Abstract
The purpose of this study is to assess the mechanical properties of the ceramic abutment with washer. In this study, ceramic abutment were used, tested with $30^{\circ}$ compression load, shear fatigue, adaptation accuracy test(rotation angle, contact interval), removal torque force test, torsional breaking force test. The $30^{\circ}$ compression load was 729 N, the shear fatigue load was 275 N, adaptation accuracy test of rotation angle was within $3^{\circ}$, contact interval within $10{\mu}m$, and removal torque force test value is $18.88N{\cdot}cm$, torsional breaking force test value is $35.52N{\cdot}cm$. Ceramic abutment with a washer fitted have sufficient mechanical strength and may be substituted for titanium abutment.
Keywords
Ceramic abutment; Washer; Implant; Mechanical strength; Screw loosening;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Young-Gyun Song, Osseointegration of Ceramics & Zirconia : A Review of Literature, Journal of Dental Rehabilitation and Applied Science 28 (2012) 219-326.
2 Yong-Hoon Jeong, Young-Pil Moon, Chung-Hwan Lee, Jin-Woo Yu, Han-Cheol Choe, Surface Characteristics of Dental Implant Fixture with Various Manufacturing Process, J. Kor. Inst. Surf. Eng. 43 (2010) 17-24.   DOI
3 Sung-ae Shin, Chang-Seop Kim, Wook Cho, Chang-Mo Jeong, Young-Chan Jeon, Ji-Hoon Yun, Mechanical strength of Zirconia Abutment in Implant Restoration, Korean academy of stomatognathic function and occlusion 25 (2009) 349-360.
4 Tae-Jung Kim, Sung-Bin Lim, Chin-Hyung Chung, The Evaluation of the atomic composition and the surface roughness of Titanium Implants following Various Laser treatment with air-powder abrasive, The journal of Korean academy of periodontology 32 (2002) 615-630.   DOI
5 Tae-Su Kim, Jae-Hyun Lee, Won-Sup Lee, Su-Young Lee, Removal of fractured implant screws: case report, Journal of Dental Rehabilitation and Applied Science 31 (2015) 60-66.   DOI
6 Byung-Ryong Bae, Yu-Sung Choi, In-Ho Cho, Influence of Implant Abutment Systems on Detorque Value and Screw Joint Stability, Journal of Dental Rehabilitation and Applied Science 26 (2010) 97-109.
7 J Bickford, An introduction of the design and behavior of bolted joints, Marcel Dekker (1995) 515-564.
8 A Versluis, TW Korioth, AC Cardoso, Numerical analysis of a dental implant system preload with a washer, Int J Oral Maxillofac Implants 14 (1999) 337-341.
9 TW Korioth, AC Cardoso, A Versluis, Effect of washers on reverse torque displacement of dental implant gold retaining screws, J Prosthet Dent 82 (1999) 312-316.   DOI
10 Gab-Woon Hwang, Man-Hyung Lee, Kyung-Woo Song, Min-Ho Lee, Tae-Sung Bae, Fatigue Characteristics of ti-6Al-4V alloy mini-implant with friction locked abutment, Korean Journal of Dental Materials 35 (2008) 359-368.
11 WC Marthin , RD Woody, BH Miller, AW Miller, Implant abutment screw retations and preloads for four different screw matrials and surfaces, J Prosthet Dent 86 (2001) 24-32.   DOI
12 DG Gratton, SA Aquilino, CM Stanford, Micromotion and dynamic fatigue properties of the dental implantabutment interface, J Prosthet Dent 85 (2001) 47-52.   DOI
13 Hyon-Mo Shin, Chang-Mo Jeong, Mi-Jeong Yun, Influence of Abutment Design and Tightening Torque on the Removal Torque Value in Three Implant Systems, Korean Journal of Dental Materials 35 (2008) 121-131.
14 Da-Un Jung, Chae-Heon Chung, Mee-Kyoung Son, Han-Cheol Choe, Effects of TiN Coating on the Fatigue Fracture of Dental Implant System with Various Cyclic Loads, J. Kor. Inst. Surf. Eng. 48 (2015) 283-291.   DOI
15 A Versluis, TW Korioth, AC Cardoso, Numerical analysis of a dental implant system preloaded with a washer, Int J Oral Maxillofac Implants 14 (1999) 337-341.
16 TW Korioth, AC Cardoso, A Versluis, Effect of washers on reverse torque displacement of dental implant gold retaining screws, J Prosthet Dent 82 (1999) 312-316.   DOI
17 O Robert, PE Parmeley, Standard handbook of fastening and joining, 1997.