Browse > Article
http://dx.doi.org/10.3340/jkns.2014.55.3.136

Reliability of Stereotactic Coordinates of 1.5-Tesla and 3-Tesla MRI in Radiosurgery and Functional Neurosurgery  

Kim, Hae Yu (Gamma Knife Center, Haeundae Paik Hospital, Inje University College of Medicine)
Lee, Sun-Il (Gamma Knife Center, Haeundae Paik Hospital, Inje University College of Medicine)
Jin, Seong Jin (Gamma Knife Center, Haeundae Paik Hospital, Inje University College of Medicine)
Jin, Sung-Chul (Department of Neurosurgery, Haeundae Paik Hospital, Inje University College of Medicine)
Kim, Jung Soo (Department of Neurosurgery, Haeundae Paik Hospital, Inje University College of Medicine)
Jeon, Kyoung Dong (Department of Neurosurgery, Haeundae Paik Hospital, Inje University College of Medicine)
Publication Information
Journal of Korean Neurosurgical Society / v.55, no.3, 2014 , pp. 136-141 More about this Journal
Abstract
Objective : The aims of this study are to identify interpersonal differences in defining coordinates and to figure out the degree of distortion of the MRI and compare the accuracy between CT, 1.5-tesla (T) and 3.0T MRI. Methods : We compared coordinates in the CT images defined by 2 neurosurgeons. We also calculated the errors of 1.5T MRI and those of 3.0T. We compared the errors of the 1.5T with those of the 3.0T. In addition, we compared the errors in each sequence and in each axis. Results : The mean difference in the CT images between the two neurosurgeons was $0.48{\pm}0.22mm$. The mean errors of the 1.5T were $1.55{\pm}0.48mm$ (T1), $0.75{\pm}0.38$ (T2), and $1.07{\pm}0.57$ (FLAIR) and those of the 3.0T were $2.35{\pm}0.53$ (T1), $2.18{\pm}0.76$ (T2), and $2.16{\pm}0.77$ (FLAIR). The smallest mean errors out of all the axes were in the x axis : 0.28-0.34 (1.5T) and 0.31-0.52 (3.0T). The smallest errors out of all the MRI sequences were in the T2 : 0.29-0.58 (1.5T) and 0.31-1.85 (3.0T). Conclusion : There was no interpersonal difference in running the Gamma $Plan^{(R)}$ to define coordinates. The errors of the 3.0T were greater than those of the 1.5T, and these errors were not of an acceptable level. The x coordinate error was the smallest and the z coordinate error was the greatest regardless of the MRI sequence. The T2 sequence was the most accurate sequence.
Keywords
MRI; Stereotactic radiosurgery; Functional neurosurgery;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Tanner SF, Finnigan DJ, Khoo VS, Mayles P, Dearnaley DP, Leach MO : Radiotherapy planning of the pelvis using distortion corrected MR images : the removal of system distortions. Phys Med Biol 45 : 2117-2132, 2000   DOI
2 Villemure JG, Marchand E, Peters T, Leroux G, Olivier A : Magnetic resonance imaging stereotaxy : recognition and utilization of the commissures. Appl Neurophysiol 50 : 57-62, 1987
3 Watanabe Y, Lee CK, Gerbi BJ : Geometrical accuracy of a 3-tesla magnetic resonance imaging unit in Gamma Knife surgery. J Neurosurg 105 Suppl : 190-193, 2006   DOI
4 Wu TH, Lee JS, Wu HM, Chu WF, Guo WY : Evaluating geometric accuracy of multi-platform stereotactic neuroimaging in radiosurgery. Stereotact Funct Neurosurg 78 : 39-48, 2002   DOI
5 Zhang B, MacFadden D, Damyanovich AZ, Rieker M, Stainsby J, Bernstein M, et al. : Development of a geometrically accurate imaging protocol at 3 Tesla MRI for stereotactic radiosurgery treatment planning. Phys Med Biol 55 : 6601-6615, 2010   DOI
6 Kangarlu A, Baertlein BA, Lee R, Ibrahim T, Yang L, Abduljalil AM, et al. : Dielectric resonance phenomena in ultra high field MRI. J Comput Assist Tomogr 23 : 821-831, 1999   DOI
7 Kondziolka D, Dempsey PK, Lunsford LD, Kestle JR, Dolan EJ, Kanal E, et al. : A comparison between magnetic resonance imaging and computed tomography for stereotactic coordinate determination. Neurosurgery 30 : 402-406; discussion 406-407, 1992   DOI
8 Loganathan AG, Chan MD, Alphonse N, Peiffer AM, Johnson AJ, McMullen KP, et al. : Clinical outcomes of brain metastases treated with Gamma Knife radiosurgery with 3.0 T versus 1.5 T MRI-based treatment planning : have we finally optimised detection of occult brain metastases? J Med Imaging Radiat Oncol 56 : 554-560, 2012   DOI
9 Lunsford LD, Martinez AJ, Latchaw RE : Stereotaxic surgery with a magnetic resonance- and computerized tomography-compatible system. J Neurosurg 64 : 872-878, 1986   DOI
10 MacFadden D, Zhang B, Brock KK, Hodaie M, Laperriere N, Schwartz M, et al. : Clinical evaluation of stereotactic target localization using 3-Tesla MRI for radiosurgery planning. Int J Radiat Oncol Biol Phys 76 : 1472-1479, 2010   DOI
11 Mack A, Wolff R, Scheib S, Rieker M, Weltz D, Mack G, et al. : Analyzing 3-tesla magnetic resonance imaging units for implementation in radiosurgery. J Neurosurg 102 Suppl : 158-164, 2005   DOI
12 Price RR, Axel L, Morgan T, Newman R, Perman W, Schneiders N, et al. : Quality assurance methods and phantoms for magnetic resonance imaging : report of AAPM nuclear magnetic resonance Task Group No. 1. Med Phys 17 : 287-295, 1990   DOI   ScienceOn
13 Dammann P, Kraff O, Wrede KH, Özkan N, Orzada S, Mueller OM, et al. : Evaluation of hardware-related geometrical distortion in structural MRI at 7 Tesla for image-guided applications in neurosurgery. Acad Radiol 18 : 910-916, 2011   DOI
14 Reinsberg SA, Doran SJ, Charles-Edwards EM, Leach MO : A complete distortion correction for MR images : II. Rectification of static-field inhomogeneities by similarity-based profile mapping. Phys Med Biol 50 : 2651-2661, 2005   DOI
15 Andoh K, Nakamae H, Ohkoshi T, Odagiri K, Kyuma Y, Hayashi A : Technical note : enhanced MR-guided stereotaxic brain surgery with the patient under general anesthesia. AJNR Am J Neuroradiol 12 : 135-138, 1991
16 Choi DR, Ahn YC, Kim DY, Huh SJ, Lee JI : Accuracy in target localization in stereotactic radiosurgery. Med Dosim 22 : 53-58, 1997   DOI
17 Fransson A, Andreo P, Pötter R : Aspects of MR image distortions in radiotherapy treatment planning. Strahlenther Onkol 177 : 59-73, 2001   DOI
18 Guo WY : Application of MR in stereotactic radiosurgery. J Magn Reson Imaging 8 : 415-420, 1998   DOI
19 Bradford R, Thomas DG, Bydder GM : MRI-directed stereotactic biopsy of cerebral lesions. Acta Neurochir Suppl (Wien) 39 : 25-27, 1987   DOI