• Title/Summary/Keyword: [$^{68}Ga$]Ga-FAPI-04

Search Result 1, Processing Time 0.015 seconds

Development of a Synthetic Method for [68Ga]Ga-FAPI-04 Using a Cassette-based Synthesizer (카세트 기반 자동합성장치를 사용한 [68Ga]Ga-FAPI-04의 합성방법 연구)

  • Jun Young PARK;Won Jun KANG
    • Korean Journal of Clinical Laboratory Science
    • /
    • v.56 no.1
    • /
    • pp.43-51
    • /
    • 2024
  • [68Ga]Ga-FAPI-04 is a promising radiopharmaceutical that binds specifically to fibroblast activation protein, which is overexpressed in more than 90% of malignant epithelial tumors but not in normal healthy tissue. This study aimed to develop an efficient method for producing 68Ga-labelled FAPI-04 using a cassette-based automated synthesizer. [68Ga]GaCl3 was eluted from an Eckert & Ziegler Medical germanium-68/gallium-68 generator using 2.5 mL of 0.1 M HCl. The synthesis of the [68Ga]Ga-FAPI-04 was performed using different concentrations of HEPES (1~2.5 M; 4-(2-hydroxyethyl) piperazine-1-ethanesulfonic acid) in 3~10 minutes; amounts of FAPI-04 precursor (5~50 ㎍) and reaction temperature (25℃~100℃) were optimized on the BIKBox® synthesizer. The labeling efficiency of [68Ga]Ga-FAPI-04 was greater than 96% (decay corrected) using 25 ㎍ FAPI-04 synthesized in 10 minutes at 100℃ in 2 M HEPES (pH 3.85), and its stability was greater than 99% at 6 hours. The total synthesis time of [68Ga]Ga-FAPI-04 was 32.4 minutes, and the product met all quality control criteria. In this study, we developed and optimized a labeling method using [68Ga]Ga-FAPI-04 using a cassette-based synthesizer. The devised method is expected to be useful for supplying [68Ga]Ga-FAPI-04 for diagnosis in clinical practice.