Browse > Article
http://dx.doi.org/10.3340/jkns.2012.51.4.222

Waffle-Cone Technique Using Solitaire AB Stent  

Park, Hye-Ran (Department of Neurosurgery, Soonchunhyang University Cheonan Hospital)
Yoon, Seok-Mann (Department of Neurosurgery, Soonchunhyang University Cheonan Hospital)
Shim, Jai-Joon (Department of Neurosurgery, Soonchunhyang University Cheonan Hospital)
Kim, Sung-Ho (Department of Neurosurgery, Soonchunhyang University Gumi Hospital)
Publication Information
Journal of Korean Neurosurgical Society / v.51, no.4, 2012 , pp. 222-226 More about this Journal
Abstract
The waffle-cone technique is a modified stent application technique, which involves protrusion of the distal portion of a stent into an aneurysm fundus to provide neck support for subsequent coiling. The authors report two cases of wide necked basilar bifurcation aneurysms, which were not amenable to stent assisted coiling, that were treated using the waffle-cone technique with a Solitaire AB stent. A 58-year-old woman presented with severe headache. Brain CT showed subarachnoid hemorrhage and angiography demonstrated a ruptured giant basilar bifurcation aneurysm with broad neck, which was treated with a Solitaire AB stent and coils using the waffle-cone technique. The second case involved an 81-year-old man, who presented with dizziness caused by brain stem infarction. Angiography also demonstrated a large basilar bifurcation unruptured aneurysm with broad neck. Solitaire AB stent deployment using the waffle-cone technique, followed by coiling resulted in near complete obliteration of aneurysm. The waffle-cone technique with a Solitaire AB stent can be a useful alternative to conventional stent application when it is difficult to catheterize bilateral posterior cerebral arteries in patients with a wide-necked basilar bifurcation aneurysm.
Keywords
Wide-necked aneurysm; Waffle-cone technique; Solitaire AB stent;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Horikoshi T, Fukamachi A, Nishi H, Fukasawa I : Detection of intracranial aneurysms by three-dimensional time-of-flight magnetic resonance angiography. Neuroradiology 36 : 203-207, 1994   DOI   ScienceOn
2 Horowitz M, Levy E, Sauvageau E, Genevro J, Guterman LR, Hanel R, et al. : Intra/extra-aneurysmal stent placement for management of complex and wide-necked-bifurcation aneurysms : eight cases using the waffle cone technique. Neurosurgery 58 : ONS258-ONS262; discussion ONS262, 2006
3 Huang Q, Xu Y, Hong B, Zhao R, Zhao W, Liu J : Stent-assisted embolization of wide-neck anterior communicating artery aneurysms : review of 21 consecutive cases. AJNR Am J Neuroradiol 30 : 1502-1506, 2009   DOI   ScienceOn
4 Kelly ME, Turner RD 4th, Moskowitz SI, Gonugunta V, Hussain MS, Fiorella D : Delayed migration of a self-expanding intracranial microstent. AJNR Am J Neuroradiol 29 : 1959-1960, 2008   DOI   ScienceOn
5 Klisch J, Clajus C, Sychra V, Eger C, Strasilla C, Rosahl S, et al. : Coil embolization of anterior circulation aneurysms supported by the Solitaire AB Neurovascular Remodeling Device. Neuroradiology 52 : 349-359, 2010   DOI   ScienceOn
6 Moret J, Cognard C, Weill A, Castaings L, Rey A : [Reconstruction technic in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases]. J Neuroradiol 24 : 30-44, 1997
7 Murao K, Yamada N, Miyamoto S, Hyuga T, Kudo T, Satow T : Identification of inflow zone of cerebral aneurysm by MRA for effective coil embolization. Surg Cereb Stroke 36 : 459-463, 2008   DOI   ScienceOn
8 Ryu CW, Koh JS, Lee CY, Kim EJ : Endovascular management of the wide-neck aneurysms : the applications of the coils and catheter. Neurointervention 5 : 71-78, 2010   DOI
9 Perez-Arjona E, Fessler RD : Basilar artery to bilateral posterior cerebral artery 'Y stenting' for endovascular reconstruction of wide-necked basilar apex aneurysms : report of three cases. Neurol Res 26 : 276-281, 2004   DOI   ScienceOn
10 Pierce DS, Rosero EB, Modrall JG, Adams-Huet B, Valentine RJ, Clagett GP, et al. : Open-cell versus closed-cell stent design differences in blood flow velocities after carotid stenting. J Vasc Surg 49 : 602-606; discussion 606, 2009   DOI   ScienceOn
11 Sychra V, Klisch J, Werner M, Dettenborn C, Petrovitch A, Strasilla C, et al. : Waffle-cone technique with SolitaireTM AB remodeling device : endovascular treatment of highly selected complex cerebral aneurysms. Neuroradiology 53 : 961-972, 2011   DOI   ScienceOn
12 Wanke I, Doerfler A, Schoch B, Stolke D, Forsting M : Treatment of wide-necked intracranial aneurysms with a self-expanding stent system : initial clinical experience. AJNR Am J Neuroradiol 24 : 1192-1199, 2003
13 Baldi S, Mounayer C, Piotin M, Spelle L, Moret J : Balloon-assisted coil placement in wide-neck bifurcation aneurysms by use of a new, compliant balloon microcatheter. AJNR Am J Neuroradiol 24 : 1222-1225, 2003
14 Biondi A, Janardhan V, Katz JM, Salvaggio K, Riina HA, Gobin YP : Neuroform stent-assisted coil embolization of wide-neck intracranial aneurysms : strategies in stent deployment and midterm follow-up. Neurosurgery 61 : 460-468; discussion 468-469, 2007   DOI   ScienceOn
15 Gruber DP, Crone KR : Endoscopic washout : a new technique for treating chronic subdural hematomas in infants. Pediatr Neurosurg 27 : 292-295, 1997   DOI   ScienceOn