Browse > Article
http://dx.doi.org/10.3340/jkns.2018.0075

Cyberknife Dosimetric Planning Using a Dose-Limiting Shell Method for Brain Metastases  

Yoon, Kyoung Jun (Radiosurgery Center, Asan Medical Center, University of Ulsan College of Medicine)
Cho, Byungchul (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Kwak, Jung Won (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Lee, Doheui (Radiosurgery Center, Asan Medical Center, University of Ulsan College of Medicine)
Kwon, Do Hoon (Department of Neurosurgery, Asan Medical Center, University of Ulsan College of Medicine)
Ahn, Seung Do (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Lee, Sang-Wook (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Kim, Chang Jin (Department of Neurosurgery, Asan Medical Center, University of Ulsan College of Medicine)
Roh, Sung Woo (Department of Neurosurgery, Asan Medical Center, University of Ulsan College of Medicine)
Cho, Young Hyun (Department of Neurosurgery, Asan Medical Center, University of Ulsan College of Medicine)
Publication Information
Journal of Korean Neurosurgical Society / v.61, no.6, 2018 , pp. 753-760 More about this Journal
Abstract
Objective : We investigated the effect of optimization in dose-limiting shell method on the dosimetric quality of CyberKnife (CK) plans in treating brain metastases (BMs). Methods : We selected 19 BMs previously treated using CK between 2014 and 2015. The original CK plans ($CK_{original}$) had been produced using 1 to 3 dose-limiting shells : one at the prescription isodose level (PIDL) for dose conformity and the others at low-isodose levels (10-30% of prescription dose) for dose spillage. In each case, a modified CK plan ($CK_{modified}$) was generated using 5 dose-limiting shells : one at the PIDL, another at intermediate isodose level (50% of prescription dose) for steeper dose fall-off, and the others at low-isodose levels, with an optimized shell-dilation size based on our experience. A Gamma Knife (GK) plan was also produced using the original contour set. Thus, three data sets of dosimetric parameters were generated and compared. Results : There were no differences in the conformity indices among the $CK_{original}$, $CK_{modified}$, and GK plans (mean 1.22, 1.18, and 1.24, respectively; p=0.079) and tumor coverage (mean 99.5%, 99.5%, and 99.4%, respectively; p=0.177), whereas the $CK_{modified}$ plans produced significantly smaller normal tissue volumes receiving 50% of prescription dose than those produced by the $CK_{original}$ plans (p<0.001), with no statistical differences in those volumes compared with GK plans (p=0.345). Conclusion : These results indicate that significantly steeper dose fall-off is able to be achieved in the CK system by optimizing the shell function while maintaining high conformity of dose to tumor.
Keywords
Radiosurgery; Brain; Neoplasm metastasis;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Blonigen BJ, Steinmetz RD, Levin L, Lamba MA, Warnick RE, Breneman JC : Irradiated volume as a predictor of brain radionecrosis after linear accelerator stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 77 : 996-1001, 2010   DOI
2 Chin LS, Ma L, DiBiase S : Radiation necrosis following gamma knife surgery: a case-controlled comparison of treatment parameters and longterm clinical follow up. J Neurosurg 94 : 899-904, 2001   DOI
3 Cho YH, Lee JM, Lee D, Park JH, Yoon K, Kim SO, et al. : Experiences on two different stereotactic radiosurgery modalities of Gamma Knife and Cyberknife in treating brain metastases. Acta neurochir (Wien) 157 : 2003-2009; discussion 2009, 2015   DOI
4 Fahrig A, Ganslandt O, Lambrecht U, Grabenbauer G, Kleinert G, Sauer R, et al. : Hypofractionated stereotactic radiotherapy for brain metastases--results from three different dose concepts. Strahlenther Onkol 183 : 625-630, 2007   DOI
5 Feuvret L, Vinchon S, Martin V, Lamproglou I, Halley A, Calugaru V, et al. : Stereotactic radiotherapy for large solitary brain metastases. Cancer Radiother 18 : 97-106, 2014   DOI
6 Gevaert T, Levivier M, Lacornerie T, Verellen D, Engels B, Reynaert N, et al. : Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of arteriovenous malformations and acoustic neuromas. Radiother Oncol 106 : 192-197, 2013   DOI
7 Giubilei C, Ingrosso G, D'Andrea M, Benassi M, Santoni R : Hypofractionated stereotactic radiotherapy in combination with whole brain radiotherapy for brain metastases. J Neurooncol 91 : 207-212, 2009   DOI
8 Jiang XS, Xiao JP, Zhang Y, Xu YJ, Li XP, Chen XJ, et al. : Hypofractionated stereotactic radiotherapy for brain metastases larger than three centimeters. Radiat Oncol 7 : 36, 2012   DOI
9 Kim YJ, Cho KH, Kim JY, Lim YK, Min HS, Lee SH, et al. : Single-dose versus fractionated stereotactic radiotherapy for brain metastases. Int J Radiat Oncol Biol Phys 81 : 483-489, 2011   DOI
10 Kaul D, Badakhshi H, Gevaert T, Pasemann D, Budach V, Tuleasca C, et al. : Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of meningioma. Acta Neurochir (Wien) 157 : 559-563; discussion 563-564, 2015   DOI
11 Massager N, Maris C, Nissim O, Devriendt D, Salmon I, Levivier M : Experimental analysis of radiation dose distribution in radiosurgery. II. Dose fall-off outside the target volume. Stereotact Funct Neurosurg 87 : 137-142, 2009   DOI
12 Korytko T, Radivoyevitch T, Colussi V, Wessels BW, Pillai K, Maciunas RJ, et al. : 12 Gy gamma knife radiosurgical volume is a predictor for radiation necrosis in non-AVM intracranial tumors. Int J Radiat Oncol Biol Phys 64 : 419-424, 2006   DOI
13 Kwon AK, Dibiase SJ, Wang B, Hughes SL, Milcarek B, Zhu Y : Hypofractionated stereotactic radiotherapy for the treatment of brain metastases. Cancer 115 : 890-898, 2009   DOI
14 Ma L, Sahgal A, Descovich M, Cho YB, Chuang C, Huang K, et al. : Equivalence in dose fall-off for isocentric and nonisocentric intracranial treatment modalities and its impact on dose fractionation schemes. Int J Radiat Oncol Biol Phys 76 : 943-948, 2010   DOI
15 Meeks SL, Buatti JM, Bova FJ, Friedman WA, Mendenhall WM : Treatment planning optimization for linear accelerator radiosurgery. Int J Radiat Oncol Biol Phys 41 : 183-197, 1998   DOI
16 Minniti G, Clarke E, Lanzetta G, Osti MF, Trasimeni G, Bozzao A, et al. : Stereotactic radiosurgery for brain metastases: analysis of outcome and risk of brain radionecrosis. Radiat Oncol 6 : 48, 2011   DOI
17 Sio TT, Jang S, Lee SW, Curran B, Pyakuryal AP, Sternick ES : Comparing Gamma Knife and CyberKnife in patients with brain metastases. J Appl Clin Med Phys 15 : 4095, 2014
18 Minniti G, D'Angelillo RM, Scaringi C, Trodella LE, Clarke E, Matteucci P, et al. : Fractionated stereotactic radiosurgery for patients with brain metastases. J Neurooncol 117 : 295-301, 2014   DOI
19 Murai T, Ogino H, Manabe Y, Iwabuchi M, Okumura T, Matsushita Y, et al. : Fractionated stereotactic radiotherapy using CyberKnife for the treatment of large brain metastases: a dose escalation study. Clin Oncol (R Coll Radiol) 26 : 151-158, 2014   DOI
20 Paddick I, Lippitz B : A simple dose gradient measurement tool to complement the conformity index. J Neurosurg 105 Suppl : 194-201, 2006   DOI
21 Solberg TD, Balter JM, Benedict SH, Fraass BA, Kavanagh B, Miyamoto C, et al. : Quality and safety considerations in stereotactic radiosurgery and stereotactic body radiation therapy: executive summary. Pract Radiat Oncol 2 : 2-9, 2012   DOI
22 Wagner TH, Bova FJ, Friedman WA, Buatti JM, Bouchet LG, Meeks SL : A simple and reliable index for scoring rival stereotactic radiosurgery plans. Int J Radiat Oncol Biol Phys 57 : 1141-1149, 2003   DOI
23 Wowra B, Muacevic A, Tonn JC : Quality of radiosurgery for single brain metastases with respect to treatment technology: a matched-pair analysis. J Neurooncol 94 : 69-77, 2009   DOI
24 Zeverino M, Jaccard M, Patin D, Ryckx N, Marguet M, Tuleasca C, et al. : Commissioning of the leksell Gamma Knife(R) $Icon^{TM}$. Med Phys 44 : 355-363, 2017   DOI
25 Yu C, Jozsef G, Apuzzo ML, Petrovich Z : Dosimetric comparison of CyberKnife with other radiosurgical modalities for an ellipsoidal target. Neurosurgery 53 : 1155-1162; discussion 1162-1163, 2003   DOI