Abstract
In this paper we present an implantable pressure sensor to measure real-time blood pressure by monitoring mechanical movement of artery. Sensor is composed of inductors (L) and capacitors (C) which are formed by microfabrication and direct bonding on two biocompatible substrates (quartz). When electrical potential is applied to the sensor, the inductors and capacitors generates a LC resonance circuit and produce characteristic resonant frequencies. Real-time variation of the resonant frequency is monitored by an external measurement system using inductive coupling. Structural and electrical simulation was performed by Computer Aided Engineering (CAE) programs, ANSYS and HFSS, to optimize geometry of sensor. Ultrafast laser (femto-second) cutting and MEMS process were executed as sensor fabrication methods with consideration of brittleness of the substrate and small radial artery size. After whole fabrication processes, we got sensors of $3mm{\times}15mm{\times}0.5mm$. Resonant frequency of the sensor was around 90 MHz at atmosphere (760 mmHg), and the sensor has good linearity without any hysteresis. Longterm (5 years) stability of the sensor was verified by thermal acceleration testing with Arrhenius model. Moreover, in-vitro cytotoxicity test was done to show biocompatiblity of the sensor and validation of real-time blood pressure measurement was verified with animal test by implant of the sensor. By integration with development of external interrogation system, the proposed sensor system will be a promising method to measure real-time blood pressure.
Keywords
Blood pressure sensor; Real-time; Implantable; Direct bonding; Ultrafast laser;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Y.H. Joung, S.J. Go, and E.J. Kim, "A study on the effective chronic disease management", Korea Institute for Health and Social Affairs, 2013.
2 Paul A. James, 2014 Evidence-Based Guideline for the Managementof High Blood Pressure in Adults Report From the Panel Members Appointed to the Eighth Joint National Committee (JNC 8), American Medical Association, 2013, pp. 1-14.
3 J.O. Han, "The Effects of Adherence on Hypertension Control among Newly Diagnosed Hypertension Patients", Health Policy and Management, vol. 24, pp. 136-142, 2014.   DOI
4 S.I. Kim, Development of Implantable Pressure Sensor using MEMS Process and Quartz Wafer Direct Bonding, pp. 4, Hanbat National University, Daejeon, 2015.
5 Jeong Bae Park, "2004 Korea Hypertension Treatment Guideline and Its Perspective", Korean Circulation J., vol. 36, pp. 405-410, 2006.   DOI
6 M.A. Fonseca, J.M. English, M. von Arx and M.G. Allen, "Wireless micromachined ceramic pressure sensor for high-temperature applications", IEEE J. MEMS, vol. 11, pp. 337-343, 2002.   DOI
7 J.V. Edmond, M. Senbos, J.F. Crul, "Invasive vs non-invasive measurement of arterial pressure", Br J Anesth, vol. 57, pp. 434-444, 1985.   DOI
8 P.M. Kearney, M. Whelton, K. Reynolds, P. Muntner, P.K. Whelton, and J. He, "Global burden of hypertension: analysis of worldwide data", The Lancet, vol. 365, pp. 217-223, 2005.   DOI
9 K.H. Shin, C.Y. Moon, T.H. Lee, C.H. Lim, and Y.J. Kim, Proc. IEEE Sensors 2004, Vienna, Austria, 2004, pp. 844.
10 S.S. Mohan, M. del Mar Hershenson, S.P. Boyd and T.H. Lee, "Simple accurate expressions for planar spiral inductances", IEEE J. Solid-State Circuits, vol. 34, pp. 1419-1424, 1999.   DOI
11 T.J. Harpster, B. Stark, and K. Najafi, "A passive wireless integrated humidity sensor", Sens. Actuators A, Phys., vol. 95, pp. 100-107, 2002.   DOI
12 J.C. Butler, A.J. Vigliotti, F.W. Verdi and S.M. Walsh, "Wireless, passive, resonant-circuit, inductively coupled, inductive strain sensor", Sens. Actuators A, Phys., vol. 102, pp. 61-66, 2002.   DOI
13 E.Y. Chow, A.L. Chlebowski, S. Chakraborty, W.J. Chappell and P.P. Irazoqui, "Fully wireless implantable cardiovascular pressure monitor integrated with a medical stent", IEEE Trans. Biomed. Circuits and Systems, vol. 57, pp. 1487-1496, 2010.
14 W.B. Kuhn and N.M. Ibrahim, "Analysis of Current Crowding Effects in Multiturn Spiral Inductors", IEEE Trans on Microwave Theory and Techniques., vol. 49, pp. 31-38, 2001.   DOI
15 Orjan Vallin, Boo Einefors, Christer Hedlund, and Greger Thornell, "DIRECT BONDED QUARTZ RESONATORS", IEEE International Frequency Control Symposium and PDA Exhibition, pp. 345-348, 2001.
16 G. Mancia and G. Gras s, "Improving the management of hypertension: reconsidering efficacy assessment", MEDICOGRAPHIA, vol. 32, pp. 221-329, 2010.
17 A.D. McNaught and A. Wilkinson, Compendium of Chemical Terminology, Oxpord; Gold Book, 1994, pp. 1175.
18 Arrhenius, S.A. Uber die Dissociationswarme und den EinflusB der Temperatur auf den Dissociationsgrad der Elektrolyte. Germany; Wilhelm Engelmann, 1889, pp. 96-116.
19 Semiconductor Device Reliability handbook, Renesas Electronics Corporation, 2013, pp. 260.