• Title/Summary/Keyword: 양전자방출단층촬영기기

Search Result 19, Processing Time 0.018 seconds

A study on image registration and fusion of MRI and SPECT/PET (뇌의 단일 광자 방출 전산화 단층촬영 영상, 양전자 방출 단층 촬영 영상 그리고 핵자기공명 영상의 융합과 등록에 관한 연구)

  • Joo, Ra-Hyung;Choi, Yong;Kwon, Soo-Il;Heo, Soo-Jin
    • Progress in Medical Physics
    • /
    • v.9 no.1
    • /
    • pp.47-53
    • /
    • 1998
  • Nuclear Medicine Images have comparatively poor spatial resolution, making it difficult to relate the functional information which they contain to precise anatomical structures. Anatomical structures useful in the interpretation of SPECT /PET Images were radiolabelled. PET/SPECT Images Provide functional information, whereas MRI mainly demonstrate morphology and anatomical. Fusion or Image Registration improves the information obtained by correlating images from various modalities. Brain Scan were studied on one or more occations using MRI and SPECT. The data were aligned using a point pair methods and surface matching. SPECT and MR Images was tested using a three dimensional water fillable Hoffman Brain Phantom with small marker and PET and MR Image was tested using a patient data. Registration of SPECT and MR Images is feasible and allows more accurate anatomic assessment of sites of abnormal uptake in radiolabeled studies. Point based registration was accurate and easily implemented three dimensional registration of multimodality data set for fusion of clinical anatomic and functional imaging modalities. Accuracy of a surface matching algorithm and homologous feature pair matching for three dimensional image registration of Single Photon Emission Computed Tomography Emission Computed Tomography (SPECT), Positron Emission Tomography (PET) and Magnetic Resonance Images(MRD was tested using a three dimensional water fill able brain phantom and Patients data. Transformation parameter for translation and scaling were determined by homologous feature point pair to match each SPECT and PET scan with MR images.

  • PDF

Digital Position Acquisition Method of PET Detector Module using Maximum Likelihood Position Estimation (최대우도함수를 이용한 양전자방출단층촬영기기의 검출기 모듈의 디지털 위치 획득 방법)

  • Lee, Seung-Jae;Baek, Cheol-Ha
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.1
    • /
    • pp.1-7
    • /
    • 2021
  • In order to acquire an image in a positron emission tomography, it is necessary to draw the position coordinates of the scintillation pixels of the detector module measured at the same time. To this end, in a detector module using a plurality of scintillation pixels and a small number of photosensors, it is necessary to obtain a flood image and divide a region of each scintillation pixel to obtain a position of a scintillation pixel interacting with a gamma ray. Alternatively, when the number of scintillation pixels and the number of photosensors to be used are the same, the position coordinates for the position of the scintillation pixels can be directly acquired as digital signal coordinates. A method of using a plurality of scintillation pixels and a small number of photosensors requires a process of obtaining digital signal coordinates requires a plurality of photosensors and a signal processing system. This complicates the signal processing process and raises the cost. To solve this problem, in this study, we developed a method of obtaining digital signal coordinates without performing the process of separating the planar image and region using a plurality of flash pixels and a small number of optical sensors. This is a method of obtaining the position coordinate values of the flash pixels interacting with the gamma ray as a digital signal through a look-up table created through the signals acquired from each flash pixel using the maximum likelihood function. Simulation was performed using DETECT2000, and verification was performed on the proposed method. As a result, accurate digital signal coordinates could be obtained from all the flash pixels, and if this is applied to the existing system, it is considered that faster image acquisition is possible by simplifying the signal processing process.

Two Layer DOI Detector Design for PET using Multiple Light Guides for Designing Light Distribution (다수의 광가이드를 통한 빛 분포 설계로 두 층의 반응 깊이를 측정하는 양전자방출단층촬영기기의 검출기 설계)

  • Seung-Jae, Lee
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.6
    • /
    • pp.703-708
    • /
    • 2022
  • A detector module measuring the depth of interaction(DOI) was designed to improve the spatial resolution of positron emission tomography(PET). The scintillation pixel array consists of two layers, and a light guide is inserted between the layers to make the light generated through the gamma-ray event different for each layer. There are four light guides, and one light guide is designed to be coupled to a 2 × 2 array of scintillation pixels. The light generated from the top layer is moved to the photosensor with a wider distribution through the light guide, and the light generated from the bottom layer is incident on the photosensor with a narrower distribution than the top layer. When a flood image is reconstructed based on the signals obtained from the photosensor by different distributions, scintillation pixels are imaged at different positions for each layer. To verify this, a DETECT2000 simulation tool that simulates the behavior of light in a scintillator was used. By designing a scintillation pixel array, a detector consisting of a light guide and a photosensor, a gamma ray event was generated in all scintillation pixels to obtain a flood imgae. As a result, it was confirmed that the top and bottom layers were imaged at different positions and completely separated. When this detector is applied to PET, it is considered that image quality can be improved through imporved spatial resolution.

Image Acquisition Study of Maximal Scintillation Pixel Array using Light Guide (광가이드를 사용한 최대 섬광 픽셀 배열의 영상 획득 연구)

  • Lee, Seung-Jae
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.3
    • /
    • pp.249-255
    • /
    • 2022
  • Positron emission tomography for small animals has very high spatial resolution for imaging very small organs. To achieve good spatial resolution, the system must be constructed using very small scintillation pixels. When a detector is constructed using very small scintillation pixels, the size of the applicable array varies depending on the photosensor pixel. In a previous study, a study was conducted to find the maximum scintillation pixel arrangement according to the size of the photosensor. In this study, a detector with a light guide was designed to configure the detector using a more extended array of scintillation pixels, and try to find the maximum arrangement in which all scintillation pixels are imaged. The detector was designed using DETECT2000, which can simulate a detector made of a scintillator. Simulations were performed by configuring the detectors from an 11 × 11 scintillation pixel array to a 16 × 16 array. After obtaining a flood image by collecting the light generated from the scintillation pixel with a photosensor, the largest arrangement without overlap was found through image analysis. As a result, the largest arrangement in which all scintillation pixels could be distinguished without overlapping was a 15 × 15 arrangement.

A Study on the Maximization of Scintillation Pixel Array According to the Size of the Photosensor (광센서 크기에 따른 섬광 픽셀 배열의 최대화 연구)

  • Lee, Seung-Jae
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.2
    • /
    • pp.157-162
    • /
    • 2022
  • Since preclinical positron emission tomography imaging is performed on small animals that are very small compared to the human body, a detector with excellent spatial resolution is required. For this purpose, a system was constructed using a detector using small scintillation pixels. Since the size of the currently developed and used photosensors is limited, excellent spatial resolution can be obtained when the minimum scintillation pixel and maximum array are used. In this study, the size of the photosensor is fixed and various scintillation pixel arrays are configured to match the size of the scintillation pixels, so that no overlap occurs in the flood image and the maximum scintillation pixel array in which all scintillation pixels are distinguished. For this purpose, DETECT2000, which can simulate a detector module composed of a scintillator and an photosensor, was used. A photosensor consisting of a 4 × 4 array of 3 mm × 3 mm pixels was used, and the scintillation pixel array was configured from 8 × 8 to 13 × 13, and simulations were performed. A flood image was constructed using the data obtained from the photosensor pixel, and the maximum scintillation pixel array that does not overlap the image was found through the flood image and the profile. As a result, the size of the scintillation pixel array in which all scintillation pixels are imaged without overlapping each other in the flood image was 11 × 11.

Design of a Depth Encoding Detector using Light Guides with Different Reflector Patterns for Each Layer (각 층별 반사체 패턴이 서로 다른 광가이드를 사용한 반응 깊이 측정 검출기 설계)

  • Seung-Jae, Lee
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.1
    • /
    • pp.31-36
    • /
    • 2023
  • Among imaging and treatment devices for small animals, positron emission tomography(PET) causes a change in spatial resolution within a field of view. This is a phenomenon caused by using a small gantry and a thin and long scintillation pixel, and detectors that measure the interaction depth are being developed and researched to solve this problem. In this study, a detector that measures the interaction depth was designed using several scintillator blocks and light guides with different reflector patterns. The scintillator block composed of 4 × 4 arrays of 3 mm × 3 mm × 5 mm scintillation pixels formed four layers, and a light guide was inserted in each layer to configure the entire detector. In order to check whether the interaction depth was measured, a gamma ray interaction was generated at the center of all scintillation pixels to acquire data and then reconstructed into a flood image. The reflector patterns of the light guides inserted between the layers were all different, so the positions of the scintillation pixels for each layer were formed in different locations. It is considered that even spatial resolution can be achieved over all regions of the field of view if all positions of the scintillation pixels thus formed are separated and used for image reconstruction.

Development of PET Detector Module Measuring DOI using Multiple Reflectors (여러 반사체를 사용한 양전자방출단층촬영기기의 반응 깊이 측정 검출기 모듈 개발)

  • Kim, Neung Gyun;Kim, Gu;Kwak, Jong Hyeok;Lee, Seung-Jae
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.6
    • /
    • pp.825-830
    • /
    • 2019
  • A detector module measuring a depth of interaction was developed using silicon photomultiplier (SiPM) and two layers of scintillation crystal array treated with multiple reflectors. When reconstructing an image based on a signal obtained by using different types of reflector of each layer, the interaction positions of scintillation pixels and gamma rays could be tracked by utilizing the feature that all scintillation pixels were recorded at different positions. The bottom layer uses a specular reflector, and the top layer uses a diffuse reflector to differently process the size of the signal obtained from the SiPM. The optical grease was used to recude the sharp refractive index change between the layers of scintillator and the SiPM. The signals obtained from the 16 SiPMs were reduced to four signals using the Anger equations, and the images were reconstructed using them. All the scintillation pixels composed of the two layers appeared in the reconstructed image, which distinguished the layer where the scintillation pixels and gamma rays interacted. If the detectors, which measure the interaction depth of two layers using different reflectors, will be applied to preclinical positron emission tomography, the degradation of spatial resolution appearing outside the field of interest could be solved.

PET System Design using a Scintillator with a Size of 0.8 mm to Improve Spatial Resolution (공간분해능 향상을 위한 0.8 mm 크기의 섬광체를 사용한 PET 시스템 설계)

  • Lee, Seung-Jae
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.5
    • /
    • pp.499-504
    • /
    • 2022
  • Positron emission tomography (PET) uses a very small scintillator to achieve exellent spatial resolution. Therefore, in this study, a PET system using a scintillator to 0.8 mm size was designed and the performance was evaluated. Anihilation radiation was generated from the center of the field of view (FOV) to the outskirts at intervals of 10 mm, and counted simultaneously. The image was reconstructed using the coincidence data, and the spatial resolution was calculated by acquiring the full width at half maximum through the profile. The spatial resolution at the center of the FOV was 1.02 mm, showing a very good result, and the spatial resolution decreased as it was located at the outer edge. To evaluate the phantom image, the Derenzo phantom was constructed to acquire the image, and the degree of classification between radiation sources was evaluated through profile analysis. The result showed that the distance between the radiation sources was larger than the spatial resolution of the radiation sources at each location, and it was confirmed that the radiation sources were distinguished through this. When the PET system designed in this study is applied to PET for small animals, it is considered that excellent performance can be secured through the characteristic of very good spatial resolution.

DOI Detector Design using Different Sized Scintillators in Each Layer (각 층의 서로 다른 크기의 섬광체를 사용한 반응 깊이 측정 검출기 설계)

  • Seung-Jae, Lee
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.1
    • /
    • pp.11-16
    • /
    • 2023
  • In preclinical positron emisson tomography(PET), spatial resolution degradation occurs outside the field of view(FOV). To solve this problem, a depth of interaction(DOI) detector was developed that measures the position where gamma rays and the scintillator interact. There are a method in which a scintillation pixel array is composed of multiple layers, a method in which photosensors are arranged at both ends of a single layer, a method in which a scintillation pixel array is constituted in several layers and a photosensor is arranged in each layer. In this study, a new type of DOI detector was designed by analyzing the characteristics of the previously developed detectors. In the two-layer detector, different sizes of scintillation pixels were used for each layer, and the array size was configured differently. When configured in this form, the positions of the scintillation pixels for each layer are arranged to be shifted from each other, so that they are imaged at different positions in a flood image. DETECT2000 simulation was performed to confirm the possibility of measuring the depth of interaction of the designed detector. A flood image was reconstructed from a light signal acquired by a gamma-ray event generated at the center of each scintillation pixel. As a result, it was confirmed that all scintillation pixels for each layer were separated from the reconstructed flood image and imaged to measure the interaction depth. When this detector is applied to preclinical PET, it is considered that excellent images can be obtained by improving spatial resolution.

Design of Small-sized Scintillation Pixel Detector with a Light Guide made of the Same Material as the Scintillation Pixel (섬광 픽셀과 동일한 물질로 광가이드를 적용한 매우 작은 섬광 픽셀 검출기 설계)

  • Seung-Jae Lee;Byungdu Jo
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.4
    • /
    • pp.523-529
    • /
    • 2023
  • In order to achieve excellent spatial resolution, very small scintillation pixels are used in detectors of positron emission tomography for small animals. However, by using these very small scintillation pixels, scintillation pixels at the edge of the array may overlap in a flood image. To solve this problem, a light guide capable of changing the distribution of light was used. Depending on the material of the light guide, the light spreading tendency is different, and accordingly, the presence or absence of overlapping is different depending on the material of the light guide used. In this study, instead of the conventional glass light guide, a detector using the same material as the scintillation pixel was designed. A scintillator light guide has a higher refractive index than a glass light guide, so the light spread is different. Flood images were acquired to evaluate the degree of separation of the scintillation pixels at the edge of the detector using the two light guides. The degree of separation was evaluated by calculating the distance between the center and the spatial resolution of the image of two scintillation pixels at the edge of the obtained flood image. As a result, when the scintillator light guide was used, better spatial resolution was shown, and the distance between centers of scintillation pixels was wider. When a detector is constructed using a scintillator light guide instead of a conventional glass light guide, it is possible to use a smaller scintillation pixel, thereby securing better spatial resolution.