Air Density Correction of Ionization Chamber using $^{90}Sr$ Radioactive Check Device ($^{90}Sr$ 방사성 동위원소를 이용한 전리함의 대기 보정계수 측정)

  • Park, Sung-Y.;Kim, Woo-C.;Shin, Dong-O.;Ji, Young-H.;Kwon, Soo-I.;Lee, Kil-D.;Cho, Young-K.;Loh, John-J.
    • Journal of Radiation Protection and Research
    • /
    • v.23 no.4
    • /
    • pp.267-271
    • /
    • 1998
  • It is required to measure air density correction factor at the time of absorbed dose calibration or measurement. In general, thermometer and barometer are widely used for air density correction. However, this can be done using the radioactive check device with better accuracy. The measurements of air density correction were performed by using the radioactive check device, Unidos electrometer, and 0.6 cc Farmer-type ion chamber of PTW under the different environmental conditions. Above experiments were repeated with thermometer and barometer. By comparing the two methods, they were within the difference of 0.2 %. The overall uncertainty for the dose found in thermometer and barometer was 1.2 - 1.6 %, depending upon either one step or two, whereas the overall uncertainty for the radioactive check device was 1.02 %. This method may reduce the possible error which could occur when thermometer and barometer are not calibrated at regular basis.

  • PDF

A Theoretical Calculation for Angular Dependence of X-ray Beams on Extremity Phantom (말단팬텀에서 X-선 빔의 방향의존성에 관한 이론적 계산)

  • Kim, Jong-Soo;Yoon, Suk-Chul;Kim, Jang-Lyul;Kim, Kwang-Pyo
    • Journal of Radiation Protection and Research
    • /
    • v.21 no.4
    • /
    • pp.263-271
    • /
    • 1996
  • The ANSI N13.32 recommends that a study of the angular response of a dosimeter be carried out once, although no pass/fail criterion is given for angular response. Gamma dose equivalent conversion and angular dependence factors were calculated by using MCNP code for the case of ANSI N13.32 extremity phantoms(finger and arm) at the depth of $7mg/cm^2$. Those extremity dosimeters were assumed to be irradiated from both monoenergitic photons and ISO X-ray narrow beams. These calculated gamma dose equivalent conversion and angular dependence factors were compared to B. Grosswendt's result calculated by using X-ray beams. The result showed that the dose equivalent conversion factors of this study agreed well with that of B. Grosswendt for all energies within 2% except 7% in the case of the low energies. In the case of angular dependence factors comparison, they agreed within 3%. It was shown that angular dependence factors of the finger phantom decreased as the horizontal angle of the phantom increased for the ISO X-ray beams less than 60keV. For the higher energy X-ray beams range they decreased slightly around 40 degree, but then increased from this energy to 90 degree.

  • PDF

Establishment of the Monoenergetic Fluorescent X-ray Radiation Fields (교정용 단일에너지 형광 X-선장의 제작)

  • Kim, Jang-Lyul;Kim, Bong-Hwan;Chang, Si-Young;Lee, Jae-Ki
    • Journal of Radiation Protection and Research
    • /
    • v.23 no.1
    • /
    • pp.33-47
    • /
    • 1998
  • Using a combination of an X-ray generator Installed in radiation calibration laboratory of Korea Atomic Energy Research Institute (KAERI) and a series of 8 radiators and filters described in ISO-4037, monoenergetic fluorescent X-rays from 8.6 keV to 75 keV were produced. This fluorescent X-rays generated by primary X-rays from radiator were discriminated $K_{\beta}$ lines with the aid of filter material and the only $K_{\alpha}$ X-rays were analyzed with the high purity Ge detector and portable MCA. The air kerma rates were measured with the 35 co ionization chamber and compared with the calculational results, and the beam uniformity and the scattered effects of radiation fields were also measured. The beam purities were more than 90 % for the energy range of 8.6 keV to 75 keV and the air kerma rates were from 1.91 mGy/h (radiator : Au, filter : W) to 54.2 mGy (radiator : Mo, filter : Zr) at 43 cm from center of the radiator. The effective area of beam at the measurement point of air kerma rates was 12 cm ${\times}$ 12 cm and the influence of scattered radiation was less than 3 %. The fluorescent X-rays established in this study could be used for the determination of energy response of the radiation measurement devices and the personal dosemeters in low photon energy regions.

  • PDF

A Study on the Characteristics of Smartphone Camera as a Medical Radiation Detector (의료 방사선 검출기로써 스마트폰 카메라의 특성에 관한 연구)

  • Kang, Han Gyu;Kim, Ho Chul
    • Journal of the Institute of Electronics and Information Engineers
    • /
    • v.53 no.5
    • /
    • pp.143-151
    • /
    • 2016
  • The aim of this study is to investigate the optimal algorithm to extract medical radiation induced pixel signal from complementary metal-oxide semiconductor (CMOS) sensors of smartphones camera. The pixel intensity and pixel number of smartphone camera were measured as the X-ray dose was increased. The front camera of the smartphone camera has low noise property and excellent dose response as compared to the back camera of the smartphone. The indirect method which uses scintillation crystal in front of the smartphone camera, couldn't improve the X-ray detection efficiency as compared to the direct method which does not use any scintillator in front of the smartphone camera. When we used the algorithm which employing threshold level on the pixel intensity and pixel number, the dose linearity was more higher for the pixel intensity rather for the pixel number. The use of pixel intensity of Y color component which represents the grey scale, would be efficient in terms of the radiation detection efficiency and reducing the complexity of the image processing. We expect that the radiation dose monitoring can be managed effectively and systematically by using the proposed radiation detection algorithm, thus eventually will contribute to the public healthcare.

MU Fluence Reconstruction based-on Delivered Leaf Position: for IMRT Quality Assurance (세기조절방사선치료의 정도관리를 위한 모니터유닛 공간분포 재구성의 효용성 평가)

  • Park, So-Yeon;Park, Yang-Kyun;Park, Jong-Min;Choi, Chang-Heon;Ye, Sung-Joon
    • Journal of Radiation Protection and Research
    • /
    • v.36 no.1
    • /
    • pp.28-34
    • /
    • 2011
  • The measurement-based verification for intensity modulated radiation therapy (IMRT) is a time-and labor-consuming procedure. Instead, this study aims to develop a MU fluence reconstruction method for IMRT QA. Total actual fluences from treatment planning system (TPS, Eclipse 8.6, Varian) were selected as a reference. Delivered leaf positions according to MU were extracted by the dynalog file generated after IMRT delivery. An in-house software was develop to reconstruct MU fluence from the acquired delivered leaf position data using MATLAB. We investigated five patient's plans delivered by both step-and-shoot IMRT and sliding window technologies. The total actual fluence was compared with the MU fluence reconstructed by using commercial software (Verisoft 3.1, PTW) and gamma analysis method (criteria: 3%/3 mm and 2%/1 mm). Gamma pass rates were $97.8{\pm}1.33$% and the reconstructed fluence was shown good agreement with RTP-based actual fluence. The fluence from step and shoot IMRT was shown slightly higher agreement with the actual fluence than that from sliding window IMRT. If moving from IMRT QA measurements toward independent computer calculations, the developed method can be used for IMRT QA. A point dose calculation method from reconstructed fluences is under development for the routine IMRT QA purpose.

Development of a cylindrical ultrasound applicator for Intracavitary Hyperthermia (강내온열 치료를 위한 원통형 초음파 치료기 개발)

  • Lee, Rena J.;Hyunsuk Suh
    • Progress in Medical Physics
    • /
    • v.13 no.2
    • /
    • pp.79-84
    • /
    • 2002
  • In this study, a cylindrical ultrasound applicator is developed for the treatment of vagina and rectum in combination with high dose rate brachytherapy. A cylindrical transducer (PZT-8, 1=1.5 cm, thickness=1.5mm OD=2.5 cm) was used as an energy source for induction of hyperthermia. Three single-element applicators were constructed to examine the performance of the PZT material. Vector impedance was measured to determine driving frequency. The efficiencies of the elements were determined using a radiation force technique to evaluate the feasibility of using the applicator as a hyperthermia source. A multi-element ultrasound applicator was designed using the PZT-8 material for the treatment of vagina. Results from the vector impedance measurements showed maximum magnitude at 1.78, 1.77, and 1.77 MHz for applicator 1,2, and 3, respectively. The radiation force measurements showed that the acoustic power of 40 watts was obtained in all three elements. The average efficiencies of the elements were 61.4, 65.2, and 54.0% for element 1, 2, and 3, respectively. The designed ultrasound hyperthermia applicator could be used in combination with high dose rate brachytherapy for the treatment of vagina and rectum. The use of this applicator with intracavitary brachytherapy could offer improved tumor control by increasing radiosensitiyity of the tumor.

  • PDF

Study for Valuing of Residential Water when Drought Conditions (가뭄상황시 생활용수 가치 추정)

  • Park, Doo-Ho;Park, Yoon-Shin
    • Proceedings of the Korea Water Resources Association Conference
    • /
    • 2007.05a
    • /
    • pp.487-492
    • /
    • 2007
  • 최근 잦은 기상이변으로 인한 자연재해는 그 빈도와 강도가 점차 증가하고 있다. 이러한 기상이변은 결국, 극단적인 가뭄이 닥칠 가능성이 농후할 것이고, 만일 극단적인 가뭄상황이 도래할 경우 제한급수 또는 절수로 인한 일반인이 느끼는 피해정도는 매우 늘어날 것이다. 본 연구는 가뭄상황시 제한급수를 고려한 소비자의 지불의사를 알아보고자, 7대 특 광역시와 32개 가뭄상습지역에 거주하는 일반인을 대상으로 생활용수 수요패턴과 가뭄상황을 고려한 생활용수의 가치를 추정하였다. 조사결과, 수돗물 수질 관련에 대해서는 가뭄상습지역보다 7대광역시가 더 민감하였고, 수돗물 수량에 대해서는 가뭄상습지역이 더 민감한 것으로 나타났다, 지불의사 추정은 양분선택형과 개방형을 혼합하여 지불의사액을 설문하였고, Tobit모형을 이용하여 지불의사액을 추정하였다. 추정결과, 교육수준이 높을 수록, 가구수가 많을수록, 소득이 높을수록 지불의사가 높은 것으로 나타났다. 또한, 가뭄으로 인해 생활용수의 25% 제한시에는 약 2010원, 50% 제한시에는 2,395원의 추가적인 지불의사가 있는 것으로 추정되었다. 결국, 조사가구의 평균 수돗물 사용량과 절감량등을 추정하여 계산한 결과, 생활용수 25% 제한시 $1m^3$당 추가적인 지불의사 금액은 약 534원으로 추정된다. 지불의사가 기대했던 것보다 낮은 이유는 소비자가 그 정도의 가뭄을 경험하지 못한데서 오는 조세저항으로 여겨진다. 그러나 이 같은 결과는 가뭄을 회피하기 위해 소비자가 기꺼이 지불할 수 있는 비용은 현재의 물 값보다 훨씬 높다는 것을 의미한다. 알 수 있었다.수 없는 실제 방사선작업 환경에서의 선량평가 수행시, 본 논문에서 제시한 선량평가방법을 적용할 경우 계획된 피폭과 예상하지 못한 방사선장에 대한 피폭을 구분할 수 있어 보다 심도 있는 피폭선량평가 및 방사선관리가 가능할 것으로 판단된다.황산기(黃酸基)의 함량은 가장 적었다내고 있었고, in vivo에서는 복강(復腔)(i. p.) 투여(投與)에서는 $12.9{\sim}22.2%$. 경구(經口)(p.o.) 투여(投與)에서는 $11.2{\sim}21.6%$의 높은 활성(活性)을 나타내고 있었다. 단 peroxidase 활성(活性)은 복강(復腔)(i. p.) 투여(投與)에서는 $129.0{\sim}188.6%$, 경구(經口)(p. o.) 투여(投與)에서는 $111.4{\sim}139.6%$의 활성(活性)을 나타내고 있었다.rucic acid를 다량함유(多量含有)하고 있는 목포단교 12는 다른 두 품종(品種)에 비(比)하여 장쇄지방산에 속하는 eicoseuoic acid가 11.7%로 많이 함유(含有)되어 있었다. 그리고 3품종(品種) 모두 중성지질(中性脂質) 획분(劃分)의 지질(脂質)은 전체지방산(全體脂肪酸) 조성(組成)의 비(比)와 별다른 차이(差異)를 나타내지 않았다.수 있으나 본원에서 사용 중인 장비에서 극히 적은 MU의 사용이 현저한 선량적 오차를 유발하진 않는 것으로 사료되어진다. 그러나 정도 관리 시 그 오차를 확인하는 과정은 장비의 사용과 수명에 따라 지속적으로 관리되어져야 할 것이다.>로 평균

  • PDF

Evaluation of the Radiochromic Film Dosimetry for a Small Curved Interface (휘어진 경계에서의 좁은 영역에 대한 Radiochromic 필름 도시메트리 평가)

  • Kang, Sei-Kwon;Park, Soah;Hwang, Taejin;Cheong, Kwang-Ho;Han, Taejin;Kim, Haeyoung;Lee, Me-Yeon;Kim, Kyoung Ju;Bae, Hoonsik
    • Progress in Medical Physics
    • /
    • v.23 no.4
    • /
    • pp.234-238
    • /
    • 2012
  • A tumor on the eyelid is often treated using a high-energy electron beam, with a metallic eye shield inserted between the eyelid and the eyeball to preserve the patient's sight. Pretreatment quality assurance of the inner eyelid dose on the metallic shield requires a very small dosimetry tool. For enhanced accuracy, a flexible device fitting the curved interface between the eyelid and the shield is also required. The radiochromic film is the best candidate for this device. To measure the doses along the curved interface and small area, a 3-mm-wide strip of EBT2 film was inserted between the phantom eyelid and the shield. After irradiation with 6 MeV electron beams, the film was evaluated for the dose profile. An acrylic eye shield of the same size as the real eye shield was machined, and CT images free from metal artifacts were obtained. Monte Carlo simulation was performed on the CT images, taking into account eye shield material, such as tungsten, aluminum, and steel. The film-based interface dose distribution agreed with the MC calculation within 2.1%. In the small (millimeter scale) and curved region, radiochromic film dosimetry promises a satisfactory result with easy handling.

Patient exposure doses from medical x-ray examinations in Korea (진단방사선검사에서 환자피폭선량에 관한 연구)

  • Kim, You-Hyun;Choi, Jong-Hak;Kim, Sung-Soo;Oh, Yuw-Han;Lee, Chan-Hyeup;Cho, Pyong-Kon;Kang, Dae-Hyun;Lee, Young-Bae;Kim, Hyung-Chul;Kim, Chel-Min
    • Journal of radiological science and technology
    • /
    • v.28 no.3
    • /
    • pp.241-248
    • /
    • 2005
  • X-ray examinations represent the largest man-made source of radiation exposure for the population. The need for standardization of radiation exposures has been suggested and the guidance levels for various radiographic and radioisotope examinations has been proposed by the International Atomic Energy Aency(IAEA) as a safety standard. In many countries, the situation of medical radiographic exposures in each country should be researched before the appropriate guidance level is established. In this study, measurements of entrance surface dose, dose-area product(DAP), computed tomograghic dose index(CTDI) and mean glandular dose(MGD) were carried out in patients who underwent routine x-ray examinations, fluoroscopy, computed tomograghy and mamography in Korea. These measured quantities were compared with the results from the calculation method in previous study. And we suggested diagnostic reference levels in medical imaging in Korea.

  • PDF

Calibration of CR-39 for Hadron Radiotherapy using 400 MeV/u C ions (400 MeV/u 탄소 이온에 대한 방사선치료 선량 측정용 고체비적검출기의 교정)

  • Kim, Sunghwan;Nam, Uk-Won;Lee, Jaejin;Park, Won-Kee;Pyo, Jeonghyun;Moon, Bong-Kon
    • Journal of radiological science and technology
    • /
    • v.39 no.1
    • /
    • pp.43-49
    • /
    • 2016
  • In this study, equivalent dose and LET (Linear Energy Transfer) calibration of CR-39 SSNTD (Solid State Nuclear Track Detector) were performed using 400 MeV/u C heavy ions in HIMAC (Heavy Ion Medical Accelerator in Chiba) for high LET radiation therapy. The irradiated CR-39 SSNDTs were etched according the etching condition of JAXA (Japan Aerospace Exploration Agency). And the etched SSNTDs were analyzed by using Image J. Determined frequency mean dose (${\bar{y_D}}$)and dose-mean lineal energy (${\bar{y_F}}$)of 400 MeV/u C are about 8.5keV/mm and 10.1 keV/mm, respectively by using the CR-39 SSNTD. This value is very similar to the results calculated by GEANT4 Monte Carlo simulation and measured with TEPC (Tissue Equivalent Proportional Counter) active radiation detector. We could determine the equivalent dose and LET calibration factors of CR-39. And we confirmed that the CR-39 SSNTD was useful for high LET radiation dosimetry in hadron radiotherapy.