This study is aimed at decreasing wrong diagnosis with corrected EF(Ejection Fraction) of 2D echo cardiography by analysing the physical time-resolution difference between Cine MRI and 2D echo cardiography and applying the corrected EF in 2D echocardiography. From February 2010 to December 2011, among the 110 patients who had undergone both 2D echo cardiography and cine MRI only 37 patient were selected suffering aortic valve regurgitation. ED, ES and SV were measured and EF was calculated in each system while normal ranges of Cine MRI and 2D echocardiography were compared to evauate misdiagnosis rate. The correlation of physical time resolution between 2D echocardiography and MRI was evaluated and the differences were corrected with linear regression coefficient which is derived from linear regression analysis. Blandt-Altman plot was used to evaluate the reliability of corrected 2D echo cardiography EF and compare the error among measured values. The values were compared with MRI normal range and misdiagnosis rate was measured again. As a result, misdiagnosis rates of physical time resolution were measured to be 32.4%(12people) before the correction of EF and 18.9%(7people) after the correction. Also, EF confirmed in Blandt-Altman plot were almost the same with MRI EF. In conclusion, when diagnosing aortic regurgitation disease, simply using 2D echocardiography can easily raise the misdiagnosis rates, therefore considering the MRI machine's physical merits, correcting the time resolution difference is important by calculating time resolution wrong diagnosis would decrease and it is considered to be useful in clinical circumstances.
Keywords
Cine MRI; 2D echo cardiography; ejection fraction; aortic valve regurgitation; time resolution;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Young Mi Hong, et al., "Comparison between Echocardiography and Cardiac Cine-MR: Left Ventricular Volume and Cardiac Output", Ewha Med J, 15, 4, pp. 327-336, 1992.   DOI
2 Leung MP, et al., "The management of symptomatic neonates with suspected congenital heart disease using combined cross-sectional echocardiography and pulsed doppler flow study as the definitive invesigation", Int J Cardiol, 24, pp. 41-40, 1989.   DOI
3 Bozio A, et al., "Cardiac surgery without catheterization: Influence of color doppler flow mapping", Pediatr cardiol, 11, pp. 245, 1990.
4 Meyer RH, et al., "Ventricular septum in right ventricular volume overload", Am J Cardiol, 30, pp. 349-353, 1972.   DOI
5 Levine RA, et al., "Echo cardiographic measurement of right ventricular volume", Circulation, 69, pp. 497-505, 1984.   DOI
6 van der Wall EE, et al., "Magnetic resonance imaging: A new approach for evaluating coronary artery disease", Am Heart J, 121, pp. 1203-1220, 1991.   DOI
7 Dodge HT, et al., "The use of the biplane angiography for the measurement of left ventricular volume in man", Am Heart, 60, pp. 762-776, 1960.   DOI
8 Dehmer Gl, et al., "Non geometric determination of left ventricular volumes from equilibrium blood pool scans", Am J Cardiol, 45, pp. 293-300, 1980.   DOI
9 Slutsky R, et al., "Left ventricular volumes by gated equilibrium radionuclide angiography: A new method", Circulation, 60, pp. 556-564, 1979.   DOI
10 Buser PT, et al., "Noninvasive evaluation of global left ventricular function with use of cine nuclear magnetic resonance", J Am Coll Cardiol, 13, pp. 1294-1300, 1989.   DOI
11 Byrd BF, et al., "Normal cardiac dimensions by magnetic resonance imaging", Am J Cardiol, 55, pp. 1440-1442, 1985.   DOI
12 Haase A, et al., "Flash imaging: rapid NMR imaging using low flipangle pulses", J Magn Reson, 67, pp. 258-266, 1986.
13 Hsu YH, et al., "Magnetic resonance imaging of total anomalous pulmonary venous drainage", Am Heart J, 121, pp. 1560-1565, 1991.   DOI
14 Marino B, et al., Pulmonary atresia with ventricular septal defect. Selection of patients for systemic to pulmonary arterial shunt based on echocardiography", Cheat, 99, pp. 158-161, 1991.