Evaluation of the Image Quality According to the Pre-set Method in PET/CT Image
PET/CT 영상 획득 시 사전설정법 차이에 따른 영상 질 평가
- Park, Sun-Myung (Department of Nuclear Medicine, Samsung Medical Center) ;
- Lee, Hyuk (Department of Nuclear Medicine, Samsung Medical Center) ;
- Hong, Gun-Chul (Department of Nuclear Medicine, Samsung Medical Center) ;
-
Chung, Eun-Kyung
(Department of Nuclear Medicine, Samsung Medical Center) ;
-
Choi, Choon-Ki
(Department of Nuclear Medicine, Samsung Medical Center) ;
-
Seok, Jae-Dong
(Department of Nuclear Medicine, Samsung Medical Center)
- 박선명 (삼성서울병원 핵의학과) ;
- 이혁 (삼성서울병원 핵의학과) ;
- 홍건철 (삼성서울병원 핵의학과) ;
-
정은경
(삼성서울병원 핵의학과) ;
-
최춘기
(삼성서울병원 핵의학과) ;
-
석재동
(삼성서울병원 핵의학과)
- Received : 2011.06.28
- Accepted : 2011.07.05
- Published : 2011.11.05
Abstract
Purpose: The result of exam using an imaging device is very closely related with the image quality. Moreover, this image quality can be changed according to the condition of image acquisition and evaluation method. In this study, we evaluated the image quality according to the difference of pre-set method in PET/CT image. Materials & Methods: PET/CT Discovery STe16 (GE Healthcare, Milwaukee, USA), Chest PET phantom (Experiment 1) and 94 NEMA phantom (Experiment 2) were used. Phantom were filled with
영상장치를 이용한 검사 시 영상의 질은 검사결과와 밀접하게 연관되어 있으며, 이는 영상획득 조건들과 이에 대한 평가방법으로 인해 달라 질 수 있다. 본 연구에서는 영상획득 시 사전설정법(Pre-set method)차이에 따른 영상의 질을 평가하였다. 장비는 PET/CT Discovery STe16(GE Healthcare, Milwaukee, USA)을 사용하였으며, Chest PET Phantom (실험 1)과 94 NEMA Phantom (실험 2)을 이용하였다. 실험 1의 경우 3.5, 6.0, 8.6 kBq/mL, 실험 2의 경우는 3.3, 5.5, 7.7, 9.9, 12.1, 16.5 kBq/mL의