Browse > Article
http://dx.doi.org/10.3340/jkns.2020.0179

Triggered Electrooculography for Identification of Oculomotor and Abducens Nerves during Skull Base Surgery  

Jeong, Ha-Neul (Department of Neurology, Yonsei University College of Medicine)
Ahn, Sang-Il (Department of Neurology, Yonsei University College of Medicine)
Na, Minkyun (Department of Neurosurgery, Yonsei University College of Medicine)
Yoo, Jihwan (Department of Neurosurgery, Yonsei University College of Medicine)
Kim, Woohyun (Department of Neurosurgery, Yonsei University College of Medicine)
Jung, In-Ho (Department of Neurosurgery, Yonsei University College of Medicine)
Kang, Soobin (Department of Neurology, Yonsei University College of Medicine)
Kim, Seung Min (Department of Neurology, Yonsei University College of Medicine)
Shin, Ha Young (Department of Neurology, Yonsei University College of Medicine)
Chang, Jong Hee (Department of Neurosurgery, Yonsei University College of Medicine)
Kim, Eui Hyun (Department of Neurosurgery, Yonsei University College of Medicine)
Publication Information
Journal of Korean Neurosurgical Society / v.64, no.2, 2021 , pp. 282-288 More about this Journal
Abstract
Objective : Electrooculography (EOG) records eyeball movements as changes in the potential difference between the negatively charged retina and the positively charged cornea. We aimed to investigate whether reliable EOG waveforms can be evoked by electrical stimulation of the oculomotor and abducens nerves during skull base surgery. Methods : We retrospectively reviewed the records of 18 patients who had undergone a skull base tumor surgery using EOG (11 craniotomies and seven endonasal endoscopic surgeries). Stimulation was performed at 5 Hz with a stimulus duration of 200 μs and an intensity of 0.1-5 mA using a concentric bipolar probe. Recording electrodes were placed on the upper (active) and lower (reference) eyelids, and on the outer corners of both eyes; the active electrode was placed on the contralateral side. Results : Reproducibly triggered EOG waveforms were observed in all cases. Electrical stimulation of cranial nerves (CNs) III and VI elicited positive waveforms and negative waveforms, respectively, in the horizontal recording. The median latencies were 3.1 and 0.5 ms for craniotomies and endonasal endoscopic surgeries, respectively (p=0.007). Additionally, the median amplitudes were 33.7 and 46.4 μV for craniotomies and endonasal endoscopic surgeries, respectively (p=0.40). Conclusion : This study showed reliably triggered EOG waveforms with stimulation of CNs III and VI during skull base surgery. The latency was different according to the point of stimulation and thus predictable. As EOG is noninvasive and relatively easy to perform, it can be used to identify the ocular motor nerves during surgeries as an alternative of electromyography.
Keywords
Abducens nerve; Electrooculography; Monitoring, Intraoperative; Oculomotor nerve;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Cornelius JF, Schipper J, Tortora A, Krause-Molle Z, Smuga M, Petridis AK, et al. : Continuous and dynamic facial nerve mapping during surgery of cerebellopontine angle tumors: clinical pilot series. World Neurosurg 119 : e855-e863, 2018   DOI
2 Fukaya C, Katayama Y, Kasai M, Kurihara J, Yamamoto T : Intraoperative electrooculographic monitoring of oculomotor nerve function during skull base surgery. Technical note. J Neurosurg 91 : 157-159, 1999   DOI
3 Hariharan P, Balzer JR, Anetakis K, Crammond DJ, Thirumala PD : Electrophysiology of extraocular cranial nerves: oculomotor, trochlear, and abducens nerve. J Clin Neurophysiol 35 : 11-15, 2018   DOI
4 Kaspera W, Adamczyk P, Slaska-Kaspera A, Ladzinski P : Usefulness of intraoperative monitoring of oculomotor and abducens nerves during surgical treatment of the cavernous sinus meningiomas. Adv Med Sci 60 : 25-30, 2015   DOI
5 Kassam AB, Prevedello DM, Carrau RL, Snyderman CH, Thomas A, Gardner P, et al. : Endoscopic endonasal skull base surgery: analysis of complications in the authors' initial 800 patients. J Neurosurg 114 : 1544-1568, 2011   DOI
6 Kawaguchi M, Ohnishi H, Sakamoto T, Shimizu K, Touho H, Monobe T, et al. : Intraoperative electrophysiologic monitoring of cranial motor nerves in skull base surgery. Surg Neurol 43 : 177-181, 1995   DOI
7 Kawamata T, Ishii N, Amano K, Namioka T, Hori T, Okada Y : A novel simple real-time electrooculographic monitoring system during transsphenoidal surgeries to prevent postoperative extraocular motor nerve dysfunction. Neurosurg Rev 36 : 371-376, 2013   DOI
8 Sheshadri V, Bharadwaj S, Chandramouli BA : Intra-operative electrooculographic monitoring to prevent post-operative extraocular motor nerve dysfunction during skull base surgeries. Indian J Anaesth 60 : 560-565, 2016   DOI
9 Singh H, Vogel RW, Lober RM, Doan AT, Matsumoto CI, Kenning TJ, et al. : Intraoperative neurophysiological monitoring for endoscopic endonasal approaches to the skull base: a technical guide. Scientifica (Cairo) 2016 : 1751245, 2016   DOI
10 Thirumala PD, Mohanraj SK, Habeych M, Wichman K, Chang YF, Gardner P, et al. : Value of free-run electromyographic monitoring of extraocular cranial nerves during expanded endonasal surgery (EES) of the skull base. J Neurol Surg Rep 74 : 43-50, 2013   DOI
11 Zhou Q, Zhang M, Jiang Y : Intraoperative oculomotor nerve monitoring predicts outcome following clipping of posterior communicating artery aneurysms. J Clin Neurosci 19 : 706-711, 2012   DOI
12 Sekiya T, Hatayama T, Iwabuchi T, Maeda S : Intraoperative recordings of evoked extraocular muscle activities to monitor ocular motor nerve function. Neurosurgery 32 : 227-235; discussion 235, 1993   DOI
13 Lopez JR : Neurophysiologic intraoperative monitoring of the oculomotor, trochlear, and abducens nerves. J Clin Neurophysiol 28 : 543-550, 2011   DOI
14 Schlake HP, Goldbrunner R, Siebert M, Behr R, Roosen K : Intra-operative electromyographic monitoring of extra-ocular motor nerves (Nn. III, VI) in skull base surgery. Acta Neurochir (Wien) 143 : 251-261, 2001   DOI