Pulsatility Estimation of a Pulsatile Decellularizing Device for the Fabrication of Organ Scaffold  

Kim, Dong Sun (Department of Mechanical and Biomedical Engineering, Kangwon National University)
Yang, Se-Ran (Department of Thoracic and Cardiovascular Surgery, Kangwon National University)
Park, Sung Min (Department of Thoracic and Cardiovascular Surgery, Kangwon National University)
Choi, Seong Wook (Department of Mechanical and Biomedical Engineering, Kangwon National University)
Publication Information
Journal of Biomedical Engineering Research / v.38, no.2, 2017 , pp. 62-73 More about this Journal
Abstract
To identify a solution for the restricted availability of healthy lungs and the high risk of immune rejections following organ transplantation, tissue engineering techniques for culturing lungs have been studied by many research groups. The most promising method for culturing lungs is the utilization of a bio-scaffold that was prepared using harvested organs from human donors or other animals by removing their original cells. In this study, a pulsatile perfusion pump was used to alleviate the cell removal effect with the high fluid-dynamic power of the perfusion stream during the decellularization process, while other conventional studies focused on chemical methods to identify efficient detergents. The purpose of this study was to analyze the developed device by using energy equivalent pressure (EEP), which is an indicator of pulsatility, to understand the characteristics of pulsatile energy transmitted according to the load size by using the artificial model and compare it with the measured EEP. The pulsatility of the device can be estimated with the concept of fluid-dynamic energy during a particular constant time period or fluid-dynamic power represented as EEP and EEP increment. Because the measured EEP of perfusion flow during decellularization can be changed by the amount of fluid leakage and the degree of clogging in the capillary vessels, EEP should be measured to determine whether the decellularization is progressing without problems. The decrement of EEP caused by the high perfusion resistance was observed from some experimental results that were obtained with artificial models. EEP can be used to monitor the decellularization process after analyzing the varying EEP according to the amount of load. It was confirmed that the EEP was maintained at a high level in the experiment using the harvested lungs from 12-13-week-old rats. In addition, it was confirmed that the cell removal time was faster than when continuous perfusion was performed. In this study, pulsatile power delivered to the lungs was measured to monitor the process of cell removal, and it serve as the evidence for efficient decellularization.
Keywords
Pulsatility; Decellularization; Scaffold; Energy Equivalent Pressure (EEP); Lung;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Denton M, Magee C, Sayegh M, "Immunosuppressive strategies in transplantation", The Lancet, vol. 353, no. 9158, pp. 1083-1091, 1999.   DOI
2 Peter M. Crapo, Thomas W. Gilbert, Stephen F. Badylak, "An overview of tissue and whole organ decellularization processes", Biomaterials, vol. 32, Issue. 12, pp. 3233-3243, 2011.   DOI
3 SF Badylak, D Taylor, K Uygun, "Whole-organ tissue engineering: decellularization and recellularization of threedimensional matrix scaffolds", Annu Rev Biomed, vol. 13, pp. 27-53, 2011.   DOI
4 SF Badylak, "The extracellular matrix as a scaffold for tissue reconstruction", Semin Cell Dev Biol, vol. 13, Issue 5, pp. 377-383.   DOI
5 Brian Cox, Andrew Emili, "Tissue subcellular fractionation and protein extraction for use in mass-spectrometry-based proteomics", Nature protocols, vol. 1, pp. 1872-1878, 2006.   DOI
6 Giusti S, Bogetti ME, Bonafina A, Fiszer de Plazas S, "An improved method to obtain a soluble nuclear fraction from embryonic brain tissue", Neurochem Res, vol. 34, no. 11, pp. 2022-2029, 2009.   DOI
7 Alhamdani MS, Schroder C, Werner J, Giese N, Bauer A, Hoheisel JD, "Single-step procedure for the isolation of proteins at near-native conditions from mammalian tissue for proteomic analysis on antibody microarrays", J Proteome Res, vol. 9, no. 2, pp. 963-971, 2010.   DOI
8 Patel N, Solanki E, Picciani R, Cavett V, Caldwell-Busby JA, Bhattacharya SK, "Strategies to recover proteins from ocular tissues for proteomics", Proteomics, vol. 8, no. 5, pp. 1055-1070, 2008.   DOI
9 Lynn M. Crosby, Christopher M. Waters, "Epithelial repair mechanisms in the lung", American Journal of Physiology, vol. 298, no. 298, pp. 715-731, 2010.
10 Rolando E. Rumbaut, Perumal Thiagarajan, "Platelet-Vessel Wall Interactions in Hemostasis and Thrombosis", Morgan & Claypool Life Sciences, vol. 2, no. 1, pp. 1-75, 2010.
11 Renata K.daPalma, NoeliaCampillo, JuanJ.Uriarte, Luis V.F.Oliveira, DanielNavajas, RamonFarre, "Pressure-and flowcontrolled media perfusion differently modify vascular mechanics in lung decellularization", journal of the mechanical behavior of biomedical materials, vol. 49, pp. 69-79, 2015.   DOI
12 Nakayama K, Tamiya, Yamamoto K, et al. "High amplitude pulsatile pump in extracorporeal circulation with particular reference to hemodynamics", Surgery, vol. 54, pp. 798-809, 1963.
13 Wright, Gordon. "Hemodynamic analysis could resolve the pulsatile blood flow controversy." The Annals of thoracic surgery, vol. 58, Issue 4, pp. 1199-1204, 1994.   DOI
14 Driessen, J. J. et al. "Pulsatile compared with nonpulsatile perfusion using a centrifugal pump for cardiopulmonary bypass during coronary artery bypass grafting. Effects on systemic haemodynamics, oxygenation, and inflammatory response parameters." Perfusion, vol. 10, no. 1, pp. 3-12. 1995.   DOI
15 Hamulu, Ahmet, et al. "Effects of flow types in cardiopulmonary bypass on gastric intramucosal pH." Perfusion, vol. 13, no. 2, pp. 129-135. 1998.   DOI
16 Undar, Akif, and Charles D. Fraser Jr. "Defining pulsatile perfusion: quantification in terms of energy equivalent pressure." Artificial organs, vol. 23, Issue 8, pp. 712-716. 1999.   DOI
17 Westerhof, Nico, Jan-Willem Lankhaar, and Berend E. Westerhof. "The arterial windkessel." Medical & biological engineering & computing, vol. 47, Issue 2 pp. 131-141, 2009.   DOI
18 Baloa, L. A., J. R. Boston, and J. F. Antaki. "Elastance-based control of a mock circulatory system." Annals of biomedical engineering, vol. 29, Issue 3, pp. 244-251, 2001.   DOI
19 Undar A, Masai T, Yang SQ, Goddard-Finegold J, Frazier OH, Fraser CD Jr. "Effects of perfusion mode on regional and global organ blood flow in a neonatal piglet model", The Ann Thoracic Surgery, vol. 68, Issue 4, pp. 1336-42, 1999.   DOI
20 Stergiopulos, Nikos, Berend E. Westerhof, and Nico Westerhof. "Total arterial inertance as the fourth element of the windkessel model." American Journal of Physiology-Heart and Circulatory Physiology, vol. 276, no.1, pp. 81-88, 1999.   DOI
21 Zuckerman, BRAM D. and F. C. Yin. "Aortic impedance and compliance in hypertensive rats." American Journal of Physiology-Heart and Circulatory Physiology, vol. 257, no. 2, pp. 553-562, 1989.   DOI
22 Bush W, "Overview of transplantation immunology and the pharmacotherapy of adult solid organ transplant recipients: focus on immunosuppression". AACN Advanced Critical Care, vol. 10, no. 2, pp. 253-69, 1999.
23 "2015 장기이식 통계연보", 보건복지부.질병관리본부 장기이식관리센터.
24 Meier-Krische HU, Schold JD, Kaplan B. "Long-erm renal allograft survival; have we made significant progress or is it time to rethink our analytic and therapeutic strategies?", Am J Transplant, vol. 4, pp. 1289-95, 2004.   DOI