Analysis of a Comparability Test between LX Detergent Cleaning Solution and OC Detergent Cleaning Solution Using OC Sensor PLEDIA  

Cha, Kyung Jae (Department of Field Application Scientist Team, Shinyang Chemical Co.,Ltd.)
Publication Information
Korean Journal of Clinical Laboratory Science / v.53, no.1, 2021 , pp. 19-31 More about this Journal
Abstract
This study aimed at comparing the performance of imported LX detergent cleaning solution (LX-CS) and the self-manufactured OC cleaning solution (OC-CS), based on functional and quantitative analysis. The functional analysis was carried out using apparent diffusion coefficient (ADC) values. For quantitative analysis, precision, linearity, and carry-over rates were measured with commercial control materials according to the Clinical and Laboratory Standards Institute (CLSI) guidelines. Using OC-Sensor PLEDIA (Eiken Chemical, Japan), the ADC value of all cuvettes satisfied the acceptance criteria. For quantitative analysis, precision was less than 5.0% for the two products, and carry-over rates were less than ±1.00%. The linearity slopes and r2 values were 1.0017 and 0.9982 in the LX-CS, and 0.9924 and 0.9996 in the OC-CS, respectively. The correlation coefficient (r) was found to be 0.9997. Also, the percent difference in correlation with 40 artificial-stool specimens was less than 10% and the p-value was less than 0.1. The result of standard deviation ratio (D: ±1 SD ratio) was similar for both products. In conclusion, the functional and quantitative analyses of the two products were compared and showed similar results. In the future, the self-manufactured OC-CS will be able to provide a much more stable and faster supply than the imported LX-CS.
Keywords
Fecal occult blood test; LX detergent cleaning solution; Precision;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Lee ES. Health insurance corporation colorectal cancer screening guidelines. 2nd ed. Goyang: National Cancer Center; 2018. p40-48.
2 Limin X. Clegg, Frederick P. Li, Benjamin F. Hankey, Kenneth Chu, Brenda K. Edwards. Cancer survival among US Whites and Minorities. Arch Intern Med. 2002;162:1985-1993.   DOI
3 Allison JE. Colon cancer screening guidelines 2005: the fecal occult blood test option has become a better FIT. Gastroenterology. 2005;129:745-748.   DOI
4 Allison JE, Tekawa IS, Ransom LJ, Adrain AL. A comparison of fecal occult blood tests for colorectal-cancer screening. N Engl J Med. 1996;334:155-159.   DOI
5 Jochim STD, Frank AO, Rene WMH, Henk AH, Rund JL, Sietze TT, et al. Higher fecal immunochemical test cutoff levels: lower positivity rates but still acceptable detection rates for early-Stage colorectal cancers. Cancer Epidemiol Biomarkers Prev. 2011;20:272-280. https://doi.org/10.1158/10559965   DOI
6 Don C. Rockey. Occult gastrointestinal bleeding. N Engl J Med. 1999;341:38-46.   DOI
7 Shinichi Touzuka, Keiko Aoyama. OC sensor series PLEDIA service manual. 2nd ed. Tokyo: Eiken chemical and Hitachi, Ltd; 2019. p343-368.
8 Jeon CH, Lee AJ, Kim KD. Annual report on external quality assessment scheme for urinalysis and faecal occult blood testing in Korea (2014). J Lab Med Qual Assur. 2015;37:179-189.   DOI
9 Inger DB. Colorectal cancer screening. Prim Care. 1999;26:179-187.   DOI
10 Walter A. Kester. Characterizing and testing A/D and D/A converters for color video applications. IEEE Trans Circuits Sys I Regul Pap.1978;25:539-550.   DOI
11 McEnroe R, Durham A, Goldford M, Kondratovich M, Lababidi S, Magari R, et al. Evaluation of precision of quantitative measurement procedures, approved guideline. CLSI document EP5-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.
12 Daniel W, Martin K, Astles J. Evaluation of the linearity of quantitative measurement procedures: a statistical approach, approved guideline. CLSI document EP6-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2003.
13 Clinical and Laboratory Standards Institute. Verification of comparability of patient results within one health care system, approved guideline. Wayne, PA: Clinical and Laboratory Standards Institute; 2012.
14 Krouwer JS, Cembrowski GS, Tholen DW. Preliminary evaluation of quantitative clinical laboratory measurement procedures, approved guideline. CLSI document EP10-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2006.
15 Nefise O, Filiz K, Dugu P. Comparison of steam technology and a two-step cleaning(water/detergent) and disinfecting(1,000 resp.5,000ppm hypochlorite) method using microfiber cloth for environmental control of multidrug-resistant organisms in an intensive care unit. GMS Hyg Infect Control. 2019;14:1-7. https://doi.org/10.3205/dgkh000330   DOI
16 Budd J, Durham A, Gwise T, Iriarte B, Kallner A, Linnet K, et al. Measurement procedure comparison and bias estimation using subject samples, approved guideline. CLSI document EP09-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2013.
17 Jeon CH, Lee AJ, Kim SG. Suh HS, Bae YC. Annual report on external quality assessment scheme for urinalysis and faecal occult blood testing in Korea (2017). J Lab Med Qual Assur. 2018;40:128-139.   DOI
18 Apolline A, Laure C, Johann-Francois O, Sebastien G, Nicolas C, Jean-Francois B. Comparison of decontamination efficacy of cleaning solutions on a biological safety cabinet workbench contaminated by cyclophosphamide. Can J Hosp Pharm. 2017;70:407-414. https://doi.org/10.4212/cjhp.v70i6.1708   DOI
19 Kim JH, Kim HS, Chung HJ, Yoon NS, Pyo YJ, Bae HG, et al. Evaluation of the OC-SENSOR neo System for testing fecal occult blood. J Lab Med Qual Assur. 2007;27:210-215.
20 Hong SB, Kim HS, Park HS, Lee DH. Evaluation of the HM-JACK automatic analyzer for fecal occult blood test. J Lab Med Qual Assur 2002;24:221-224.
21 Kim DC, Cho SS, Song J, Kim EC, Kim JQ. Evaluation of the OC-SENSOR (Automatic Measuring Apparatus) for immunological fecal occult blood test. J Clin Pathol Qual Control. 1998;20:281-287.
22 Kim JH, Chung HJ, Yoon NS, Pyo YJ, Bae HG. Evaluation of the OC-SENSOR neo System for testing fecal occult blood. Korean J Lab Med. 2007;27:210-215.   DOI
23 Kim BH, Cha YJ. Verification of comparability among quantitative results obtained using multiple instruments within a health care system. Lab Med Online. 2014;4:65-70.   DOI
24 Park YM, Choi Q, Kwon GC, Koo SH. Performance evaluation of the HM-JACKarc analyser for fecal occult blood test. J Lab Med Qual Assur. 2016;38:137-142.   DOI