Purpose: The result of exam using an imaging device is very closely related with the image quality. Moreover, this image quality can be changed according to the condition of image acquisition and evaluation method. In this study, we evaluated the image quality according to the difference of pre-set method in PET/CT image. Materials & Methods: PET/CT Discovery STe16 (GE Healthcare, Milwaukee, USA), Chest PET phantom (Experiment 1) and 94 NEMA phantom (Experiment 2) were used. Phantom were filled with $^{18}F$-FDG maintaining hot sphere and background ratio to 4:1. In the case of experiment 1, we set the radio activity concentration on 3.5, 6.0, 8.6 kBq/mL. In the case of experiment 2, we set the radio activity concentration on 3.3, 5.5, 7.7, 9.9, 12.1, 16.5 kBq/mL. All experiments were performed with the time-set method for 2 minutes 30 seconds per frame and the count-set method with one hundred million counts in 3D mode after CT transmission scan. For the evaluation of the image quality, we compared each results by using the NECR and SNR. Results: In the experiment 1, both the NECR and SNR were increased as radioactivity concentration getting increased. The NECR was shown as 53.7, 66.9, 91.4. and SNR was shown as 7.9, 10.0, 11.7. Both the NECR and SNR were increased in time-set method. But the count-set method's pattern was not similar with the time-set method. The NECR was shown as 53.8, 69.1, 97.8, and SNR was shown as 14.1, 14.7 14.4. The SNR was not increased in count-set method. In experiment 2, results of both the NECR and SNR were shown as 45.1, 70.6, 95.3, 115.6, 134.6, 162.2 and 7.1, 8.8, 10.6, 11.5, 12.7, 14.0. These results were shown similar patten with the experiment 1. Moreover, when the count-set method was applied, the NECR was shown as 42.1, 67.3, 92.1, 112.2, 130.7, 158.7, and SNR was shown as 15.2, 15.9, 15.6, 15.4, 15.5, 14.9. The NECR was increased but SNR was not shown same pattern. Conclusion: Increment of administered radioactivity improves the quality of image unconcerned with the pre-set method. However, NECR was not influenced by increment of total acquisition counts through simple increasing scan duration without increment of administered activity. In case of count-set method, the SNR was shown similar value despite of increment of radioactivity. So, the administered activity is more important than the scan duration. And we have to consider that evaluation of image quality using only SNR may not be appropriate.