Foreign Body Reaction after Implantation of a Device for Intervertebral Assisted Motion  

Seo, Jun-Yeong (Department of Orthopaedic Surgery, Jeju National University Hospital, School of Medicine, Jeju National University)
Ha, Kee-Yong (Department of Orthopaedic Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea)
Kim, Young-Hoon (Department of Orthopaedic Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea)
Ahn, Joo-Hyun (Department of Orthopaedic Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea)
Publication Information
Journal of Korean Neurosurgical Society / v.59, no.6, 2016 , pp. 647-649 More about this Journal
Abstract
The device for intervertebral assisted motion (DIAM) is a dynamic implant that consists of a silicone bumper enveloped by a polyethylene terephthalate (PET) fiber sack. Silicone and PET were used because of their biological inertness, but repetitive motion of the spine can cause wear on the implant nonetheless. The purpose of this study is to report a case of foreign body reaction (FBR) against a DIAM. A 72-year-old female patient presented with lower back pain and both legs radiating pain. She had undergone DIAM implantation at L4-5 for spinal stenosis 5 years previously. The intervertebral disc space of L4-5, where the DIAM was inserted, had collapsed and degenerative scoliosis had developed due to left-side collapse. MRI showed L3-4 thecal sac compression and left L4-5 foraminal stenosis. The patient underwent removal of the DIAM and instrumented fusion from L3 to L5. During surgery, fluid and granulation tissue were evident around the DIAM. Histopathology showed scattered wear debris from the DIAM causing chronic inflammation due to the resulting FBR. A FBR due to wear debris of a DIAM can induce a hypersensitivity reaction and bone resorption around the implant, causing it to loosen.
Keywords
Device for intervertebral assisted motion; Foreign body reaction; Polyethylene terephthalate; Silicone;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Akhaddar A, Boulahroud O, Naama O, Al-Bouzidi A, Boucetta M : Paraspinal textiloma after posterior lumbar surgery : a wolf in sheep's clothing. World Neurosurg 77 : 375-380, 2012   DOI
2 Anderson JM : Inflammatory response to implants. ASAIO Trans 34 : 101-107, 1988   DOI
3 Anderson JM : Multinucleated giant cells. Curr Opin Hematol 7 : 40-47, 2000   DOI
4 Cunningham BW, Hallab NJ, Hu N, McAfee PC : Epidural application of spinal instrumentation particulate wear debris : a comprehensive evaluation of neurotoxicity using an in vivo animal model. J Neurosurg Spine 19 : 336-350, 2013   DOI
5 Daniel J, Holland J, Quigley L, Sprague S, Bhandari M : Pseudotumors associated with total hip arthroplasty. J Bone Joint Surg Am 94 : 86-93, 2012   DOI
6 Ha KY, Seo JY, Kwon SE, Son IN, Kim KW, Kim YH : Posterior dynamic stabilization in the treatment of degenerative lumbar stenosis : validity of its rationale. J Neurosurg Spine 18 : 24-31, 2013   DOI
7 Jerosch J, Moursi MG : Foreign body reaction due to polyethylene's wear after implantation of an interspinal segment. Arch Orthop Trauma Surg 128 : 1-4, 2008
8 Kim DH, Albert TJ : Interspinous process spacers. J Am Acad Orthop Surg 15 : 200-207, 2007   DOI
9 Potter EH, Rohrich RJ, Bolden KM : The role of silicone granulomas in recurrent capsular contracture : a review of the literature and an approach to management. Plast Reconstr Surg 131 : 888e-895e, 2013   DOI
10 Rodriguez A, Meyerson H, Anderson JM : Quantitative in vivo cytokine analysis at synthetic biomaterial implant sites. J Biomed Mater Res A 89 : 152-159, 2009