• Title/Summary/Keyword: 반치환 무릎 인공관절

Search Result 1, Processing Time 0.013 seconds

Analysis of Contact Pressure for Material Combination in Unicompartmental Knee Implant (반치환 무릎 인공관절에서의 재료조합에 따른 접촉압력 분석)

  • Noh, Tae-Heon;Chun, Heoung-Jae
    • Journal of the Computational Structural Engineering Institute of Korea
    • /
    • v.31 no.1
    • /
    • pp.23-29
    • /
    • 2018
  • In knee implants, contact pressure has a significant effect on wear. In this study, finite element analysis is performed using the knee implant model developed in the previous research. The contact pressures for a total of 10 knee implant materials combinations were analyzed using the combinations actually used in research and industry. In order to calculate the contact pressure, The load was applied when the flection angle of knee was $30^{\circ}$, $45^{\circ}$ and $60^{\circ}$. The result of contact pressure revealed the smallest contact pressure in the titanium alloy-UHMWPE combination. In the case of UHMWPE, contact pressure did not change much with any material used in the femur. Compared the combination with the largest contact pressure and the smallest contact pressure, the difference was 0.77%. On the other hand, Carbon / PEEK composites showed 5.3% difference when the contact pressure was the largest and the smallest. It can be seen that when the Carbon / PEEK composite material is used as the bearing part, the material of the femoral part affects the wear. This study will contribute to the prediction of knee implant wear and minimization of wear.