A Study of Orthognathic Surgical Guides with Two-stage Split Path  

Min Uk, Kim (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Chung Hwan, Park (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Ji Hyoung, Rho (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Eui Sung, Jung (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Young Sang, Park (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Dong Guk, Kim (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Yohan, Seo (New Drug Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Young Jea, Woo (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation)
Jong Min, Lee (Department of Radiology, School of Medicine, Kyungpook National University)
Publication Information
Journal of Biomedical Engineering Research / v.43, no.6, 2022 , pp. 382-389 More about this Journal
Abstract
In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ± 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.
Keywords
Orthognathic surgery; Maxilla; 3D Print; 3D CAD; 3D Scan;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Hashemi S, Parvizi S, Baghbanijavid H, Tan A, Nematollahi M, Ramazani A, Fang N, Elahinia M. Computational modelling of process-structure-property-performance relationships in metal additive manufacturing: a review. International materials reviews. 2022. pp. 1-46. 
2 Ministry of Food and Drug Safety. Understanding Medical Device 3D Printing Technology: Part I Medical Device 3D Printing Technology. 2018. 
3 Schouman T, Rouch P, Imholz B, Fasel J, Courvoisier, D. Accuracy Evaluation of CAD/CAM Generated Splints in Orthognathic Surgery : A Cadaveric Study. Head & Face Medicine. 2015;11(1):11-24. 
4 Rinaldo Z, Giorgio O, Carlo T, Gaetano I, Domenico D, Umberto Z, Ilenia B, Stefano B. Clinical Accuracy of Splintless Maxillary Positioning with Aid of CAD/CAM Fabricated Surgical Cutting Guides and Titanium Plates. J Coatings. 2022;12(10):147-157.    DOI
5 Kim MU, Kim YH, Kim YS. A Study of Orthognathic Surgical Splints Development Using 3D Convergence Technology. J Korean Society for Precision Engineering. 2019;36(3):301-309.    DOI
6 http://snuh.org/health/nMedInfo/nView.do?category=DIS&medid=AA000517 Accessed on 7 Sep 2022. 
7 http://www.yon.co.kr/bbs/content.php?co_id=bjkp03 Accessed on 12 Feb 2022. 
8 http://www.lineupdent.co.kr/58/?pCode=58 Accessed on 17 Feb 2022. 
9 Kang SH, Kim HJ, Park HW, and Lee SH. Maxillary Cutting Guide for Executing a Simulated Osteotomy and Removing the Bony Interference During Orthognathic Surgery. Journal of Medical Devices, 2015, pp. 44-50. 
10 Kim YS, Hong SM, Jung YH, No GW. Computer Aided Design and Applications. Sigma Press. 2018. 
11 Kim MU, Kim YH, Kim YS. A Study of Orthognathic Surgical Splints Development Using 3D Convergence Technology. J Korean Society for Precision Engineering. 2019;36(3):301-309.    DOI
12 https://www.hani.co.kr/arti/economy/biznews Accessed on 27 Dec 2021. 
13 https://www.amc.seoul.kr/asan/healthinfo/management/managementDetail.do?managementId=360 Accessed on 17 Nov 2022. 
14 Vale F, Scherzberg J, Cavaleiro J, Sanz D, Caramelo Fl. 3D Virtual Planning in Orthognathic Surgery and CAD/CAM Surgical Splints Generation in One Patient with Craniofacial Microsomia : A Case Report. Dental Press J Orthodontics. 2016;21(1):89-100.    DOI
15 Li B, Zhang L, Sun H, Yuan J, Shen S. G. A Novel Method of Computer Aided Orthognathic Surgery Using Individual CAD/CAM Templates: A Combination of Osteotomy and Repositioning Guides. British J Oral and Maxillofacial Surgery. 2013;51(8):239-244. 
16 Kim MU, Kim YH, Kim YS. A Study of Orthognathic Surgical Splints Development Using 3D Convergence Technology. J Korean Society for Precision Engineering. 2019;36(3):301-309.   DOI