Performance Characteristics of 3D GSO PET/CT Scanner (Philips GEMINI PET/DT)  

Kim, Jin-Su (Department of Nuclear Medicine, Interdisciplinary Program in Radiation Applied Life Science Major, Seoul National University College of Medicine)
Lee, Jae-Sung (Department of Nuclear Medicine, Interdisciplinary Program in Radiation Applied Life Science Major, Seoul National University College of Medicine)
Lee, Byeong-Il (Department of Nuclear Medicine, Interdisciplinary Program in Radiation Applied Life Science Major, Seoul National University College of Medicine)
Lee, Dong-Soo (Department of Nuclear Medicine, Interdisciplinary Program in Radiation Applied Life Science Major, Seoul National University College of Medicine)
Chung, June-Key (Department of Nuclear Medicine, Interdisciplinary Program in Radiation Applied Life Science Major, Seoul National University College of Medicine)
Lee, Myung-Chul (Department of Nuclear Medicine, Interdisciplinary Program in Radiation Applied Life Science Major, Seoul National University College of Medicine)
Publication Information
The Korean Journal of Nuclear Medicine / v.38, no.4, 2004 , pp. 318-324 More about this Journal
Abstract
Purpose: Philips GEMINI is a newly introduced whole-body GSO PET/CT scanner. In this study, performance of the scanner including spatial resolution, sensitivity, scatter fraction, noise equivalent count ratio (NECR) was measured utilizing NEMA NU2-2001 standard protocol and compared with performance of LSO, BGO crystal scanner. Methods: GEMINI is composed of the Philips ALLEGRO PET and MX8000 D multi-slice CT scanners. The PET scanner has 28 detector segments which have an array of 29 by 22 GSO crystals ($4{\times}6{\times}20$ mm), covering axial FOV of 18 cm. PET data to measure spatial resolution, sensitivity, scatter fraction, and NECR were acquired in 3D mode according to the NEMA NU2 protocols (coincidence window: 8 ns, energy window: $409[\sim}664$ keV). For the measurement of spatial resolution, images were reconstructed with FBP using ramp filter and an iterative reconstruction algorithm, 3D RAMLA. Data for sensitivity measurement were acquired using NEMA sensitivity phantom filled with F-18 solution and surrounded by $1{\sim}5$ aluminum sleeves after we confirmed that dead time loss did not exceed 1%. To measure NECR and scatter fraction, 1110 MBq of F-18 solution was injected into a NEMA scatter phantom with a length of 70 cm and dynamic scan with 20-min frame duration was acquired for 7 half-lives. Oblique sinograms were collapsed into transaxial slices using single slice rebinning method, and true to background (scatter+random) ratio for each slice and frame was estimated. Scatter fraction was determined by averaging the true to background ratio of last 3 frames in which the dead time loss was below 1%. Results: Transverse and axial resolutions at 1cm radius were (1) 5.3 and 6.5 mm (FBP), (2) 5.1 and 5.9 mm (3D RAMLA). Transverse radial, transverse tangential, and axial resolution at 10 cm were (1) 5.7, 5.7, and 7.0 mm (FBP), (2) 5.4, 5.4, and 6.4 mm (3D RAMLA). Attenuation free values of sensitivity were 3,620 counts/sec/MBq at the center of transaxial FOV and 4,324 counts/sec/MBq at 10 cm offset from the center. Scatter fraction was 40.6%, and peak true count rate and NECR were 88.9 kcps @ 12.9 kBq/mL and 34.3 kcps @ 8.84 kBq/mL. These characteristics are better than that of ECAT EXACT PET scanner with BGO crystal. Conclusion: The results of this field test demonstrate high resolution, sensitivity and count rate performance of the 3D PET/CT scanner with GSO crystal. The data provided here will be useful for the comparative study with other 3D PET/CT scanners using BGO or LSO crystals.
Keywords
NEMA NU2-2001; GSO; PET/CT;
Citations & Related Records
연도 인용수 순위
  • Reference
1 American College of Radiology. ACR technical standard for medical nuclear physics performance monitoring of PET imaging equipment. 2001
2 Casey ME, Young J, Wheelock T, Schmand M, Bendriem B, Nutt R, et al. Physical performance of a high resolution PET/CT scanner. J Nucl Med 2004;45(suppl):100 [abstract]
3 Daube-Witherspoon ME, Matej S, Karp JS, Lewitt RM. Application of the row action maximum likelihood algorithm with spherical basis functions to clinical PET imaging. IEEE Trans Nucl Sci 2001;48:24-30   DOI   ScienceOn
4 Yang Y, Tai YC, Siegel S, Newport DF, Bai B, Li Q, et al. Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging. Phys Med Biol 2004;49:2527-45   DOI   ScienceOn
5 Cherry SR, Sorenson JA, Phelps ME. Physics in nuclear medicine. 3rd ed. Philadelphia: Elsevier Science; 2003. p. 137
6 Surti S, Karp JS. Imaging characteristics of a 3-dimensional GSO whole-body PET camera. J Nucl Med 2004;45:1040-9   PUBMED
7 Kim JS, Lee JS, Lee DS, Chung J-K, Lee MC. Performance evaluation of Philips GEMINI PET/CT using NEMA NU2-2001 standard. J Nucl Med 2004;45(suppl):101 [abstract]
8 Bailey DL. Quantitative procedures in 3D PET. In: Bendriem B, Townsend DW, editors. The theory and practice of 3D PET. Dordrecht: Kluwer Academic Publishers; 1998. p. 118-9
9 Laymon CM, Bowsher JE. Performance of the log likelihood function for discriminating artifactual from artifact free PET atteunatioin correction images. J Nucl Med 2004;45(suppl):160 [abstract]
10 Cherry SR, Shao Y, Silverman RW, Meadors K, Siegel S, Chatziioannou A et al. MicroPET: a high resolution PET scanner for imaging small animals. IEEE Trans Nucl Sci 1997;44:1161-6   DOI   ScienceOn
11 Cherry SR, Sorenson JA, Phelps ME. Physics in nuclear medicine. 3rd ed. Philadelphia: Elsevier Science; 2003. p. 104
12 Karp JS, Surti S, Daube-Witherspoon ME, Freifelder R, Cardi CA, Adam LE, et al. Performance of a brain PET camera based on anger-logic gadolinium oxyorthosilicate detectors. J Nucl Med 2003;44:1340-9   PUBMED
13 Daube-Witherspoon ME, Karp JS, Casey ME, DiFilippo FP, Hines H, Muehllehner G, et al. PET performance measurements using the NEMA NU 2-2001 standard. J Nucl Med 2002;43:1398-409   PUBMED
14 National Electrical Manufacturers Association: NEMA standards publication NU2-2001: performance measurements of positron emission tomographs. Rosslyn, VA, National Electrical Manufacturers Association, 2001
15 Barnes D, Egan G, O'Keefe G, Abbott D. Characterization of dynamic 3-D PET imaging for functional brain mapping. IEEE Trans Med Imaging 1997;16:261-9   DOI   ScienceOn
16 Kim JS, Lee JS, Lee DS, Chung J-K, Lee MC. Performance evaluation of Siemens CTI ECAT EXACT 47 scanner using NEMA NU2-2001. Korean J Nucl Med 2004;38:259-67
17 Lodge MA, Dilsizian V, Line BR. Performance assessment of the Philips GEMINI PET/CT Scanner. J Nucl Med 2004;45(suppl):425[abstract]
18 Browne J, de Pierro AB. A row-action alternative to the EM algorithm for maximizing likelihood in emission tomography. IEEE Trans Med Imaging 1996;15:687-99   DOI   ScienceOn
19 Tai C, Chatziioannou A, Siegel S, Young J, Newport D, Goble RN, et al. Performance evaluation of the microPET P4: a PET system dedicated to animal imaging. Phys Med Biol 2001;46:1845-62   DOI   ScienceOn