{BM}$ values were 1.384, 1.662, 1.825, 1.913, and 2.001 for the corresponding BM thickness values. The BM yielded 76% of the bolus efficiency. We expect BM to become useful devices for deep-set patient body parts to which it is difficult to apply a bolus.