The goal of this study is to reduce patient exposure dose by providing image quality and radiation dose according to inspection methods. Volume Computed Tomography Dose Index(CTDIvol) and Dose Length Product(DLP) of prospective and retrospective ECG gating snapshot segment of Coronary CT angiography(CTA) were measured each snapshot segment methods. CT number, noise, uniformity, and resolution were also measured using phantom under the same condition of coronary CTA. The results showed that CT number, noise, uniformity and resolution are similar to each other. In terms of CTDIvol and DLP, however, measurement dose of prospective ECG gating snapshot segment was lower than the retrospective case by 37.5% and 40.3%. Therefore, it is highly recommended that in the coronary CTA, prospective ECG gating scan mode should be chosen to reduce patient dose.
Kim, Jin-Su;Lee, Jae-Sung;Lee, Byeong-Il;Lee, Dong-Soo;Chung, June-Key;Lee, Myung-Chul
The Korean Journal of Nuclear Medicine
/
v.38
no.4
/
pp.318-324
/
2004
Purpose: Philips GEMINI is a newly introduced whole-body GSO PET/CT scanner. In this study, performance of the scanner including spatial resolution, sensitivity, scatter fraction, noise equivalent count ratio (NECR) was measured utilizing NEMA NU2-2001 standard protocol and compared with performance of LSO, BGO crystal scanner. Methods: GEMINI is composed of the Philips ALLEGRO PET and MX8000 D multi-slice CT scanners. The PET scanner has 28 detector segments which have an array of 29 by 22 GSO crystals ($4{\times}6{\times}20$ mm), covering axial FOV of 18 cm. PET data to measure spatial resolution, sensitivity, scatter fraction, and NECR were acquired in 3D mode according to the NEMA NU2 protocols (coincidence window: 8 ns, energy window: $409[\sim}664$ keV). For the measurement of spatial resolution, images were reconstructed with FBP using ramp filter and an iterative reconstruction algorithm, 3D RAMLA. Data for sensitivity measurement were acquired using NEMA sensitivity phantom filled with F-18 solution and surrounded by $1{\sim}5$ aluminum sleeves after we confirmed that dead time loss did not exceed 1%. To measure NECR and scatter fraction, 1110 MBq of F-18 solution was injected into a NEMA scatter phantom with a length of 70 cm and dynamic scan with 20-min frame duration was acquired for 7 half-lives. Oblique sinograms were collapsed into transaxial slices using single slice rebinning method, and true to background (scatter+random) ratio for each slice and frame was estimated. Scatter fraction was determined by averaging the true to background ratio of last 3 frames in which the dead time loss was below 1%. Results: Transverse and axial resolutions at 1cm radius were (1) 5.3 and 6.5 mm (FBP), (2) 5.1 and 5.9 mm (3D RAMLA). Transverse radial, transverse tangential, and axial resolution at 10 cm were (1) 5.7, 5.7, and 7.0 mm (FBP), (2) 5.4, 5.4, and 6.4 mm (3D RAMLA). Attenuation free values of sensitivity were 3,620 counts/sec/MBq at the center of transaxial FOV and 4,324 counts/sec/MBq at 10 cm offset from the center. Scatter fraction was 40.6%, and peak true count rate and NECR were 88.9 kcps @ 12.9 kBq/mL and 34.3 kcps @ 8.84 kBq/mL. These characteristics are better than that of ECAT EXACT PET scanner with BGO crystal. Conclusion: The results of this field test demonstrate high resolution, sensitivity and count rate performance of the 3D PET/CT scanner with GSO crystal. The data provided here will be useful for the comparative study with other 3D PET/CT scanners using BGO or LSO crystals.
A detector module measuring the depth of interaction(DOI) was designed to improve the spatial resolution of positron emission tomography(PET). The scintillation pixel array consists of two layers, and a light guide is inserted between the layers to make the light generated through the gamma-ray event different for each layer. There are four light guides, and one light guide is designed to be coupled to a 2 × 2 array of scintillation pixels. The light generated from the top layer is moved to the photosensor with a wider distribution through the light guide, and the light generated from the bottom layer is incident on the photosensor with a narrower distribution than the top layer. When a flood image is reconstructed based on the signals obtained from the photosensor by different distributions, scintillation pixels are imaged at different positions for each layer. To verify this, a DETECT2000 simulation tool that simulates the behavior of light in a scintillator was used. By designing a scintillation pixel array, a detector consisting of a light guide and a photosensor, a gamma ray event was generated in all scintillation pixels to obtain a flood imgae. As a result, it was confirmed that the top and bottom layers were imaged at different positions and completely separated. When this detector is applied to PET, it is considered that image quality can be improved through imporved spatial resolution.
To assess the clinical value of time resolved imaging of contrast kinetics(TRICKS) MRA by comparison with conventional time of flight(TOF) MR angiography. Both TOF-MRA and TRICKS-MRA were performed in 17 patients with cerebrovascular disease and in 6 patients with brain tumor. Among 17 cerebraovascular patients, digital subtraction angiography(DSA) data were also obtained in 11 patients. TOF-MRA showed good spatial resolution but short in temporal resolution. Although TRICKS-MRA showed somewhat low spatial resolution, it showed superior temporal resolution by distinguishing vessel and tumor in all patients. Also, from the analysis of vessel-tumor relationship, TRICKS-MRA showed better performance than TOF-MRA. TRICKS-MRA makes it possible to image arterial, capillary and venous phase sequentially with very speedy manner and therefore, the clinical use of this method is highly suggestive for future use.
Our objective was to evaluate the clinical feasibility of spatial domain filtering as an alternative to additional image reconstruction using different kernels in CT. Kernels were grouped as H30 (head medium smooth), B30 (body medium smooth), S80 (special) and U95 (ultra sharp). Derived from thin coilimated source images, four sets of images were generated using phantom kernels. MTF (50%, 10%, 2%) measured with H30 (3.25, 5.68, 7.45 Ip/cm) B30 (3.84, 6.25, 7.72 Ip/cm), S80 (4.69, 9.49, 12.34 Ip/cm), and U95 (14.19, 20.31, 24.67 Ip/cm). Spatial resolution for the U95 kernel (0.6 mm) was 33.3% greater than that of the H30 and B30 (0.8 mm) kernels. Initially scanned kernels images were rated for subjective image qualify, using a five-point scale. Image scanned with a convolution kernel led to an increase in noise (U95), whereas the results for CT attenuation coefficient were comparable. CT images increase the diagnostic accuracy in head (H30), abdomen (B30), temporal bone and lung (U95) kernels may be controlled by adjusting CT various algorithms, which should be adjusted to take into account the kernels of the CT undergoing the examination.
Among imaging and treatment devices for small animals, positron emission tomography(PET) causes a change in spatial resolution within a field of view. This is a phenomenon caused by using a small gantry and a thin and long scintillation pixel, and detectors that measure the interaction depth are being developed and researched to solve this problem. In this study, a detector that measures the interaction depth was designed using several scintillator blocks and light guides with different reflector patterns. The scintillator block composed of 4 × 4 arrays of 3 mm × 3 mm × 5 mm scintillation pixels formed four layers, and a light guide was inserted in each layer to configure the entire detector. In order to check whether the interaction depth was measured, a gamma ray interaction was generated at the center of all scintillation pixels to acquire data and then reconstructed into a flood image. The reflector patterns of the light guides inserted between the layers were all different, so the positions of the scintillation pixels for each layer were formed in different locations. It is considered that even spatial resolution can be achieved over all regions of the field of view if all positions of the scintillation pixels thus formed are separated and used for image reconstruction.
근적외영역의 타이탄 측광영상에서 나타나는 주연감광(Limb Darkening) 및 주연증광(Limb Brightening) 현상의 원인을 설명하기 위해 구면기하(spherical geometry)에 의한 효과 외에 헤이즈(haze) 자체의 성질 변화량을 추정해 본다. 근적외영역의 강한 메탄 흡수 밴드 부근에서 타이탄의 대기를 관측하면 파장에 따라 표면 부근에서 성층권 사이의 정보를 얻을 수 있으며, 이를 복사전달모델과 비교해 대기 중에 존재하는 연무 입자의 성질을 유추할 수 있다. 관측자료는 2006년 2월 Gemini/North Observatory의 Near-Infrared Integral Field Spectrometer(NIFS)를 이용해 얻은 측 분광 자료를 사용하였다. NIFS의 공간분해능은 화소당 0.05"로, 타이탄의 적도부분을 약 17화소로 분해할 수 있으므로, 각 지역별로 분광선 모델을 만들어 관측자료와 비교하는 데에 용이하였다.
Jang, Eui Sun;Kwak, In Suk;Park, Sun Myung;Choi, Choon Ki;Lee, Hyuk;Kim, Soo Young;Choi, Sung Wook
The Korean Journal of Nuclear Medicine Technology
/
v.17
no.2
/
pp.67-71
/
2013
Purpose: The Change of CT exposure condition have a effect on image quality and patient exposure dose. In this study, we evaluated effect CT image quality and SUV when CT parameters (Pitch, Rotation time) were changed. Materials and Methods: Discovery Ste (GE, USA) was used as a PET/CT scanner. Using GE QA Phantom and AAPM CT Performance Phantom for evaluate Noise of CT image. Images are acquired by using 24 combinations that four stages pitch (0.562, 0.938, 1.375, 1.75:1) and six stages X-ray tube rotation time (0.5s-1.0s). PET images are acquired using 1994 NEMA PET Phantom ($^{18}F-FDG$ 5.3 kBq/mL, 2.5 min/frame). For noise test, noise are evaluated by standard deviation of each image's CT numbers. And then we used expectation noise according to change of DLP (Dose Length Product) to experimental noise ratio for index of effectiveness. For spatial resolution test, we confirmed that it is possible to identify to 1.0 mm size of the holes at the AAPM CT Performance Phantom. Finally we evaluated each 24 image's SUV. Results: Noise efficiency were 1.00, 1.03, 1.01, 0.96 and 1.00, 1.04, 1.02, 0.97 when pitch changes at the QA Phantom and AAPM Phantom. In case of X-ray tube rotation time changes, 0.99, 1.02, 1.00, 1.00, 0.99, 0.99 and 1.01, 1.01, 0.99, 1.01, 1.01, 1.01 at the QA Phantom and AAPM Phantom. We could identify 1.0 mm size of the holes all 24 images. Also, there were no significant change of SUV and all image's average SUV were 1.1. Conclusion: 1.75:1 pitch is the most effective value at the CT image evaluation according to pitch change and It doesn't affect to the spatial resolution and SUV. However, the change of rotation time doesn't affect anything. So, we recommend to use the effective pitch like 1.75:1 and adequate X-ray tube rotation time according to patient size.
Proceedings of the Korea Contents Association Conference
/
2009.05a
/
pp.1105-1110
/
2009
To assess the clinical value of time resolved imaging of contrast kinetics(TRICKS) MRA by comparison with conventional time of flight(TOF) MR angiography. Both TOF-MRA and TRICKS-MRA were performed in 17 patients with cerebrovascular disease and in 6 patients with brain tumor. Among 17 cerebraovascular patients, digital subtraction angiography(DSA) data were also obtained in 11 patients. TOF-MRA showed good spatial resolution but short in temporal resolution. Although TRICKS-MRA showed somewhat low spatial resolution, it showed superior temporal resolution by distinguishing vessel and tumor in all patients. Also, from the analysis of vessel-tumor relationship, TRICKS-MRA showed better performance than TOF-MRA. TRICKS-MRA makes it possible to image arterial, capillary and venous phase sequentially with very speedy manner and therefore, the clinical use of this method is highly suggestive for future use.
Purpose: SPECT/CT, a combination of SPECT and CT, is capable of expressing the results of attenuation correction on images biased by automatic program. As a result, this research evaluates the usefulness of images with CT attenuation correction, using various phantoms and images of patients. Materials and Methods: From July of 2012 to September of 2012, this research was conducted on the contrast, spatial resolution, and images of patients. We studied the contrast with IEC body phantom and Jaszczak phantom, while the spatial resolution was evaluated with NEMA triple line phantom. Further, a comparative study was carried out on the quality of the images, on the difference between the images before and after the CT attenuation correction. Results: Compared the differences between the contrast before and after the CT attenuation correction in IEC body phantom. The contrast was improved by 33.6% at minimum, 89.8% at maximum. In case of Jaszczak Phantom, the contrast was enhanced by 9.9% at minimum, 27.8% at maximum. In NEMA Triple line phantom, the resolution was raised by 4.5% in average: 4.4% in horizontal, 4.5% in vertical. In Anthropomorphic Torso Phantom, the perfusion score of the interior wall with the most severe attenuation was measured to be 29.4%. In the experiment carried out on myocardial perfusion SPECT/CT patients, 9% improvement was discovered in the interior wall, where the most dramatic attenuation occurred, after the CT attenuation correction. Conclusion: SPECT/CT proved its clinical usefulness by enabling the acquisition of images with enhanced contrast and spatial resolution compare to the ones resulted from SPECT.
본 웹사이트에 게시된 이메일 주소가 전자우편 수집 프로그램이나
그 밖의 기술적 장치를 이용하여 무단으로 수집되는 것을 거부하며,
이를 위반시 정보통신망법에 의해 형사 처벌됨을 유념하시기 바랍니다.
[게시일 2004년 10월 1일]
이용약관
제 1 장 총칙
제 1 조 (목적)
이 이용약관은 KoreaScience 홈페이지(이하 “당 사이트”)에서 제공하는 인터넷 서비스(이하 '서비스')의 가입조건 및 이용에 관한 제반 사항과 기타 필요한 사항을 구체적으로 규정함을 목적으로 합니다.
제 2 조 (용어의 정의)
① "이용자"라 함은 당 사이트에 접속하여 이 약관에 따라 당 사이트가 제공하는 서비스를 받는 회원 및 비회원을
말합니다.
② "회원"이라 함은 서비스를 이용하기 위하여 당 사이트에 개인정보를 제공하여 아이디(ID)와 비밀번호를 부여
받은 자를 말합니다.
③ "회원 아이디(ID)"라 함은 회원의 식별 및 서비스 이용을 위하여 자신이 선정한 문자 및 숫자의 조합을
말합니다.
④ "비밀번호(패스워드)"라 함은 회원이 자신의 비밀보호를 위하여 선정한 문자 및 숫자의 조합을 말합니다.
제 3 조 (이용약관의 효력 및 변경)
① 이 약관은 당 사이트에 게시하거나 기타의 방법으로 회원에게 공지함으로써 효력이 발생합니다.
② 당 사이트는 이 약관을 개정할 경우에 적용일자 및 개정사유를 명시하여 현행 약관과 함께 당 사이트의
초기화면에 그 적용일자 7일 이전부터 적용일자 전일까지 공지합니다. 다만, 회원에게 불리하게 약관내용을
변경하는 경우에는 최소한 30일 이상의 사전 유예기간을 두고 공지합니다. 이 경우 당 사이트는 개정 전
내용과 개정 후 내용을 명확하게 비교하여 이용자가 알기 쉽도록 표시합니다.
제 4 조(약관 외 준칙)
① 이 약관은 당 사이트가 제공하는 서비스에 관한 이용안내와 함께 적용됩니다.
② 이 약관에 명시되지 아니한 사항은 관계법령의 규정이 적용됩니다.
제 2 장 이용계약의 체결
제 5 조 (이용계약의 성립 등)
① 이용계약은 이용고객이 당 사이트가 정한 약관에 「동의합니다」를 선택하고, 당 사이트가 정한
온라인신청양식을 작성하여 서비스 이용을 신청한 후, 당 사이트가 이를 승낙함으로써 성립합니다.
② 제1항의 승낙은 당 사이트가 제공하는 과학기술정보검색, 맞춤정보, 서지정보 등 다른 서비스의 이용승낙을
포함합니다.
제 6 조 (회원가입)
서비스를 이용하고자 하는 고객은 당 사이트에서 정한 회원가입양식에 개인정보를 기재하여 가입을 하여야 합니다.
제 7 조 (개인정보의 보호 및 사용)
당 사이트는 관계법령이 정하는 바에 따라 회원 등록정보를 포함한 회원의 개인정보를 보호하기 위해 노력합니다. 회원 개인정보의 보호 및 사용에 대해서는 관련법령 및 당 사이트의 개인정보 보호정책이 적용됩니다.
제 8 조 (이용 신청의 승낙과 제한)
① 당 사이트는 제6조의 규정에 의한 이용신청고객에 대하여 서비스 이용을 승낙합니다.
② 당 사이트는 아래사항에 해당하는 경우에 대해서 승낙하지 아니 합니다.
- 이용계약 신청서의 내용을 허위로 기재한 경우
- 기타 규정한 제반사항을 위반하며 신청하는 경우
제 9 조 (회원 ID 부여 및 변경 등)
① 당 사이트는 이용고객에 대하여 약관에 정하는 바에 따라 자신이 선정한 회원 ID를 부여합니다.
② 회원 ID는 원칙적으로 변경이 불가하며 부득이한 사유로 인하여 변경 하고자 하는 경우에는 해당 ID를
해지하고 재가입해야 합니다.
③ 기타 회원 개인정보 관리 및 변경 등에 관한 사항은 서비스별 안내에 정하는 바에 의합니다.
제 3 장 계약 당사자의 의무
제 10 조 (KISTI의 의무)
① 당 사이트는 이용고객이 희망한 서비스 제공 개시일에 특별한 사정이 없는 한 서비스를 이용할 수 있도록
하여야 합니다.
② 당 사이트는 개인정보 보호를 위해 보안시스템을 구축하며 개인정보 보호정책을 공시하고 준수합니다.
③ 당 사이트는 회원으로부터 제기되는 의견이나 불만이 정당하다고 객관적으로 인정될 경우에는 적절한 절차를
거쳐 즉시 처리하여야 합니다. 다만, 즉시 처리가 곤란한 경우는 회원에게 그 사유와 처리일정을 통보하여야
합니다.
제 11 조 (회원의 의무)
① 이용자는 회원가입 신청 또는 회원정보 변경 시 실명으로 모든 사항을 사실에 근거하여 작성하여야 하며,
허위 또는 타인의 정보를 등록할 경우 일체의 권리를 주장할 수 없습니다.
② 당 사이트가 관계법령 및 개인정보 보호정책에 의거하여 그 책임을 지는 경우를 제외하고 회원에게 부여된
ID의 비밀번호 관리소홀, 부정사용에 의하여 발생하는 모든 결과에 대한 책임은 회원에게 있습니다.
③ 회원은 당 사이트 및 제 3자의 지적 재산권을 침해해서는 안 됩니다.
제 4 장 서비스의 이용
제 12 조 (서비스 이용 시간)
① 서비스 이용은 당 사이트의 업무상 또는 기술상 특별한 지장이 없는 한 연중무휴, 1일 24시간 운영을
원칙으로 합니다. 단, 당 사이트는 시스템 정기점검, 증설 및 교체를 위해 당 사이트가 정한 날이나 시간에
서비스를 일시 중단할 수 있으며, 예정되어 있는 작업으로 인한 서비스 일시중단은 당 사이트 홈페이지를
통해 사전에 공지합니다.
② 당 사이트는 서비스를 특정범위로 분할하여 각 범위별로 이용가능시간을 별도로 지정할 수 있습니다. 다만
이 경우 그 내용을 공지합니다.
제 13 조 (홈페이지 저작권)
① NDSL에서 제공하는 모든 저작물의 저작권은 원저작자에게 있으며, KISTI는 복제/배포/전송권을 확보하고
있습니다.
② NDSL에서 제공하는 콘텐츠를 상업적 및 기타 영리목적으로 복제/배포/전송할 경우 사전에 KISTI의 허락을
받아야 합니다.
③ NDSL에서 제공하는 콘텐츠를 보도, 비평, 교육, 연구 등을 위하여 정당한 범위 안에서 공정한 관행에
합치되게 인용할 수 있습니다.
④ NDSL에서 제공하는 콘텐츠를 무단 복제, 전송, 배포 기타 저작권법에 위반되는 방법으로 이용할 경우
저작권법 제136조에 따라 5년 이하의 징역 또는 5천만 원 이하의 벌금에 처해질 수 있습니다.
제 14 조 (유료서비스)
① 당 사이트 및 협력기관이 정한 유료서비스(원문복사 등)는 별도로 정해진 바에 따르며, 변경사항은 시행 전에
당 사이트 홈페이지를 통하여 회원에게 공지합니다.
② 유료서비스를 이용하려는 회원은 정해진 요금체계에 따라 요금을 납부해야 합니다.
제 5 장 계약 해지 및 이용 제한
제 15 조 (계약 해지)
회원이 이용계약을 해지하고자 하는 때에는 [가입해지] 메뉴를 이용해 직접 해지해야 합니다.
제 16 조 (서비스 이용제한)
① 당 사이트는 회원이 서비스 이용내용에 있어서 본 약관 제 11조 내용을 위반하거나, 다음 각 호에 해당하는
경우 서비스 이용을 제한할 수 있습니다.
- 2년 이상 서비스를 이용한 적이 없는 경우
- 기타 정상적인 서비스 운영에 방해가 될 경우
② 상기 이용제한 규정에 따라 서비스를 이용하는 회원에게 서비스 이용에 대하여 별도 공지 없이 서비스 이용의
일시정지, 이용계약 해지 할 수 있습니다.
제 17 조 (전자우편주소 수집 금지)
회원은 전자우편주소 추출기 등을 이용하여 전자우편주소를 수집 또는 제3자에게 제공할 수 없습니다.
제 6 장 손해배상 및 기타사항
제 18 조 (손해배상)
당 사이트는 무료로 제공되는 서비스와 관련하여 회원에게 어떠한 손해가 발생하더라도 당 사이트가 고의 또는 과실로 인한 손해발생을 제외하고는 이에 대하여 책임을 부담하지 아니합니다.
제 19 조 (관할 법원)
서비스 이용으로 발생한 분쟁에 대해 소송이 제기되는 경우 민사 소송법상의 관할 법원에 제기합니다.
[부 칙]
1. (시행일) 이 약관은 2016년 9월 5일부터 적용되며, 종전 약관은 본 약관으로 대체되며, 개정된 약관의 적용일 이전 가입자도 개정된 약관의 적용을 받습니다.